Non-contrast-enhanced renal and abdominal MR angiography using velocity-selective inversion preparation

被引:36
|
作者
Shin, Taehoon [1 ]
Worters, Pauline W. [2 ]
Hu, Bob S. [3 ,4 ]
Nishimura, Dwight G. [1 ]
机构
[1] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
[2] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[3] Palo Alto Med Fdn, Palo Alto, CA USA
[4] Heart Vista Inc, Palo Alto, CA USA
基金
美国国家卫生研究院;
关键词
non-contrast-enhanced MRA; velocity-selective magnetization preparation; renal and abdominal MRA; 3D balanced steady-state free precession; STATE FREE-PRECESSION; LOWER-EXTREMITIES; K-SPACE; RECOVERY; ARTERIES; SSFP; TIME; SUBTRACTION; EXCITATION; PULSES;
D O I
10.1002/mrm.24356
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Non-contrast-enhanced MR angiography is a promising alternative to the established contrast-enhanced approach as it reduces patient discomfort and examination costs and avoids the risk of nephrogenic systemic fibrosis. Inflow-sensitive slab-selective inversion recovery imaging has been used with great promise, particularly for abdominal applications, but has limited craniocaudal coverage due to inflow time constraints. In this work, a new non-contrast-enhanced MR angiography method using velocity-selective inversion preparation is developed and applied to renal and abdominal angiography. Based on the excitation k-space formalism and Shinnar-Le-Roux transform, a velocity-selective excitation pulse is designed that inverts stationary tissues and venous blood while preserving inferiorly flowing arterial blood. As the magnetization of the arterial blood in the abdominal aorta and iliac arteries is well preserved during the magnetization preparation, artery visualization over a large abdominal field of view is achievable with an inversion delay time that is chosen for optimal background suppression. Healthy volunteer tests demonstrate that the proposed method significantly increases the extent of visible arteries compared with the slab-selective approach, covering renal arteries through iliac arteries over a craniocaudal field of view of 340 mm. Magn Reson Med, 2013. (c) 2012 Wiley Periodicals, Inc.
引用
收藏
页码:1268 / 1275
页数:8
相关论文
共 50 条
  • [21] Non-contrast-Enhanced Flow-Independent Peripheral MR Angiography with Balanced SSFP
    Cukur, Tolga
    Lee, Jin H.
    Bangerter, Neal K.
    Hargreaves, Brian A.
    Nishimura, Dwight G.
    MAGNETIC RESONANCE IN MEDICINE, 2009, 61 (06) : 1533 - 1539
  • [22] Non-contrast-enhanced MR angiography for detecting arteriovenous fistula dysfunction in haemodialysis patients
    Jin, W. T.
    Zhang, G. F.
    Liu, H. C.
    Zhang, H.
    Li, B.
    Zhu, X. Q.
    CLINICAL RADIOLOGY, 2015, 70 (08) : 852 - 857
  • [23] Non-Contrast-Enhanced MR Imaging of Renal Artery Stenosis at 1.5 Tesla
    Wilson, Gregory J.
    Maki, Jeffrey H.
    MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2009, 17 (01) : 13 - +
  • [24] Accelerating Non-Contrast-Enhanced MR Angiography with Inflow Inversion Recovery Imaging by Skipped Phase Encoding and Edge Deghosting (SPEED)
    Chang, Zheng
    Xiang, Qing-San
    Shen, Hao
    Yin, Fang-Fang
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2010, 31 (03) : 757 - 765
  • [25] Comparison of axial and coronal acquisitions by non-contrast-enhanced renal 3D MR angiography using flow-in time-spatial labeling inversion pulse
    Junji Takahashi
    Yuki Ohmoto-Sekine
    Takashi Yoshida
    Mitsue Miyazaki
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2020, 33 : 95 - 102
  • [26] Comparison of axial and coronal acquisitions by non-contrast-enhanced renal 3D MR angiography using flow-in time-spatial labeling inversion pulse
    Takahashi, Junji
    Ohmoto-Sekine, Yuki
    Yoshida, Takashi
    Miyazaki, Mitsue
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2020, 33 (01) : 95 - 102
  • [27] Hybrid dynamic bright and black blood angiography by non-contrast-enhanced vessel selective saturation angiography
    Entelmann, Wiebke
    Lindner, Thomas
    Nawka, Marie Teresa
    Fiehler, Jens
    Jansen, Olav
    Huhndorf, Monika
    MAGNETIC RESONANCE IMAGING, 2024, 108 : 22 - 28
  • [28] Initial Clinical Evaluation of a Non-Contrast-Enhanced MR Angiography Method in the Distal Lower Extremities
    Priest, Andrew Nicholas
    Joubert, Ilse
    Winterbottom, Andrew Philip
    See, Teik Choon
    Graves, Martin John
    Lomas, David John
    MAGNETIC RESONANCE IN MEDICINE, 2013, 70 (06) : 1644 - 1652
  • [29] Use of Non-Contrast-Enhanced MR Angiography to Assess Recanalization after Uterine Artery Embolization
    Ishisaki, Juliana Yumi
    Kato, Hitomi
    Kuwatsuru, Yoshiki
    Toei, Hiroshi
    Hoshina, Ayako
    Takemasa, Naoki
    Arai, Masafumi
    Kuwatsuru, Ryohei
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (05)
  • [30] Non-contrast-enhanced peripheral angiography using a sliding interleaved cylinder acquisition
    Kwon, Kie Tae
    Kerr, Adam B.
    Wu, Holden H.
    Hu, Bob S.
    Brittain, Jean H.
    Nishimura, Dwight G.
    MAGNETIC RESONANCE IN MEDICINE, 2015, 74 (03) : 727 - 738