Atrial fibrillation;
photoplethysmogram;
electrocardiogram;
arrhythmia;
data quality;
D O I:
暂无
中图分类号:
R318 [生物医学工程];
学科分类号:
0831 ;
摘要:
A preliminary study into the signal quality of the Empatica's E4 portable photoplethysmogram device as validation for further research on using the device to detect the heart arrhythmia atrial fibrillation. In order to pursue this research, it was necessary to verify the quality of the data produced by the E4 device against a device currently used by clinicians to detect atrial fibrillation. This was done by having healthy volunteers wear both the Empatica E4 and the clinician standard device while following standard clinician procedure for atrial fibrillation diagnosis. The device was compared qualitatively against General Electric's SEER Light Extend Recorder holter portable electrocardiogram. The analysis was done by non-experts at varying academic levels. The primary reviewer's results showed similar data quality between both devices 85% of the time and the holter performed better 5% of the time.