A comparison of the influence of three different lumbar interbody fusion approaches on stress in the pedicle screw fixation system: Finite element static and vibration analyses

被引:23
|
作者
Fan, Wei [1 ]
Guo, Li-Xin [1 ]
机构
[1] Northeastern Univ, Sch Mech Engn & Automat, 3-11 Wenhua Rd, Shenyang 110819, Liaoning, Peoples R China
基金
中国国家自然科学基金;
关键词
bilateral pedicle screw fixation; breakage; finite element; lumbar interbody fusion; static and vibration analyses; BIOMECHANICAL ANALYSIS; SPINE; INSTRUMENTATION; LOAD; COMPLICATIONS; STABILITY; POSTERIOR; BEHAVIOR; FAILURE; DEVICE;
D O I
10.1002/cnm.3162
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study aimed to examine breakage risk of the bilateral pedicle screw (BPS) fixation system under static and vibration loadings after three different types of lumbar interbody fusion surgery. A previously validated intact L1-sacrum finite element model was modified to simulate anterior, posterior, and transforaminal lumbar interbody fusion (ALIF, PLIF, and TLIF, respectively) with BPS fixation system (consisting of pedicle screws and rigid connecting rods) at L4-L5. As a risk parameter for breakage, the von Mises stresses in the pedicle screws and the rods for the ALIF, PLIF, and TLIF models under static loading (flexion, extension, lateral bending, and axial torsion moments) and vibration loading (sinusoidal vertical load) were calculated and compared. The calculated von Mises stresses were different in the ALIF, PLIF, and TLIF models, but these stresses for all the fusion models were found to be concentrated in neck of the pedicle screw and middle of the rod under both the static and vibration loadings. The results from static analyses showed that the maximum stress in the BPS fixation system was greater in the TLIF model than in the ALIF and PLIF models under all the applied static loadings. The results from transient dynamic analyses also showed that the TLIF generated greater dynamic responses of the stress in the BPS fixation system to the vertical vibration compared with the ALIF and PLIF. It implies that the TLIF procedure might incur a higher risk of breakage for the BPS fixation system than the ALIF and PLIF procedures.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Biomechanical Comparison of Unilateral and Bilateral Pedicle Screw Fixation after Multilevel Lumbar Lateral Interbody Fusion
    Mok, James M.
    Forsthoefel, Craig
    Diaz, Roberto Leonardo
    Lin, Ye
    Amirouche, Farid
    GLOBAL SPINE JOURNAL, 2024, 14 (05) : 1524 - 1531
  • [22] Unilateral versus bilateral pedicle screw fixation with anterior lumbar interbody fusion: a comparison of postoperative outcomes
    Levy, Hannah A.
    Karamian, Brian A.
    Freedman, Brett A.
    Sebastian, Arjun S.
    EUROPEAN SPINE JOURNAL, 2024, 33 (11) : 4406 - 4407
  • [23] Comparison of unilateral versus bilateral pedicle screw fixation in lumbar interbody fusion: a meta-analysis
    Wenbin Ding
    Yile Chen
    Hui Liu
    Jianru Wang
    Zhaomin Zheng
    European Spine Journal, 2014, 23 : 395 - 403
  • [24] Comparison of unilateral versus bilateral pedicle screw fixation in lumbar interbody fusion: a meta-analysis
    Ding, Wenbin
    Chen, Yile
    Liu, Hui
    Wang, Jianru
    Zheng, Zhaomin
    EUROPEAN SPINE JOURNAL, 2014, 23 (02) : 395 - 403
  • [25] Computational comparison of anterior lumbar interbody fusion and oblique lumbar interbody fusion with various supplementary fixation systems: a finite element analysis
    Ouyang, Pengrong
    Tan, Qinghua
    He, Xijing
    Zhao, Bo
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2023, 18 (01)
  • [26] Computational comparison of anterior lumbar interbody fusion and oblique lumbar interbody fusion with various supplementary fixation systems: a finite element analysis
    Pengrong Ouyang
    Qinghua Tan
    Xijing He
    Bo Zhao
    Journal of Orthopaedic Surgery and Research, 18
  • [27] Posterior Lumbar Interbody Fusion: Comparison of Single Intervertebral Cage and Single Side Pedicle Screw Fixation versus Bilateral Cages and Screw Fixation
    Moreland, D. B.
    Asch, H. L.
    Czajka, G. A.
    Overkamp, J. A.
    Sitzman, D. M.
    MINIMALLY INVASIVE NEUROSURGERY, 2009, 52 (03) : 132 - 136
  • [28] Letter to the Editor Regarding: "Biomechanical Comparison of Stand-Alone and Bilateral Pedicle Screw Fixation for Oblique Lumbar Interbody Fusion Surgery-A Finite Element Analysis"
    Tartara, Fulvio
    Cofano, Fabio
    Di Perna, Giuseppe
    Tancioni, Flavio
    Garbossa, Diego
    WORLD NEUROSURGERY, 2021, 150 : 248 - 249
  • [29] Biomechanical comparison of unilateral and bilateral pedicle screws fixation for transforaminal lumbar interbody fusion after decompressive surgery - a finite element analysis
    Chen, Shih-Hao
    Lin, Shang-Chih
    Tsai, Wen-Chi
    Wang, Chih-Wei
    Chao, Shih-Heng
    BMC MUSCULOSKELETAL DISORDERS, 2012, 13
  • [30] Biomechanical comparison of unilateral and bilateral pedicle screws fixation for transforaminal lumbar interbody fusion after decompressive surgery -- a finite element analysis
    Shih-Hao Chen
    Shang-Chih Lin
    Wen-Chi Tsai
    Chih-Wei Wang
    Shih-Heng Chao
    BMC Musculoskeletal Disorders, 13