Towards an inline reconstruction architecture for micro-CT systems

被引:28
|
作者
Brasse, D
Humbert, B
Mathelin, C
Rio, MC
Guyonnet, JL
机构
[1] Univ Strasbourg 1, CNRS, Inst Rech Subatom, IN2P3, Strasbourg, France
[2] Inst Genet & Biol Mol & Cellulaire, Illkirch Graffenstaden, France
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2005年 / 50卷 / 24期
关键词
D O I
10.1088/0031-9155/50/24/003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Recent developments in micro-CT have revolutionized the ability to examine in vivo living experimental animal models such as mouse with a spatial resolution less than 50 mu m. The main requirements of in vivo imaging for biological researchers are a good spatial resolution, a low dose induced to the animal during the full examination and a reduced acquisition and reconstruction time for screening purposes. We introduce inline acquisition and reconstruction architecture to obtain in real time the 3D attenuation map of the animal fulfilling the three previous requirements. The micro-CT system is based on commercially available x-ray detector and micro-focus x-ray source. The reconstruction architecture is based on a cluster of PCs where a dedicated communication scheme combining serial and parallel treatments is implemented. In order to obtain high performance transmission rate between the detector and the reconstruction architecture, a dedicated data acquisition system is also developed. With the proposed solution, the time required to filter and backproject a projection of 2048 x 2048 pixels inside a volume of 140 mega voxels using the Feldkamp algorithm is similar to 500 ms, the time needed to acquire the same projection.
引用
收藏
页码:5799 / 5811
页数:13
相关论文
共 50 条
  • [11] 3D micro-CT analysis of cancellous bone architecture
    Rossi, M
    Casali, F
    Romani, D
    Carabini, ML
    [J]. DEVELOPMENTS IN X-RAY TOMOGRAPHY III, 2002, 4503 : 349 - 358
  • [12] Evaluation of a micro-CT scanner applied to the characterization of pulmonary architecture in the rat
    Yang, K
    Miller, D
    Plopper, C
    Lee, D
    Wexler, A
    Boone, J
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1906 - 1906
  • [13] Micro-CT characterization of variability in 3D textile architecture
    Desplentere, F
    Lomov, SV
    Woerdeman, DL
    Verpoest, I
    Wevers, M
    Bogdanovich, A
    [J]. COMPOSITES SCIENCE AND TECHNOLOGY, 2005, 65 (13) : 1920 - 1930
  • [14] Mojette tomographic reconstruction for micro-CT: a bone and vessels quality evaluation
    Sarkissian, H. Der
    Recur, B.
    Guedon, Jp.
    Blerya, P.
    Pilet, P.
    Amouriq, Y.
    [J]. MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [15] Micro-CT for nondestructive 3D reconstruction of MEMS and sensors
    Sasov, A
    [J]. MEMS DESIGN, FABRICATION, CHARACTERIZATION, AND PACKAGING, 2001, 4407 : 388 - 394
  • [16] In-line phase contrast micro-CT reconstruction for biomedical specimens
    Fu, Jian
    Tan, Renbo
    [J]. BIO-MEDICAL MATERIALS AND ENGINEERING, 2014, 24 (01) : 431 - 437
  • [17] Fast high-resolution micro-CT with exact reconstruction methods
    Varslot, T.
    Kingston, A.
    Sheppard, A.
    Sakellariou, A.
    [J]. DEVELOPMENTS IN X-RAY TOMOGRAPHY VII, 2010, 7804
  • [18] Improved Dynamic Micro-CT Imaging Binning for Optimized Image Reconstruction
    Ismail, M. K.
    Xin, Y.
    Hamedani, H.
    Amzajerdian, F.
    Baron, R.
    Loza, L.
    Ruppert, K.
    Kadlecek, S.
    Rizi, R. R.
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2023, 207
  • [19] Micro-CT Guided 3D Reconstruction of Histological Images
    Nagara, Kai
    Roth, Holger R.
    Nakamura, Shota
    Oda, Hirohisa
    Moriya, Takayasu
    Oda, Masahiro
    Mori, Kensaku
    [J]. PATCH-BASED TECHNIQUES IN MEDICAL IMAGING (PATCH-MI 2017), 2017, 10530 : 93 - 101
  • [20] Distributed image reconstruction of micro-CT data: Computation and communication issues
    Benson, TM
    Gregor, J
    [J]. CIC '05: Proceedings of the 2005 International Conference on Communications in Computing, 2005, : 153 - 159