Rapid diagnosis of catheter-related sepsis using the acridine orange leukocyte cytospin test and an endoluminal brush

被引:26
|
作者
Tighe, MJ
Kite, P
Thomas, D
Fawley, WN
McMahon, MJ
机构
[1] UNIV LEEDS,GEN INFIRM,DEPT SURG,DEPT MICROBIOL,LEEDS LS1 3EX,W YORKSHIRE,ENGLAND
[2] UNIV LEEDS,GEN INFIRM,NUTR SUPPORT SERV,LEEDS LS1 3EX,W YORKSHIRE,ENGLAND
关键词
D O I
10.1177/0148607196020003215
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Background: In neonates, the acridine orange leukocyte cytospin (AOLC) test has been found to be a highly sensitive test for the detection of infected IV catheters in situ, which provides a result in less than 1 hour. Preliminary data suggested that the AOLC test was of limited value in adults. We report here a modification of the test for adult patients with indwelling central venous catheters. Methods: A prospective study was performed on two groups of 50 adult patients with suspected sepsis and a central venous catheter The AOLC test was carried out after the clinical decision to remove the catheter had been made. In group 1 patients, a blood sample was withdrawn from the catheter for the AOLC test. In the patients in group 2, an endoluminal brush was used to ''sweep'' the catheter before the collection of the blood sample. Results of the AOLC test were compared with culture of the removed catheter tip. Results: From the catheters in group 1 (no brush), 17 catheter tips were found to be infected, but the AOLC was positive in only two patients (12%). In group 2 (brush), 18 tips were infected, and the AOLC test was positive in 15 patients (83%). The use of the endoluminal brush significantly improved the yield of the AOLC test (p < .01) to levels reported in neonates. The AOLC test produced no false positives in either group. Conclusion: When used independently, the AOLC test was not sensitive enough to detect catheter-related sepsis. However, in combination with an endoluminal brush, the AOLC test was much more sensitive and has the potential to provide a simple, rapid, and accurate diagnostic test for catheter-related sepsis, which does not require removal of the catheter.
引用
收藏
页码:215 / 218
页数:4
相关论文
共 50 条
  • [41] In situ diagnosis of intravascular catheter-related bloodstream infection:: A comparison of quantitative culture, differential time to positivity, and endoluminal brushing
    Catton, JA
    Dobbins, BM
    Kite, P
    Wood, JM
    Eastwood, K
    Sugden, S
    Sandoe, JAT
    Burke, D
    McMahon, MJ
    Wilcox, MH
    CRITICAL CARE MEDICINE, 2005, 33 (04) : 787 - 791
  • [42] A rapid ELISA for the diagnosis of intravascular catheter related sepsis caused by coagulase negative staphylococci
    Worthington, T
    Lambert, PA
    Traube, A
    Elliott, TSJ
    JOURNAL OF CLINICAL PATHOLOGY, 2002, 55 (01) : 41 - 43
  • [43] Nosocomial coagulase negative staphylococcal (CoNS) catheter-related sepsis in preterm infants: Definition, diagnosis, prophylaxis, and prevention
    Craft A.
    Finer N.
    Journal of Perinatology, 2001, 21 (3) : 186 - 192
  • [44] The use of a rapid in situ test in the detection of central venous catheter-related bloodstream infection:: A prospective study
    Bong, JJ
    Kite, P
    Ammori, BJ
    Wilcox, MH
    McMahon, MJ
    JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 2003, 27 (02) : 146 - 150
  • [45] Rapid Diagnosis of Staphylococcal Catheter-Related Bacteraemia in Direct Blood Samples by Real-Time PCR
    Zboromyrska, Yuliya
    De la Calle, Cristina
    Soto, Marcelo
    Sampietro-Colom, Laura
    Soriano, Alex
    Jose Alvarez-Martinez, Mriam
    Almela, Manel
    Marco, Francesc
    Arjona, Ruth
    Cobos-Trigueros, Nazaret
    Morata, Laura
    Mensa, Jose
    Antonio Martinez, Jose
    Mira, Aurea
    Vila, Jordi
    PLOS ONE, 2016, 11 (08):
  • [46] Evaluation of Acridine-Orange microscopy and the Paracheck Pf rapid antigen-detection test in the diagnosis of Plasmodium falciparum malaria
    Arora, S
    Shinkre, N
    Koppikar, G
    ANNALS OF TROPICAL MEDICINE AND PARASITOLOGY, 2003, 97 (06): : 655 - 656
  • [47] Diagnosis of catheter-related bloodstream infection in a total parenteral nutrition population: inclusion of sepsis defervescence after removal of culture-positive central venous catheter
    Walshe, C. M.
    Boner, K. S.
    Bourke, J.
    Hone, R.
    Phelan, D.
    JOURNAL OF HOSPITAL INFECTION, 2010, 76 (02) : 119 - 123
  • [48] Differential time to positivity of central and peripheral blood cultures is inaccurate for the diagnosis of Staphylococcus aureus long-term catheter-related sepsis
    Bouzidi, H.
    Emirian, A.
    Marty, A.
    Chachaty, E.
    Laplanche, A.
    Gachot, B.
    Blot, F.
    JOURNAL OF HOSPITAL INFECTION, 2018, 99 (02) : 192 - 199
  • [49] Catheter-related Mycobacterium fortuitum Bloodstream Infection: Rapid Identification Using MALDI-TOF Mass Spectrometry
    Artacho-Reinoso, M. J.
    Olbrich, P.
    Solano-Paez, P.
    Ybot-Gonzalez, P.
    Lepe, J. A.
    Neth, O.
    Aznar, J.
    KLINISCHE PADIATRIE, 2014, 226 (02): : 68 - 71
  • [50] Improved diagnosis of central venous catheter-related bloodstream infections using the HB&L UROQUATTRO™ system
    C. Fontana
    M. Favaro
    M. C. Bossa
    S. Minelli
    A. Altieri
    M. Pelliccioni
    F. Falcione
    L. Di Traglia
    O. Cicchetti
    C. Favalli
    European Journal of Clinical Microbiology & Infectious Diseases, 2012, 31 : 3139 - 3144