An improved methodology for design of custom-made hip prostheses to be fabricated using additive manufacturing technologies

被引:22
|
作者
Rahmati, Sadegh [1 ]
Abbaszadeh, Farid [2 ]
Farahmand, Farzam [3 ,4 ]
机构
[1] Islamic Azad Univ, Majlesi Branch, Dept Mech Engn, Esfahan, Iran
[2] Islamic Azad Univ, Young Researchers Club, Sci & Res Branch, Tehran, Iran
[3] Univ Tehran Med Sci, RCSTIM, Tehran, Iran
[4] Sharif Univ Technol, Sch Mech Engn, Tehran, Iran
关键词
Rapid prototypes; Manufacturing systems; Prosthetic devices; Computer aided design; Total hip replacement; Fill and fit; Medical rapid prototyping; FEMORAL-NECK ANTEVERSION; INTERNAL CALCAR SEPTUM; COMPUTED-TOMOGRAPHY; PROXIMAL FEMUR; STEM; MORPHOLOGY; 3D;
D O I
10.1108/13552541211250382
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Purpose - The purpose of this paper is to present an improved methodology for design of custom-made hip prostheses, through integration of advanced image processing, computer aided design (CAD) and additive manufacturing (AM) technologies. Design/methodology/approach - The proposed methodology for design of custom-made hip prostheses is based on an independent design criterion for each of the intra-medullary and extra-medullary portions of the prosthesis. The intra-medullar part of the prosthesis is designed using a more accurate and detailed description of the 3D geometry of the femoral intra-medullary cavity, including the septum calcar ridge, so that an improved fill and fit performance is achieved. The extra-medullary portion of the prosthesis is designed based on the anatomical features of the femoral neck, in order to restore the original biomechanical characteristics of the hip joint. The whole design procedure is implemented in a systematic framework to provide a fast, repeatable and non-subjective response which can be further evaluated and modified in a preplanning simulation environment. Findings - The efficacy of the proposed methodology for design of custom-made hip prostheses was evaluated in a case study on a hip dysplasia patient. The cortical bone was distinguished from cancellous in CT images using a thresholding procedure. In particular the septum calcar ridge could be recognized and was incorporated in the design to improve the primary stability of the prosthesis. The lateral and frontal views of the prosthesis, with the patient's images at the background, indicated a close geometrical match with the cortical bone of femoral shaft, and a good compatibility with the anatomy of the proximal femur. Also examination of the cross sections of the prosthesis and the patient's intra-medullary canal at five critical levels revealed close geometrical match in distal stem but less conformity in proximal areas due to preserving the septum calcar ridge. The detailed analysis of the fitting deviation between the prosthesis and point cloud data of the patient's femoral intra-medullary canal, indicated a rest fitting deviation of 0.04 to 0.11 mm in stem. However, relatively large areas of interference fit of 0.04 mm were also found which are considered to be safe and not contributing to the formation of bone cracks. The geometrical analysis of the extra-medullary portion of the prosthesis indicated an anteversion angle of 12.5 degrees and a neck-shaft angle of 131, which are both in the acceptable range. Finally, a time and cost effective investment casting technique, based on AM technology, was used for fabrication of the prosthesis. Originality/value - The proposed design methodology helps to improve the fixation stability of the custom made total hip prostheses and restore the original biomechanical characteristics of the joint. The fabrication procedure, based on AM technology, enables the production of the customized hip prosthesis more accurately, quickly and economically.
引用
收藏
页码:389 / 400
页数:12
相关论文
共 50 条
  • [31] Analysis of principles inspiring design of three-dimensional-printed custom-made prostheses in two referral centres
    Angelini, Andrea
    Kotrych, Daniel
    Trovarelli, Giulia
    Szafranski, Andrzej
    Bohatyrewicz, Andrzej
    Ruggieri, Pietro
    [J]. INTERNATIONAL ORTHOPAEDICS, 2020, 44 (05) : 829 - 837
  • [32] Analysis of principles inspiring design of three-dimensional-printed custom-made prostheses in two referral centres
    Andrea Angelini
    Daniel Kotrych
    Giulia Trovarelli
    Andrzej Szafrański
    Andrzej Bohatyrewicz
    Pietro Ruggieri
    [J]. International Orthopaedics, 2020, 44 : 829 - 837
  • [33] Design of custom-made navigational template of femoral head and pilot research in total hip resurfacing arthroplasty
    Jinlong Liang
    Yonghui Zhao
    Xinjian Gao
    Xuewei Fang
    Yongqing Xu
    Sheng Lu
    [J]. BMC Surgery, 20
  • [34] Design of custom-made navigational template of femoral head and pilot research in total hip resurfacing arthroplasty
    Liang, Jinlong
    Zhao, Yonghui
    Gao, Xinjian
    Fang, Xuewei
    Xu, Yongqing
    Lu, Sheng
    [J]. BMC SURGERY, 2020, 20 (01)
  • [35] A Knowledge-Based Decision-Making Framework for the Design of Manufacturing Networks for Custom-Made Products
    Mourtzis, Dimitris
    Doukas, Michalis
    [J]. ADVANCES IN PRODUCTION MANAGEMENT SYSTEMS: INNOVATIVE AND KNOWLEDGE-BASED PRODUCTION MANAGEMENT IN A GLOBAL-LOCAL WORLD, PT 1, 2014, 438 : 563 - 571
  • [36] Methodology for design process of a snap-fit joint made by additive manufacturing
    Ramirez, Emilio A.
    Caicedo, Fausto
    Hurel, Jorge
    Helguero, Carlos G.
    Luis Amaya, Jorge
    [J]. 12TH CIRP CONFERENCE ON INTELLIGENT COMPUTATION IN MANUFACTURING ENGINEERING, 2019, 79 : 113 - 118
  • [37] The outcome of two-stage arthroplasty using a custom-made interval spacer to treat the infected hip
    Younger, ASE
    Duncan, CP
    Masri, BA
    McGraw, RW
    [J]. JOURNAL OF ARTHROPLASTY, 1997, 12 (06): : 615 - 623
  • [38] Cementless total hip arthroplasty using a custom-made femoral component with sand-blasted surface
    Sakai, T
    Sugano, N
    Matsui, M
    Lee, SB
    Nishii, T
    Haraguchi, K
    Nakata, K
    Ochi, T
    Yoshikawa, H
    Ohzono, K
    [J]. ARTHROPLASTY 2000: RECENT ADVANCES IN TOTAL JOINT REPLACEMENT, 2001, : 111 - 116
  • [39] DESIGN OF CUSTOM HIP STEM PROSTHESES USING 3-DIMENSIONAL CT MODELING
    ROBERTSON, DD
    WALKER, PS
    GRANHOLM, JW
    NELSON, PC
    WEISS, PJ
    FISHMAN, EK
    MAGID, D
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1987, 11 (05) : 804 - 809
  • [40] Additive manufacturing structural redesign of hip prostheses for stress-shielding reduction and improved functionality and safety
    Cortis, Gabriele
    Mileti, Ilaria
    Nalli, Filippo
    Palermo, Eduardo
    Cortese, Luca
    [J]. MECHANICS OF MATERIALS, 2022, 165