Characteristics of detection thresholds and maximum comfortable loudness levels as a function of pulse rate in human cochlear implant users

被引:32
|
作者
Zhou, Ning [1 ]
Xu, Li [1 ,2 ]
Pfingst, Bryan E. [1 ]
机构
[1] Univ Michigan, Dept Otolaryngol, Kresge Hearing Res Inst, Ann Arbor, MI 48109 USA
[2] Ohio Univ, Sch Rehabil & Commun Sci, Athens, OH 45701 USA
关键词
PULSATILE ELECTRICAL-STIMULATION; MODULATION DETECTION; INTENSITY DISCRIMINATION; TEMPORAL INTEGRATION; SPEECH-PERCEPTION; ELECTRODE; POSITION; RECOGNITION; FREQUENCY; GROWTH;
D O I
10.1016/j.heares.2011.12.008
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
The ability of an implanted ear to integrate multiple pulses, as measured by the slopes of detection threshold level (T level) versus pulse rate functions, may reflect cochlear health in the cochlea, as suggested by previous animal studies (Kang et al., 2010; Pfingst et al., 2011). In the current study, we examined the slopes of T level versus pulse rate functions in human subjects with cochlear implants. Typically, T levels decrease as a function of pulse rate, consistent with a multipulse integration mechanism. The magnitudes of the slopes of the T level versus pulse rate functions obtained from the human subjects were comparable to those reported in the animal studies. The slopes varied across stimulation sites, but did not change systematically along the tonotopic axis. This suggests that the slopes are dependent on local conditions near the individual stimulation sites. The characteristics of these functions were also similar to those found in animals in that the slopes for higher pulse rates were steeper than those for the lower pulse rates, consistent with a combined effect of multipulse integration and cumulative partial depolarization mechanisms at rates above 1000 pps. The maximum comfortable loudness level (C level) versus pulse rate functions were also examined to determine the effect of level on the slopes. Slopes of C-level functions were shallower than those for the T-level functions and were not correlated with those of the T-level functions, so the mechanisms underlying these two functions are probably not identical. The slopes of the T- or C-level functions were not dependent on stimulus-current level. Based on these results, we suggest that slopes of T level versus pulse rate functions might be a useful measure for estimating nerve survival in the cochlea in regions close to the stimulation sites. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:25 / 32
页数:8
相关论文
共 50 条
  • [1] Across-site variation in detection thresholds and maximum comfortable loudness levels for cochlear implants
    Pfingst, BE
    Xu, L
    JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2004, 5 (01): : 11 - 24
  • [2] Across-Site Variation in Detection Thresholds and Maximum Comfortable Loudness Levels for Cochlear Implants
    Bryan E. Pfingst
    Li Xu
    Journal of the Association for Research in Otolaryngology, 2004, 5 : 11 - 24
  • [3] Effect of Pulse Rate on Loudness Discrimination in Cochlear Implant Users
    Azadpour, Mahan
    McKay, Colette M.
    Svirsky, Mario A.
    JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2018, 19 (03): : 287 - 299
  • [4] Effect of Pulse Rate on Loudness Discrimination in Cochlear Implant Users
    Mahan Azadpour
    Colette M. McKay
    Mario A. Svirsky
    Journal of the Association for Research in Otolaryngology, 2018, 19 : 287 - 299
  • [5] Analysis of electrical thresholds and maximum comfortable levels in cochlear implant patients
    Luis Vargas, Jose
    Sainz, Manuel
    Roldan, Cristina
    Alvarez, Isaac
    de la Torre, Angel
    AURIS NASUS LARYNX, 2013, 40 (03) : 260 - 265
  • [6] Effect of pulse rate on loudness growth functions in cochlear implant users
    Geissler, Gunnar
    Hey, Matthias
    Boehnke, Britta
    Fredelake, Stefan
    Frohne-Buechner, Caroline
    Buechner, Andreas
    Mueller-Deile, Joachim
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2014, 59 : S761 - +
  • [7] Rate modulation detection thresholds for cochlear implant users
    Brochier, Tim
    McKay, Colette
    McDermott, Hugh
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2018, 143 (02): : 1214 - 1222
  • [8] Relationship between gap detection thresholds and loudness in cochlear-implant users
    Garadat, Soha N.
    Pfingst, Bryan E.
    HEARING RESEARCH, 2011, 275 (1-2) : 130 - 138
  • [9] Effect of Pulse Polarity on Thresholds and on Non-monotonic Loudness Growth in Cochlear Implant Users
    Olivier Macherey
    Robert P. Carlyon
    Jacques Chatron
    Stéphane Roman
    Journal of the Association for Research in Otolaryngology, 2017, 18 : 513 - 527
  • [10] Effect of Pulse Polarity on Thresholds and on Non-monotonic Loudness Growth in Cochlear Implant Users
    Macherey, Olivier
    Carlyon, Robert P.
    Chatron, Jacques
    Roman, Stephane
    JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2017, 18 (03): : 513 - 527