Prediction of Cyclic Fatigue Life of Nickel-Titanium Rotary Files by Virtual Modeling and Finite Elements Analysis

被引:26
|
作者
Scattina, Alessandro [1 ]
Alovisi, Mario [2 ]
Paolino, Davide Salvatore [1 ]
Pasqualini, Damiano [2 ]
Scotti, Nicola [2 ]
Chiandussi, Giorgio [1 ]
Berutti, Elio [2 ]
机构
[1] Politecn Torino, Dept Mech & Aerosp Engn, Turin, Italy
[2] Univ Turin, Dept Surg Sci, Div Endodont, Sch Dent, Turin, Italy
关键词
C-S criterion; finite element analysis; multiaxial random fatigue; nickel-titanium rotary instrument; virtual modeling; CLINICAL-USE; ENDODONTIC INSTRUMENTS; MECHANICAL-PROPERTIES; PROTAPER-S1; INSTRUMENTS; GLIDE PATH; M-WIRE; RESISTANCE; DEFECTS; PROFILE; BEHAVIOR;
D O I
10.1016/j.joen.2015.07.010
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: The finite element method (FEM) has been proposed as a method to analyze stress distribution in nickel-titanium (NiTi) rotary instruments but has not been assessed as a method of predicting the number of cycles to failure (NCF). The objective of this study was to predict NCF and failure location of NiTi rotary instruments by FEM virtual simulation of an experimental nonstatic fatigue test. Methods: Pro Taper Next (PTN) X1, X2, and X3 files (Dentsply Maillefer, Baillagues, Switzerland) (n = 20 each) were tested to failure using a customized fatigue testing device. The device and file geometries were replicated with computer-aided design software. Computer-aided design geometries (geometric model) were imported and discretized (numeric model). The typical material model of an M-Wire alloy was applied. The numeric model of the device and file geometries were exported for finite element analysis (FEA). Multiaxial random fatigue methodology was used to analyze stress history and predict instrument life. Experimental data from PIN X2 and X3 were used for virtual model tuning through a reverse engineering approach to optimize material mechanical properties. Tuned material parameters were used to predict the average NCF and failure locations of PTN X1 by FEA; t tests were used to compare FEA and experimental findings (P < .05). Results: Experimental NCF and failure locations did not differ from those predicted with FEA (P = .098). Conclusions: File NCF and failure location may be predicted by FEA. Virtual design, testing, and analysis of file geometries could save considerable time and resources during instrument development.
引用
收藏
页码:1867 / 1870
页数:4
相关论文
共 50 条
  • [31] Cyclic Fatigue Resistance of ProTaper Next and NRT Nickel Titanium Rotary Files
    Elbatouty, Kariem Mostafa
    Elbatal, Laila M.
    Obeid, Maram Mohamed
    ENDO-ENDODONTIC PRACTICE TODAY, 2016, 10 (02): : 89 - 94
  • [32] Cyclic fatigue resistance of two nickel-titanium rotary instruments in interrupted rotation
    Pedulla, E.
    Lizio, A.
    Scibilia, M.
    Grande, N. M.
    Plotino, G.
    Boninelli, S.
    Rapisarda, E.
    Lo Giudice, G.
    INTERNATIONAL ENDODONTIC JOURNAL, 2017, 50 (02) : 194 - 201
  • [33] Effect of Repeated Autoclaving on Cyclic Fatigue of Three Nickel-Titanium Rotary Instruments
    Elbatal, Laila M.
    El Batouty, Kariem Mostafa
    Obeid, Maram Mohamed
    ENDO-ENDODONTIC PRACTICE TODAY, 2016, 10 (01): : 15 - 22
  • [34] Influence of Rotational Speed on the Cyclic Fatigue of Rotary Nickel-Titanium Endodontic Instruments
    Lopes, Helio P.
    Ferreira, Alessandra A. P.
    Elias, Carlos N.
    Moreira, Edson J. L.
    Machado de Oliveira, Julio C.
    Siqueira, Jose F., Jr.
    JOURNAL OF ENDODONTICS, 2009, 35 (07) : 1013 - 1016
  • [35] Response: Cyclic fatigue resistance of ProTaper Next nickel-titanium rotary instruments
    Elnaghy, A. M.
    INTERNATIONAL ENDODONTIC JOURNAL, 2015, 48 (11) : 1101 - 1101
  • [36] Morphometric analysis of shank-to-flute ratio in rotary nickel-titanium files
    Biz, MT
    Figueiredo, JAP
    INTERNATIONAL ENDODONTIC JOURNAL, 2004, 37 (06) : 353 - 358
  • [37] Effects of simulated clinical conditions on nickel-titanium rotary files
    Svec, TA
    Powers, JM
    JOURNAL OF ENDODONTICS, 1999, 25 (11) : 759 - 760
  • [38] Effect of rotational speed on the breakage of nickel-titanium rotary files
    Dietz, DB
    Di Fiore, PM
    Bahcall, JK
    Lautenschlager, EP
    JOURNAL OF ENDODONTICS, 2000, 26 (02) : 68 - 71
  • [39] Defects in rotary nickel-titanium files after clinical use
    Sattapan, B
    Nervo, GJ
    Palamara, JEA
    Messer, HH
    JOURNAL OF ENDODONTICS, 2000, 26 (03) : 161 - 165
  • [40] The deterioration of rotary nickel-titanium files under controlled conditions
    Svec, TA
    Powers, JM
    JOURNAL OF ENDODONTICS, 2002, 28 (02) : 105 - 107