An Implantable Bio-Micro-system for Drug Monitoring

被引:0
|
作者
Ghoreishizadeh, Sara [1 ]
Kilinc, Enver G. [2 ]
Baj-Rossi, Camilla [1 ]
Dehollain, Catherine [2 ]
Carrara, Sandro [1 ]
De Micheli, Giovanni [1 ]
机构
[1] Ecole Polytech Fed Lausanne, LSI, CH-1015 Lausanne, Switzerland
[2] RFIC, CH-1015 Lausanne, Switzerland
基金
瑞士国家科学基金会;
关键词
Cyclic voltammetry measurement; Drug detection; Power management; Inductive coil; Readout circuit; Integrated circuit;
D O I
暂无
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Multi-target and continuous monitoring by wireless implantable devices is of increasing interest for personalized therapy. In this work an implantable system is presented which is capable of measuring different drugs with Cyclic Voltammetry (CV) method. The wireless microsystem consists of four modules, namely (i) The inductive coil; (ii) Power management IC; (iii) Readout and control IC; (iv) Biosensor array. The power management IC provides 1.8 V with as high as 2 mW power for the readout IC. The configurable readout IC is able to control the biosensor array and measure the sensor current in CV method. CV experiments performed with this microsystem well agree with a commercial equipment for two well known anti-cancer drugs, Etoposide and Mitoxantrone, detection.
引用
收藏
页码:218 / 221
页数:4
相关论文
共 50 条
  • [1] Fabrication of Bio-Composite Scaffold for Implantable Drug Delivery System (DDS)
    Chu, Won-Shik
    Pandey, Jitendra K.
    Ahn, Sung-Hoon
    JOURNAL OF BIOBASED MATERIALS AND BIOENERGY, 2014, 8 (02) : 230 - 239
  • [2] Study on structural optimum design of implantable drug delivery micro-system
    Wang, Xiaopeng
    Chen, Tianning
    Yang, Zhanxiao
    Wang, Wanjun
    SIMULATION MODELLING PRACTICE AND THEORY, 2007, 15 (01) : 47 - 56
  • [3] Characteristics of Biodegradable Implantable Drug Delivery System with Micro-porous Structure
    Gao, Yang
    Wang, Xiaopeng
    Chen, Tianning
    NEMS/MEMS TECHNOLOGY AND DEVICES, 2009, 74 : 307 - 310
  • [4] Implantable Resorbable Micro-prism Arrays for Real-time Drug Release Monitoring
    Tao, H.
    Siebert, S. M.
    Pritchard, E. M.
    Sassaroli, A.
    Panilaitis, B.
    Brenckle, M. A.
    Amsden, J. J.
    Levitt, J. M.
    Fantini, S.
    Kaplan, D. L.
    Omenetto, F. G.
    2012 CONFERENCE ON LASERS AND ELECTRO-OPTICS (CLEO), 2012,
  • [5] Flexible electronics for bio-signal monitoring in implantable applications
    Yoshimoto, Shusuke
    Araki, Teppei
    Uemura, Takafumi
    Noda, Yuki
    Sekitani, Tsuyoshi
    IEICE ELECTRONICS EXPRESS, 2017, 14 (20):
  • [6] A Wireless Powered Implantable Bio-Sensor Tag System-on-Chip for Continuous Glucose Monitoring
    Guan, Shuo
    Gu, Jingren
    Shen, Zhonghan
    Wang, Junyu
    Huang, Yue
    Mason, Andrew
    2011 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE (BIOCAS), 2011, : 193 - 196
  • [7] An implantable drug-delivery system based on shape memory alloy micro-actuation
    Reynaerts, D
    Peirs, J
    VanBrussel, H
    SENSORS AND ACTUATORS A-PHYSICAL, 1997, 61 (1-3) : 455 - 462
  • [8] Release Model and Characteristics of Biodegradable Implantable Drug Delivery System with Micro-porous Structure
    Gao, Yang
    Wang, Xiaopeng
    Chen, Tianning
    PROCEEDINGS OF THE THIRD INTERNATIONAL CONFERENCE ON MECHANICAL ENGINEERING AND MECHANICS, VOLS 1 AND 2, 2009, : 1637 - 1642
  • [9] A MULTIPURPOSE IMPLANTABLE DRUG ADMINISTRATION SYSTEM
    BARTELT, K
    STEWART, B
    HANKNER, D
    WESTON, D
    VANKAMPEN, K
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (08) : 523 - 523
  • [10] THE INFUSAID - AN IMPLANTABLE DRUG DELIVERY SYSTEM
    POLLOCK, S
    SADOW, S
    VETERINARY MEDICINE & SMALL ANIMAL CLINICIAN, 1982, 77 (06): : 897 - 901