Numerical Model Reduction for the Prediction of Interface Pressure Applied by Compression Bandages on the Lower Leg

被引:16
|
作者
Chassagne, Fanette [1 ,2 ]
Molimard, Jerome [1 ]
Convert, Reynald [2 ]
Giraux, Pascal [3 ]
Badel, Pierre [1 ]
机构
[1] Ecole Natl Super Mines St Etienn, CIS EMSE, SAINBOSE, F-42023 St Etienne, France
[2] Thuasne, F-92307 Levallois Perret, France
[3] Univ Jean Monnet, Fac Med, Dept Phys Med & Rehabil, St Etienne, France
关键词
Finite element simulation; model reduction; compression bandages; interface pressure; IDENTIFICATION; FRICTION; THERAPY; SKIN;
D O I
10.1109/TBME.2017.2774598
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: To develop a new method for the prediction of interface pressure applied by medical compression bandages. Methods: A finite element simulation of bandage application was designed, based on patient-specific leg geometries. For personalized interface pressure prediction, a model reduction approach was proposed, which included the parametrization of the leg geometry. Pressure values computed with this reduced model were then confronted to experimental pressure values. Results: The most influencing parameters were found to be the bandage tension, the skin-to-bandage friction coefficient and the leg morphology. Thanks to the model reduction approach, it was possible to compute interface pressure as a linear combination of these parameters. The pressures computed with this reduced model were in agreement with experimental pressure values measured on 66 patients' legs. Conclusion: This methodology helps to predict patient-specific interface pressure applied by compression bandages within a few minutes whereas it would take a few days for the numerical simulation. The results of this method show less bias than Laplace's Law, which is for now the only other method for interface pressure computation.
引用
收藏
页码:449 / 457
页数:9
相关论文
共 50 条
  • [21] Is compression bandaging accurate? The routine use of interface pressure measurements in compression bandaging of venous leg ulcers
    Satpathy, A
    Hayes, S
    Dodds, S
    PHLEBOLOGY, 2006, 21 (01) : 36 - 40
  • [22] Pressure Prediction Model for Compression Garment Design
    Leung, W. Y.
    Yuen, D. W.
    Ng, Sun Pui
    Shi, S. Q.
    JOURNAL OF BURN CARE & RESEARCH, 2010, 31 (05): : 716 - 727
  • [23] Numerical steady flow solutions of the lower leg venous circulation: effects of external compression
    Fullana, J-M
    Flaud, P.
    EUROPEAN PHYSICAL JOURNAL-APPLIED PHYSICS, 2007, 38 (03): : 287 - 297
  • [24] Numerical simulation of skin pressure distribution applied by graduated compression stockings
    Dai, X. Q.
    Liu, R.
    Li, Y.
    Zhang, M.
    Kwok, Y. L.
    COMPUTATIONAL TEXTILE, 2007, : 301 - +
  • [25] Research on prediction model for formation pressure in compression structure
    Yang, Jin
    Liu, Shujie
    Shi, Lei
    Zhou, Jianliang
    Wang, Pingshuang
    Yang, Xuechang
    Shiyou Xuebao/Acta Petrolei Sinica, 2009, 30 (05): : 764 - 768
  • [26] Controlled, comparative study of relation between volume changes and interface pressure under short-stretch bandages in leg lymphedema patients
    Damstra, Robert J.
    Brouwer, Els R.
    Partsch, Hugo
    DERMATOLOGIC SURGERY, 2008, 34 (06) : 773 - 779
  • [27] Compression stockings for treating venous leg ulcers: measurement of interface pressure under a new ulcer kit
    Partsch, B.
    Partsch, H.
    PHLEBOLOGY, 2008, 23 (01) : 40 - 46
  • [28] A three-parameter model for the venous pressure volume relation of the lower leg
    van Langen, H
    PHLEBOLOGY, 2003, 18 (02) : 83 - 91
  • [29] Intraoperative pressure monitoring of the lower leg for preventing compression-related complications associated with the lithotomy position
    Ryuji Kajitani
    Maiko Minami
    Yuka Kubo
    Haruka Iwaihara
    Yurie Takishita
    Mie Isayama
    Ryo Ohno
    Takaomi Hayashi
    Takahide Sasaki
    Yoshiko Matsumoto
    Hideki Nagano
    Akira Komono
    Naoya Aisu
    Gumpei Yoshimatsu
    Yoichiro Yoshida
    Suguru Hasegawa
    Surgical Endoscopy, 2022, 36 (8) : 5873 - 5881
  • [30] Intraoperative pressure monitoring of the lower leg for preventing compression-related complications associated with the lithotomy position
    Kajitani, Ryuji
    Minami, Maiko
    Kubo, Yuka
    Iwaihara, Haruka
    Takishita, Yurie
    Isayama, Mie
    Ohno, Ryo
    Hayashi, Takaomi
    Sasaki, Takahide
    Matsumoto, Yoshiko
    Nagano, Hideki
    Komono, Akira
    Aisu, Naoya
    Yoshimatsu, Gumpei
    Yoshida, Yoichiro
    Hasegawa, Suguru
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2022, 36 (08): : 5873 - 5881