Investigation of Hyfrecators and Their In Vitro Interference with Implantable Cardiac Devices

被引:13
|
作者
Weyer, Christopher [1 ]
Siegle, Ronald J. [2 ,3 ]
Girard, Guillaume [4 ]
机构
[1] Cleveland Clin Fdn, Dept Dermatol, Plast Surg Inst, Cleveland, OH 44195 USA
[2] Ctr Surg Dermatol, Westerville, OH USA
[3] Ohio State Univ, Div Dermatol, Columbus, OH 43210 USA
[4] Medtronic Inc, Cardiac Rhythm Management, Mounds View, MN USA
关键词
DERMATOLOGICAL SURGERY;
D O I
10.1111/j.1524-4725.2012.02526.x
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
BACKGROUND Guidelines exist for minimizing potential electromagnetic interference (EMI) with electrosurgical equipment in patients with cardiac rhythm management (CRM) devices. These guidelines encompass all electrosurgical devices but are not specific for hyfrecators. OBJECTIVE To investigate the potential interference of CRM devices by hyfrecators. MATERIALS AND METHODS Using a collagen-based saline gel, three implantable pulse generators (pacemakers) and three implantable cardioverter defibrillators were tested to measure the EMI from two commonly used hyfrecators. The six devices were tested using the hyfrecator under normal use settings and on maximum power. RESULTS Hyfrecators did not interfere with defibrillators and affected pacemakers only when used in close proximity to the device. For the pacemakers, atrial inhibition was observed at a distance of 3 cm on maximum hyfrecator settings and 1 cm at normal use settings. Ventricular inhibition occurred in very close proximity to the device (<1 cm) or in direct contact. CONCLUSION Hyfrecators are safe to use in patients with defibrillators and can be used in pacemaker patients within 2 inches of the device perimeter.
引用
收藏
页码:1843 / 1848
页数:6
相关论文
共 50 条
  • [21] Implantable rhythm devices and electromagnetic interference
    Lamaison, D.
    ARCHIVES DES MALADIES PROFESSIONNELLES ET DE L ENVIRONNEMENT, 2012, 73 (03) : 332 - 333
  • [22] Effects of medically generated electromagnetic interference from medical devices on cardiac implantable electronic devices: A review
    Barmore, Walker
    Patel, Himax
    Voong, Cassandra
    Tarallo, Caroline
    Calkins, Joe B., Jr.
    WORLD JOURNAL OF CARDIOLOGY, 2022, 14 (08): : 446 - 453
  • [23] SENSORS IN IMPLANTABLE CARDIAC DEVICES
    FURMAN, S
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (07): : 1087 - 1088
  • [24] Cardiac Implantable Electronic Devices
    Al-Khatib, Sana M.
    NEW ENGLAND JOURNAL OF MEDICINE, 2024, 390 (05): : 442 - 454
  • [25] Electromagnetic interference effect of dental equipment on cardiac implantable electrical devices: A systematic review
    Niu, Yuting
    Chen, Yufei
    Li, Wenjing
    Xie, Ruiqin
    Deng, Xuliang
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2020, 43 (12): : 1588 - 1598
  • [27] Electromagnetic Interference and Cardiac Implantable Electronic Devices Is the Ride in Electric Cars Still Cool?
    Turagam, Mohit K.
    Deering, Thomas
    Chung, Mina
    Cheng, Jie
    Fisher, John
    Lakkireddy, Dhanunjaya
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2019, 73 (02) : 210 - 213
  • [28] Magnetic Interference on Cardiac Implantable Electronic Devices From Apple iPhone MagSafe Technology
    Nadeem, Fahd
    Nunez Garcia, Arismendy
    Thach Tran, Cao
    Wu, Michael
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2021, 10 (12):
  • [29] Electromagnetic interference between implantable cardiac devices and continuous-flow left ventricular assist devices: a review
    Jonathan S. Gordon
    Elizabeth J. Maynes
    Thomas J. O’Malley
    Behzad B. Pavri
    Vakhtang Tchantchaleishvili
    Journal of Interventional Cardiac Electrophysiology, 2021, 61 : 1 - 10
  • [30] Electromagnetic interference between implantable cardiac devices and continuous-flow left ventricular assist devices: a review
    Gordon, Jonathan S.
    Maynes, Elizabeth J.
    O'Malley, Thomas J.
    Pavri, Behzad B.
    Tchantchaleishvili, Vakhtang
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2021, 61 (01) : 1 - 10