Aspheric and free-form surfaces test with non-null sub-aperture stitching

被引:1
|
作者
Liu, Dong [1 ]
Zhou, Yuhao [1 ]
Bai, Jian [1 ]
Shi, Tu [1 ]
Shen, Yibing [1 ]
Yang, Yongying [1 ]
Zhang, Lei [1 ]
机构
[1] Zhejiang Univ, Coll Opt & Engn, State Key Lab Modern Opt Instrumentat, Hangzhou 310027, Zhejiang, Peoples R China
来源
基金
中国国家自然科学基金;
关键词
free-form surface; irregular sub-aperture; aspheric testing; interferometry;
D O I
10.1117/12.2246149
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A free-form surface sub-aperture stitching interferometry (FSSI) is proposed for aspheric and free-form surfaces testing. Different from normal annular and circular sub-aperture stitching method, non-null interferometric configuration and irregular sub-apertures are employed to make the best use of the resolved fringes. Meanwhile, a multi-aperture simultaneous reverse optimizing reconstruction (MSROR) process is proposed for full aperture figure error reconstruction. With the multi-configuration model, full aperture figure error would be extracted in form of Zernike polynomials from sub-apertures wavefront data by the MSROR method. This method concurrently accomplishes sub-aperture retrace error and misalignment correction, requiring neither complex mathematical algorithms, nor sub-aperture overlaps. The experiments show that, the aspheric annular sub-aperture stitching accuracy with the proposed FSSI is more than 1/50 lambda (rms) comparing with the result of ZYGO Verifier Asphere interferometer. The bi-conic surface sector sub-aperture stitching accuracy is about 1/50 lambda (rms) comparing with result of Taylor Hobson contourgraph in longitudinal section contour.
引用
收藏
页数:12
相关论文
共 46 条
  • [1] Non-null testing for aspheric surfaces using elliptical sub-aperture stitching technique
    Zhao, Zixin
    Zhao, Hong
    Gu, Feifei
    Du, Hubing
    Li, Kaixing
    OPTICS EXPRESS, 2014, 22 (05): : 5512 - 5521
  • [2] Research on non-null convex aspherical sub-aperture stitching detection technology
    Yan Gong-jing
    Zhang Xian-zhong
    CHINESE OPTICS, 2018, 11 (05): : 798 - 803
  • [3] Study of annular sub-aperture stitching interferometry for aspheric surfaces
    Zhao, Zixin
    Xiao, Zhaoxian
    Zhang, Hangying
    APPLIED OPTICAL METROLOGY II, 2017, 10373
  • [4] Calibration of a non-null test interferometer for the measurement of aspheres and free-form surfaces
    Baer, Goran
    Schindler, Johannes
    Pruss, Christof
    Siepmann, Jens
    Osten, Wolfgang
    OPTICS EXPRESS, 2014, 22 (25): : 31200 - 31211
  • [5] Sub-aperture stitching of aspheric surfaces in precision in-situ measurement
    Pan, Jintao
    Zhang, Xiangchao
    Xu, Min
    AOPC 2017: 3D MEASUREMENT TECHNOLOGY FOR INTELLIGENT MANUFACTURING, 2017, 10458
  • [6] Non-null annular subaperture stitching interferometry for aspheric test
    Zhang, Lei
    Liu, Dong
    Shi, Tu
    Yang, Yongying
    Chong, Shiyao
    Miao, Liang
    Huang, Wei
    Shen, Yibing
    Bai, Jian
    OPTIFAB 2015, 2015, 9633
  • [7] Non-null stitching test convex aspheric metal mirror
    Cai Z.
    Wang X.
    Hu H.
    Zeng X.
    Wang J.
    Hongwai yu Jiguang Gongcheng/Infrared and Laser Engineering, 2021, 50 (11):
  • [8] Stitching of off-axis sub-aperture null measurements of an aspheric surface
    Zhao, Chunyu
    Burge, James H.
    INTERFEROMETRY XIV: TECHNIQUES AND ANALYSIS, 2008, 7063
  • [9] Model analysis of sub-aperture stitching interferometry for aspheric surface test
    Qiao, Yujing
    Tan, Jiubin
    Wang, Weibo
    Yi Qi Yi Biao Xue Bao/Chinese Journal of Scientific Instrument, 2007, 28 (10): : 1880 - 1885
  • [10] Correction of misalignment introduced aberration in non-null test measurements of free-form surfaces
    Baer, G.
    Schindler, J.
    Pruss, C.
    Osten, W.
    JOURNAL OF THE EUROPEAN OPTICAL SOCIETY-RAPID PUBLICATIONS, 2013, 8