Reassessment of the Access Testosterone chemiluminescence assay and comparison with LC-MS method

被引:18
|
作者
Dittadi, Ruggero [1 ]
Matteucci, Mara [2 ]
Meneghetti, Elisa [1 ]
Ndreu, Rudina [3 ]
机构
[1] ULSS 3 Serenissima, Osped Angelo, UOC Lab Anal, Lab Med Dept, Venice, Italy
[2] Eureka Srl, Lab Div, Chiaravalle, AN, Italy
[3] CNR Clin Physiol Inst, Pisa, Italy
关键词
cross-reaction; DxI platform; immunoassay method; liquid chromatography-mass spectrometry; LoQ; testosterone; IMMUNOASSAYS; MS/MS; SERUM;
D O I
10.1002/jcla.22286
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Aim of the studyTo reassess the imprecision and Limit of Quantitation, to evaluate the cross-reaction with dehydroepiandrosterone-sulfate (DHEAS), the accuracy toward liquid chromatography-mass spectrometry (LC-MS) and the reference interval of the Access Testosterone method, performed by DxI immunoassay platform (Beckman Coulter). Material and MethodsImprecision was evaluated testing six pool samples assayed in 20 different run using two reagents lots. The cross-reaction with DHEAS was studied both by a displacement curve and by spiking DHEAS standard in two serum samples with known amount of testosterone. The comparison with LC-MS was evaluated by Passing-Bablock analysis in 21 routine serum samples and 19 control samples from an External Quality Assurance (EQA) scheme. The reference interval was verified by an indirect estimation on 2445 male and 2838 female outpatients. ResultsThe imprecision study showed a coefficient of variation (CV) between 2.7% and 34.7% for serum pools from 16.3 and 0.27nmol/L. The value of Limit of Quantitation at 20% CV was 0.53nmol/L. The DHEAS showed a cross-reaction of 0.0074%. A comparison with LC-MS showed a trend toward a slight underestimation of immunoassay vs LC-MS (Passing-Bablock equations: DxI=-0.24+0.906 LCMS in serum samples and DxI=-0.299+0.981 LCMS in EQA samples). The verification of reference interval showed a 2.5th-97.5th percentile distribution of 6.6-24.3nmol/L for male over 14years and <0.5-2.78nmol/L for female subjects, in accord with the reference intervals reported by the manufacturer. ConclusionsThe Access Testosterone method could be considered an adequately reliable tool for the testosterone measurement.
引用
下载
收藏
页数:5
相关论文
共 50 条
  • [1] Comparison of a routine LC-MS/MS assay for testosterone against an isotope dilution GCMS reference method
    Keevil, B. G.
    Owen, L. J.
    Blincko, S.
    Ramsay, C. S.
    Sluss, P. M.
    Van Uytfanghe, K.
    Thienpont, L. M.
    CLINICAL CHEMISTRY, 2008, 54 (06) : A104 - A104
  • [2] The bioavailable testosterone assay with LC-MS/MS in aging male
    Horiuchi, T
    Chiba, Y
    Kurosaki, T
    Murayama, T
    Onouchi, T
    PROCEEDINGS OF THE 12TH INTERNATIONAL CONGRESS OF ENDOCRINOLOGY, 2004, : 1283 - 1286
  • [3] The Analysis of Testosterone by LC-MS/MS (A Comparison with Immunoassay).
    McFarlane, John
    Liu, Hua-fen
    Fernandez, Beth
    Stone, Judith
    ENDOCRINE REVIEWS, 2010, 31 (03)
  • [4] Comparison of a method for androstenedione and testosterone using LC-MS/MS versus immunoassay
    Gallagher, L
    Owen, L
    Keevil, B
    CLINICA CHIMICA ACTA, 2005, 355 : S237 - S237
  • [5] Data to support a simultaneous testosterone and estradiol assay in serum by LC-MS/MS
    Schofield, Ryan C.
    Ramanathan, Lakshmi V.
    Pessin, Melissa S.
    Carlow, Dean C.
    DATA IN BRIEF, 2018, 20 : 1160 - 1165
  • [6] A robust LC-MS/MS assay with online cleanup for measurement of serum testosterone
    Zhou, Yu
    Wang, Sihe
    JOURNAL OF SEPARATION SCIENCE, 2019, 42 (15) : 2561 - 2568
  • [7] A sensitive assay for the measurement of testosterone in human serum by LC-MS/MS.
    Nelson, RE
    Singh, RJ
    Grebe, SK
    CLINICAL CHEMISTRY, 2004, 50 (06) : A106 - A106
  • [8] Comparison of two immunoassays with an LC-MS/MS assay for serum cortisol
    Owen, L.
    Wear, J.
    Keevil, B.
    CLINICAL CHEMISTRY, 2008, 54 (06) : A106 - A106
  • [9] VALIDATION OF AN LC-MS/MS ASSAY FOR METHOTREXATE WITH GOOD COMPARISON TO FPIA
    Bjergum, M. W.
    Langman, L. J.
    Snozek, C. L. H.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S801 - S801
  • [10] Development of LC-MS/MS confirmatory method for the determination of testosterone in bovine serum
    Wozniak, Barbara
    Matraszek-Zuchowska, Iwona
    Witek, Sebastian
    Posyniak, Andrzej
    JOURNAL OF VETERINARY RESEARCH, 2017, 61 (01) : 81 - 89