Transcranial MR Imaging-Guided Focused Ultrasound Interventions Using Deep Learning Synthesized CT

被引:18
|
作者
Su, P. [1 ,2 ]
Guo, S. [1 ,3 ]
Roys, S. [1 ,3 ]
Maier, F. [4 ]
Bhat, H. [2 ]
Melhem, E. R. [1 ]
Gandhi, D. [1 ]
Gullapalli, R. [1 ,3 ]
Zhuo, J. [1 ,3 ]
机构
[1] Univ Maryland, Sch Med, Dept Diagnost Radiol & Nucl Med, 670 W Baltimore St, Baltimore, MD 21201 USA
[2] Siemens Med Solut USA, Malvern, PA USA
[3] Univ Maryland, Med Ctr, Ctr Metab Imaging & Therapeut, Baltimore, MD 21201 USA
[4] Siemens Healthcare GmbH, Erlangen, Germany
关键词
MAGNETIC-RESONANCE;
D O I
10.3174/ajnr.A6758
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: Transcranial MR imaging-guided focused ultrasound is a promising novel technique to treat multiple disorders and diseases. Planning for transcranial MR imaging-guided focused ultrasound requires both a CT scan for skull density estimation and treatment-planning simulation and an MR imaging for target identification. It is desirable to simplify the clinical workflow of transcranial MR imaging-guided focused ultrasound treatment planning. The purpose of this study was to examine the feasibility of deep learning techniques to convert MR imaging ultrashort TE images directly to synthetic CT of the skull images for use in transcranial MR imaging-guided focused ultrasound treatment planning. MATERIALS AND METHODS: The U-Net neural network was trained and tested on data obtained from 41 subjects (mean age, 66.4 +/- 11.0 years; 15 women). The derived neural network model was evaluated using a k-fold cross-validation method. Derived acoustic properties were verified by comparing the whole skull-density ratio from deep learning synthesized CT of the skull with the reference CT of the skull. In addition, acoustic and temperature simulations were performed using the deep learning CT to predict the target temperature rise during transcranial MR imaging-guided focused ultrasound. RESULTS: The derived deep learning model generates synthetic CT of the skull images that are highly comparable with the true CT of the skull images. Their intensities in Hounsfield units have a spatial correlation coefficient of 0.80 +/- 0.08, a mean absolute error of 104.57 +/- 21.33 HU, and a subject-wise correlation coefficient of 0.91. Furthermore, deep learning CT of the skull is reliable in the skull-density ratio estimation (r = 0.96). A simulation study showed that both the peak target temperatures and temperature distribution from deep learning CT are comparable with those of the reference CT. CONCLUSIONS: The deep learning method can be used to simplify workflow associated with transcranial MR imaging-guided focused ultrasound.
引用
收藏
页码:1841 / 1848
页数:8
相关论文
共 50 条
  • [21] MR Imaging-Guided Focused Ultrasound- Clinical Applications in Managing Malignant Gliomas
    Chen, Huanwen
    Anastasiadis, Pavlos
    Woodworth, Graeme F.
    MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2024, 32 (04) : 673 - 679
  • [22] Magnetic resonance imaging-guided focused ultrasound synovectomy
    Foldes, K
    Hynynen, K
    Shortkroff, S
    Winalski, CS
    Collucci, V
    Koskinen, SK
    McDannold, N
    Jolesz, F
    SCANDINAVIAN JOURNAL OF RHEUMATOLOGY, 1999, 28 (04) : 233 - 237
  • [23] MR Imaging-Guided Cardiovascular Interventions in Young Children
    Tzifa, Aphrodite
    Schaeffter, Tobias
    Razavi, Reza
    MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2012, 20 (01) : 117 - +
  • [24] Uterine leiomyomas: MR imaging-guided focused ultrasound surgery - Results of different treatment protocols
    Fennessy, Fiona M.
    Tempany, Clare M.
    McDannold, Nathan J.
    So, Minna J.
    Hesley, Gina
    Gostout, Bobbie
    Kim, Hyun S.
    Holland, George A.
    Sarti, Dennis A.
    Hynynen, Kullervo
    Jolesz, Ferenc A.
    Stewart, Elizabeth A.
    RADIOLOGY, 2007, 243 (03) : 885 - 893
  • [25] FOCUSED US SYSTEM FOR MR IMAGING-GUIDED TUMOR ABLATION
    CLINE, HE
    HYNYNEN, K
    WATKINS, RD
    ADAMS, WJ
    SCHENCK, JF
    ETTINGER, RH
    FREUND, WR
    VETRO, JP
    JOLESZ, FA
    RADIOLOGY, 1995, 194 (03) : 731 - 737
  • [26] Treatment of uterine fibroids using MR imaging guided focused ultrasound
    Sklair-Levy, M
    Gomori, JM
    Revel, A
    Shoshan, A
    Yagel, S
    RADIOLOGY, 2002, 225 : 440 - 440
  • [27] MR imaging-Guided interventions in the genitourinary tract: An evolving concept
    Fennessy, Fiona M.
    Tuncali, Kemal
    Morrison, Paul R.
    Tempany, Clare M.
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2008, 46 (01) : 149 - +
  • [28] FEASIBILITY OF REAL-TIME MR IMAGING-GUIDED INTERVENTIONS
    JOLESZ, FA
    SILVERMAN, SG
    FRIED, MP
    ALEXANDER, E
    HSU, L
    GLEASON, PL
    RADIOLOGY, 1995, 197 : 278 - 278
  • [29] Thalamic Connectivity in Patients with Essential Tremor Treated with MR Imaging-guided Focused Ultrasound: In Vivo Fiber Tracking by Using Diffusion-Tensor MR Imaging
    Wintermark, Max
    Huss, Diane S.
    Shah, Binit B.
    Tustison, Nicholas
    Druzgal, T. Jason
    Kassell, Neal
    Elias, W. Jeff
    RADIOLOGY, 2014, 272 (01) : 202 - 209
  • [30] Transcranial Magnetic Resonance Imaging-Guided Focused Ultrasound Surgery of Brain Tumors: Initial Findings in 3 Patients
    McDannold, Nathan
    Clement, Greg T.
    Black, Peter
    Jolesz, Ferenc
    Hynynen, Kullervo
    NEUROSURGERY, 2010, 66 (02) : 323 - 332