Image quality of 4D in-treatment CBCT acquired during lung SBRT using FFF beam: a phantom study

被引:3
|
作者
Kim, Jihun
Keum, Ki Chang
Lee, Ho
Hong, Chae-Seon
Park, Kwangwoo
Kim, Jin Sung
机构
[1] Department of Radiation Oncology, Yonsei University College of Medicine, 50-1 Yonsei-ro, Seodaemoon-gu, Seoul
基金
新加坡国家研究基金会;
关键词
Image quality; In-treatment; Cone-beam computed tomography; Stereotactic body radiation therapy; Lung; Flattening filter free; MEGAVOLTAGE SCATTER CORRECTION; BODY RADIATION-THERAPY; MODULATED ARC THERAPY; COMPUTED-TOMOGRAPHY; TUMOR MOTION; VERIFICATION; 3D;
D O I
10.1186/s13014-020-01668-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background Rotational beam delivery enables concurrent acquisition of cone-beam CT (CBCT), thereby facilitating further geometric verification of patient setup during radiation treatment. However, it is challenging to acquire CBCT during stereotactic body radiation therapy (SBRT) using flattening-filter free X-ray beams, in which a high radiation dose is delivered. This study presents quantitative evaluation results of the image quality in four-dimensional (4D) in-treatment CBCT acquired during SBRT delivery. Methods The impact of megavoltage (MV) scatter and acquisition parameters on the image quality was evaluated using Catphan 503 and XSight lung tracking phantoms. The in-treatment CBCT images of the phantoms were acquired while delivering 16 SBRT plans. The uniformity, contrast, and contrast-to-noise ratio (CNR) of the in-treatment CBCT images were calculated and compared to those of CBCT images acquired without SBRT delivery. Furthermore, the localizing accuracy of the moving target in the XSight lung phantom was evaluated for 10 respiratory phases. Results The CNR of the 3D-reconstucted Catphan CBCT images was reduced from 6.3 to 2.6 due to the effect of MV treatment scatter. Both for the Catphan and XSight phantoms, the CBCT image quality was affected by the tube current and monitor units (MUs) of the treatment plan. The lung target in the XSight tracking phantom was most visible for extreme phases; the mean CNRs of the lung target in the in-treatment CBCT images (with 40 mA tube current) across the SBRT plans were 3.2 for the end-of-exhalation phase and 3.0 for the end-of-inhalation phase. The lung target was localized with sub-millimeter accuracy for the extreme respiratory phases. Conclusions Full-arc acquisition with an increased tube current (e.g. 40 mA) is recommended to compensate for degradation in the CBCT image quality due to unflattened MV beam scatter. Acquiring in-treatment CBCT with a high-MU treatment beam is also suggested to improve the resulting CBCT image quality.
引用
收藏
页数:13
相关论文
共 50 条
  • [31] A Prototype Program for Analyzing 4D Image Guidance Shifts for Lung SBRT
    Wang, B.
    Sarkar, V.
    Busselberg, C.
    Rassiah-Szegedi, P.
    Zhao, H.
    Huang, Y.
    Szegedi, M.
    Salter, B.
    MEDICAL PHYSICS, 2012, 39 (06) : 3797 - 3797
  • [32] Difference in performance between 3D and 4D CBCT for lung imaging: a dose and image quality analysis
    Thengumpallil, Sheeba
    Smith, Kathleen
    Monnin, Pascal
    Bourhis, Jean
    Bochud, Francois
    Moeckli, Raphael
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (06): : 97 - 106
  • [33] What is the best reference image for IGRT using 4D CBCT?
    van Herk, M.
    Bryce-Atkinson, A.
    Lindsay, J.
    Faivre-Finn, C.
    Eccles, C.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1480 - S1481
  • [34] BEAM'S EYE VIEW BASED VERIFICATION OF ACTUAL DELIVERY USING IN-TREATMENT 4D CONE-BEAM CT
    Nakagawa, K.
    Haga, A.
    Yamashita, H.
    Kida, S.
    Saotome, N.
    Sakumi, A.
    Imae, T.
    Ohtomo, K.
    Masutani, Y.
    Yoda, K.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S438 - S438
  • [35] Evaluation of Uncertainties in Lung SBRT Treatment Planning Using 4D Dose Calculation and DIR
    Azcona, J.
    Moreno-Jimenez, M.
    Huesa, C.
    Barbes, B.
    Aristu, J.
    Burguete, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E523 - E524
  • [36] Evaluation of 4D cone beam CT-based dose calculation for SBRT lung cancer treatment
    Bellefqih, S.
    Benadon, B.
    Roque, A.
    Gaillot, N.
    Servagi-Vernat, S.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S1105 - S1105
  • [37] Impact of baseline shifts on 4D cone-beam CT images using a 4D phantom driven by lung tumor motions
    Moriwaki, H.
    Shiraishi, K.
    Sakumi, A.
    Ikeda, T.
    Shimizu, W.
    Yoda, K.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S981 - S981
  • [38] 4D Monte Carlo dose calculations for pre-treatment quality assurance of VMAT SBRT: a phantom-based feasibility study
    Roberts, Natalia F.
    Williams, Matthew
    Holloway, Lois
    Metcalfe, Peter
    Oborn, Bradley M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (21):
  • [39] A Novel Workflow for SBRT Image Guidance Using 4D CT Method
    Salter, B.
    Wang, B.
    Zhao, H.
    Rassiah-Szegedi, P.
    Szegedi, M.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [40] 4D dose calculation for SBRT using deformable image registration and probability density function of lung tumor
    Shiinoki, T.
    Kawamura, S.
    Hanazawa, H.
    Uehara, T.
    Yuasa, Y.
    Park, S.
    Koike, M.
    Kanzaki, R.
    Shibuya, K.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S208 - S208