Image quality of 4D in-treatment CBCT acquired during lung SBRT using FFF beam: a phantom study

被引:3
|
作者
Kim, Jihun
Keum, Ki Chang
Lee, Ho
Hong, Chae-Seon
Park, Kwangwoo
Kim, Jin Sung
机构
[1] Department of Radiation Oncology, Yonsei University College of Medicine, 50-1 Yonsei-ro, Seodaemoon-gu, Seoul
基金
新加坡国家研究基金会;
关键词
Image quality; In-treatment; Cone-beam computed tomography; Stereotactic body radiation therapy; Lung; Flattening filter free; MEGAVOLTAGE SCATTER CORRECTION; BODY RADIATION-THERAPY; MODULATED ARC THERAPY; COMPUTED-TOMOGRAPHY; TUMOR MOTION; VERIFICATION; 3D;
D O I
10.1186/s13014-020-01668-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background Rotational beam delivery enables concurrent acquisition of cone-beam CT (CBCT), thereby facilitating further geometric verification of patient setup during radiation treatment. However, it is challenging to acquire CBCT during stereotactic body radiation therapy (SBRT) using flattening-filter free X-ray beams, in which a high radiation dose is delivered. This study presents quantitative evaluation results of the image quality in four-dimensional (4D) in-treatment CBCT acquired during SBRT delivery. Methods The impact of megavoltage (MV) scatter and acquisition parameters on the image quality was evaluated using Catphan 503 and XSight lung tracking phantoms. The in-treatment CBCT images of the phantoms were acquired while delivering 16 SBRT plans. The uniformity, contrast, and contrast-to-noise ratio (CNR) of the in-treatment CBCT images were calculated and compared to those of CBCT images acquired without SBRT delivery. Furthermore, the localizing accuracy of the moving target in the XSight lung phantom was evaluated for 10 respiratory phases. Results The CNR of the 3D-reconstucted Catphan CBCT images was reduced from 6.3 to 2.6 due to the effect of MV treatment scatter. Both for the Catphan and XSight phantoms, the CBCT image quality was affected by the tube current and monitor units (MUs) of the treatment plan. The lung target in the XSight tracking phantom was most visible for extreme phases; the mean CNRs of the lung target in the in-treatment CBCT images (with 40 mA tube current) across the SBRT plans were 3.2 for the end-of-exhalation phase and 3.0 for the end-of-inhalation phase. The lung target was localized with sub-millimeter accuracy for the extreme respiratory phases. Conclusions Full-arc acquisition with an increased tube current (e.g. 40 mA) is recommended to compensate for degradation in the CBCT image quality due to unflattened MV beam scatter. Acquiring in-treatment CBCT with a high-MU treatment beam is also suggested to improve the resulting CBCT image quality.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Image quality of 4D in-treatment CBCT acquired during lung SBRT using FFF beam: a phantom study
    Jihun Kim
    Ki Chang Keum
    Ho Lee
    Chae-Seon Hong
    Kwangwoo Park
    Jin Sung Kim
    Radiation Oncology, 15
  • [2] Image quality of in-treatment 4D-CBCT obtained at various doses in VMAT for SBRT: a phantom study
    Shimohigashi, Y.
    Doi, Y.
    Maruyama, M.
    Yotsuji, Y.
    Kai, Y.
    Toya, R.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S1066 - S1067
  • [3] Determination of PTV Margin for Lung Tumor Using In-Treatment 4D CBCT
    Haga, A.
    Nakagawa, K.
    Yamashita, H.
    Takahashi, W.
    Sakumi, A.
    Masutani, Y.
    Kida, S.
    Saotome, N.
    Itoh, S.
    Shiraki, T.
    Ohtomo, K.
    MEDICAL PHYSICS, 2012, 39 (06) : 3699 - 3699
  • [4] Imaging Quality of Intrafraction 4D CBCT for Lung SBRT Using Flattening Filter-Free Beams: A Phantom Study
    Kim, J.
    Kim, J.
    Keum, K.
    Lee, H.
    MEDICAL PHYSICS, 2019, 46 (06) : E357 - E357
  • [5] Accuracy Verification of 4D Cone Beam CT (CBCT)-Based Lung Tumor Registration by Way of In-Treatment 4D CBCT Analysis
    Nakagawa, K.
    Haga, A.
    Takahashi, W.
    Yamashita, H.
    Imae, T.
    Kida, S.
    Masutani, Y.
    Kotoku, J.
    Sakumi, A.
    Ohtomo, K.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S752 - S752
  • [6] 4D CBCT image guidance in liver SBRT
    Perez-Rozos, A.
    Roman, A.
    Sainz, I. Jerez
    Otero, A.
    Lupianez, Y.
    Medina, J. A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E711 - E711
  • [7] Feasibility of Using Average Intensity Projection of 4D In-Treatment CBCT for Patient Setup Verification
    Kim, J.
    Hong, C.
    Keum, K.
    Kim, J.
    MEDICAL PHYSICS, 2020, 47 (06) : E536 - E536
  • [8] ITV versus mid-ventilation for treatment planning in lung SBRT: a comparison of target coverage and PTV adequacy by using in-treatment 4D cone beam CT
    Bellec, J.
    Arab-Ceschia, F.
    Castelli, J.
    Lafond, C.
    Chajon, E.
    RADIATION ONCOLOGY, 2020, 15 (01)
  • [9] ITV versus mid-ventilation for treatment planning in lung SBRT: a comparison of target coverage and PTV adequacy by using in-treatment 4D cone beam CT
    J. Bellec
    F. Arab-Ceschia
    J. Castelli
    C. Lafond
    E. Chajon
    Radiation Oncology, 15
  • [10] IN-TREATMENT DOSE CALCULATION USING INTREATMENT 4D KILOVOLTAGE CONE-BEAM CT AND IN-TREATMENT LINAC PARAMETERS DURING VOLUMETRIC MODULATED ARC THERAPY FOR A LUNG TUMOR
    Sakumi, A.
    Haga, A.
    Satoshi, K.
    Naoya, S.
    Yoda, K.
    Nakagawa, K.
    RADIOTHERAPY AND ONCOLOGY, 2012, 102 : S125 - S125