Imaging and spectroscopic performance studies of pixellated CdTe Timepix detector

被引:22
|
作者
Maneuski, D. [1 ]
Astromskas, V. [1 ]
Frojdh, E. [2 ]
Frojdh, C. [2 ]
Gimenez, E. N. [3 ]
Marchal, J. [3 ]
O'Shea, V. [1 ]
Stewart, G. [1 ]
Tartoni, N. [3 ]
Wilhelm, H. [3 ]
Wraight, K. [1 ]
Zain, R. M. [1 ]
机构
[1] Univ Glasgow, Sch Phys & Astron, Glasgow G12 8QQ, Lanark, Scotland
[2] Mid Sweden Univ, S-85170 Sundsvall, Sweden
[3] Diamond Light Source Ltd, Didcot OX11 0DE, Oxon, England
来源
关键词
Gamma detectors (scintillators; CZT; HPG; HgI etc); Hybrid detectors; Instrumentation for synchrotron radiation accelerators; READOUT CHIP;
D O I
10.1088/1748-0221/7/01/C01038
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this work the results on imaging and spectroscopic performances of 14 x 14 x 1 mm CdTe detectors with 55 x 55 mu m and 110 x 110 mu m pixel pitch bump-bonded to a Timepix chip are presented. The performance of the 110 x 110 mu m pixel detector was evaluated at the extreme conditions beam line I15 of the Diamond Light Source. The energy of X-rays was set between 25 and 77 keV. The beam was collimated through the edge slits to 20 mu m FWHM incident in the middle of the pixel. The detector was operated in the time-over-threshold mode, allowing direct energy measurement. Energy in the neighbouring pixels was summed for spectra reconstruction. Energy resolution at 77 keV was found to be Delta E/E = 3.9%. Comparative imaging and energy resolution studies were carried out between two pixel size detectors with a fluorescence target X-ray tube and radioactive sources. The 110 x 110 mu m pixel detector exhibited systematically better energy resolution in comparison to 55 x 55 mu m. An imaging performance of 55 x 55 mu m pixellated CdTe detector was assessed using the Modulation Transfer Function (MTF) technique and compared to the larger pixel. A considerable degradation in MTF was observed for bias voltages below -300V. Significant room for improvement of the detector performance was identified both for imaging and spectroscopy and is discussed.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Investigations on performance and spectroscopic capabilities of a 3 mm CdTe Timepix detector
    Fey, Julian
    Procz, Simon
    Schuetz, Michael K.
    Fiederle, Michael
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 977
  • [2] Multiple Module Pixellated CdTe Spectroscopic X-Ray Detector
    Wilson, Matthew D.
    Bell, Steven J.
    Cernik, Robert J.
    Christodoulou, Christiana
    Egan, Christopher K.
    O'Flynn, Daniel
    Jacques, Simon
    Pani, Silvia
    Scuffham, James
    Seller, Paul
    Sellin, Paul J.
    Speller, Robert
    Veale, Matthew C.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (02) : 1197 - 1200
  • [3] Characterization of a X-ray pixellated CdTe detector with TIMEPIX photon-counting readout chip
    Ruat, Marie
    Ponchut, Cyril
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 4799 - 4803
  • [4] Investigations on the Performance of a 5 mm CdTe Timepix3 Detector for Compton Imaging Applications
    Useche Parra, Juan S.
    Roque, Gerardo
    Schuetz, Michael K.
    Fiederle, Michael
    Procz, Simon
    SENSORS, 2024, 24 (24)
  • [5] High efficiency pixellated CdTe detector
    Bennett, PR
    Shah, KS
    Klugerman, M
    Squillante, MR
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 392 (1-3): : 260 - 263
  • [6] Evaluation of Timepix3 based CdTe photon counting detector for fully spectroscopic small animal SPECT imaging
    Trojanova, E.
    Jakubek, J.
    Turecek, D.
    Sykora, V.
    Francova, P.
    Kolarova, V.
    Sefc, L.
    JOURNAL OF INSTRUMENTATION, 2018, 13
  • [7] Optimization of a pixellated CdZnTe/CdTe detector for a multi-modality imaging system
    Guerra, Pedro
    Dararnbara, Dimitra G.
    Visvikis, Dimitris
    Santos, Andres
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 2976 - +
  • [8] Characteristic of X-ray Tomography Performance using CdTe Timepix Detector
    Zain, R. M.
    O'Shea, V.
    Maneuski, D.
    ADVANCING NUCLEAR SCIENCE AND ENGINEERING FOR SUSTAINABLE NUCLEAR ENERGY KNOWLEDGE, 2017, 1799
  • [9] Optimization of the spectroscopic response of the Timepix detector
    Kroupa, M.
    Jakubek, J.
    Soukup, P.
    JOURNAL OF INSTRUMENTATION, 2012, 7
  • [10] A study of the spectroscopic performance of a CdTe microstrip detector
    Dusi, W
    Auricchio, N
    Brigliadori, L
    Cola, A
    Donati, A
    Landini, G
    Mengoni, D
    Perillo, E
    Siffert, P
    Ventura, G
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (04) : 1026 - 1030