Extending laboratory automation to the wards: effect of an innovative pneumatic tube system on diagnostic samples and transport time

被引:15
|
作者
Suchsland, Juliane [2 ]
Winter, Theresa [2 ]
Greiser, Anne [2 ]
Streichert, Thomas [3 ]
Otto, Benjamin [1 ,4 ]
Mayerle, Julia [5 ]
Runge, Soeren [5 ]
Kallner, Anders [6 ]
Nauck, Matthias [2 ,7 ]
Petersmann, Astrid [2 ]
机构
[1] Eppendorf AG, Hamburg, Germany
[2] Univ Med Greifswald, Inst Clin Chem & Lab Med, Ferdinand Sauerbruch Str, D-17475 Greifswald, Germany
[3] Univ Cologne, Fac Med, Inst Clin Chem, Cologne, Germany
[4] Univ Med Ctr, Dept Internal Med, Hamburg, Germany
[5] Univ Med Greifswald, Dept Med A, Greifswald, Germany
[6] Karolinska Univ Hosp, Dept Clin Chem, Stockholm, Sweden
[7] Univ Med, Partner Site Greifswald, DZHK German Ctr Cardiovasc Res, Greifswald, Germany
关键词
laboratory automation; pneumatic tube system; pre-analytical phase; sample transport; INCREASED HEMOLYSIS; DELIVERY-SYSTEM; INTERFERENCE;
D O I
10.1515/cclm-2016-0380
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The innovative pneumatic tube system (iPTS) transports one sample at a time without the use of cartridges and allows rapid sending of samples directly into the bulk loader of a laboratory automation system (LAS). We investigated effects of the iPTS on samples and turn-around time (TAT). Methods: During transport, a mini data logger recorded the accelerations in three dimensions and reported them in arbitrary area under the curve (AUC) units. In addition representative quantities of clinical chemistry, hematology and coagulation were measured and compared in 20 blood sample pairs transported by iPTS and courier. Results: Samples transported by iPTS were brought to the laboratory (300 m) within 30 s without adverse effects on the samples. The information retrieved from the data logger showed a median AUC of 7 and 310 arbitrary units for courier and iPTS transport, respectively. This is considerably below the reported limit for noticeable hemolysis of 500 arbitrary units. Conclusions: iPTS reduces TAT by reducing the hands-on time and a fast transport. No differences in the measurement results were found for any of the investigated 36 analytes between courier and iPTS transport. Based on these findings the iPTS was cleared for clinical use in our hospital.
引用
收藏
页码:225 / 230
页数:6
相关论文
共 28 条
  • [21] Validation of a pneumatic tube system to transport surgical pathology biopsy samples (vol 56, pg 99, 2017)
    Guibourg, Briac
    Marcorelles, Pascale
    Uguen, Arnaud
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (04) : 681 - 681
  • [22] Effect of a pneumatic tube delivery system on quality and timeliness of laboratory test results for emergency department patients
    Fernandes, CM
    Worster, A
    Eva, K
    Hill, S
    McCallum, C
    [J]. ANNALS OF EMERGENCY MEDICINE, 2003, 42 (04) : S22 - S22
  • [23] Assessment of Platelet Function in Whole Blood by Multiple Electrode Aggregometry: Transport of Samples Using a Pneumatic Tube System Reply
    Paniccia, Rita
    Gori, Anna Maria
    Marcucci, Rossella
    Prisco, Domenico
    Abbate, Rosanna
    Antonucci, Emilia
    Maggini, Niccolo
    Romano, Eloisa
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2009, 132 (05) : 803 - 804
  • [24] The Effect of a Pneumatic Tube Transport System on Platelet Aggregation Using Optical Aggregometry and the PFA-100™
    Huebner, Ulrich
    Boeckel-Frohnhoefer, Nicole
    Hummel, Bjoern
    Geisel, Juergen
    [J]. CLINICAL LABORATORY, 2010, 56 (1-2) : 59 - 64
  • [25] EFFECT OF SAMPLE TRANSPORT BY TEMPUS600 PNEUMATIC TUBE SYSTEM ON PLATELETS, PLATELET FUNCTION TESTS AND THROMBOELASTOMETRY
    De Kesel, Pieter
    Devreese, Katrien
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2019, 41 : 49 - 49
  • [26] The comparison of blood gas analysis (BGA) parameters in samples transported by the pneumatic tube system and manual transport at Dr. Soetomo Hospital, Surabaya, Indonesia
    Gunawan, Hantoro
    Marpaung, Ferdy Royland
    [J]. BALI MEDICAL JOURNAL, 2019, 8 (02) : 400 - 403
  • [27] THE RAPID-TRANSPORT OF ARTERIAL BLOOD-GAS SPECIMENS VIA PNEUMATIC TUBE AND A MICROCOMPUTER BASED SYSTEM FOR QUALITY-CONTROL OF TURNAROUND TIME
    JONES, JB
    SHULSKI, A
    MILLER, C
    MORAN, JJ
    [J]. CLINICAL CHEMISTRY, 1982, 28 (07) : 1610 - 1610
  • [28] Pneumatic tube system transport does not alter platelet function in optical and whole blood aggregometry, prothrombin time, activated partial thromboplastin time, platelet count and fibrinogen in patients on anti-platelet drug therapy
    Enko, Dietmar
    Mangge, Harald
    Muench, Andreas
    Niedrist, Tobias
    Mahla, Elisabeth
    Metzler, Helfried
    Prueller, Florian
    [J]. BIOCHEMIA MEDICA, 2017, 27 (01) : 217 - 224