Clinical Results and Pump Analysis of the Gyro Pump for Long-Term Extracorporeal Life Support

被引:1
|
作者
Terasaki, Takamitsu [1 ]
Takano, Tamaki [1 ]
Michinaga, Yuuki [1 ]
Yokokawa, Michihiro [1 ]
Wada, Yuko [1 ]
Seto, Tatsuichirou [1 ]
Fukui, Daisuke [1 ]
Amano, Jun [1 ]
机构
[1] Shinshu Univ, Dept Cardiovasc Surg, Sch Med, Matsumoto, Nagano 3908621, Japan
关键词
Gyro pump; Extracorporeal life support; Long term; PERCUTANEOUS CARDIOPULMONARY SUPPORT; CENTRIFUGAL PUMP; VENTRICULAR SUPPORT; C1E3; BYPASS; OXYGENATOR; EXPERIENCE; MEMBRANE; IMPELLER; ASSIST;
D O I
10.1111/aor.12171
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Rescuing patients in severe cardiac failure with extracorporeal support remains challenging. The Gyro pump is a centrifugal blood pump and was now used for cardiopulmonary bypass, although it was originally developed for long-term cardiac assist. Little is known about clinical experiences using this pump. Here, we report on the clinical results of long-term extracorporeal life support for over 4 days using the Gyro pump with Excelung, a hollow fiber oxygenator coated with silicone and heparin. Seven patients underwent extracorporeal life support with 15 pump and oxygenator combinations. Gyro and Excelung were used for venoarterial extracorporeal support in six patients and for right ventricular support in one patient. Patient characteristics, pump driving conditions, and blood chemistry were obtained retrospectively. All pumps were subsequently disassembled and examined macroscopically, with 6 of 15 pumps also examined by scanning electron microscopy (SEM). The patient mortality rate was 57.1%. Mean duration of support was 10.5 +/- 7.2 days per pump and oxygenator combination. Lactate dehydrogenase and aspartate aminotransferase were generally maintained below 1000 and 100IU/L, respectively, after the first 4 days of pump driving. Thrombi were found in two pumps, one used without anticoagulation and the other driven at a very slow rotational speed. SEM revealed no wear in the male bearings and very low wear and deformation (0.02 +/- 0.03mm) in the female bearings. The combination of Gyro and Excelung may be applicable for long-term biventricular and right ventricular support, although proper anticoagulation should be administrated to avoid thrombus formation inside the pump.
引用
收藏
页码:810 / 815
页数:6
相关论文
共 50 条
  • [21] Pediatric Extracorporeal Life Support Using a Third Generation Diagonal Pump
    Speth, Marlene
    Muench, Frank
    Purbojo, Ariawan
    Gloeckler, Martin
    Toka, Okan
    Cesnjevar, Robert A.
    Rueffer, Andre
    ASAIO JOURNAL, 2016, 62 (04) : 482 - 490
  • [22] A new pneumatic pump for extracorporeal circulation: TPP (True Pulsatile Pump). Experimental and first clinical results
    Waldenberger, FR
    Vandezande, E
    Janssens, P
    Morishige, N
    Demeyere, R
    deRuyter, E
    Wiebalck, A
    Flameng, W
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1997, 20 (08): : 447 - 454
  • [23] Long-term outcomes of patients bridged to recovery with venoarterial extracorporeal life support
    Dardik, Gabriel
    Ning, Yuming
    Kurlansky, Paul
    Cruz, Guillermo Almodovar
    Vinogradsky, Alice
    Fried, Justin
    Topkara, Veli K.
    Takeda, Koji
    PERFUSION-UK, 2024, 39 (08): : 1629 - 1635
  • [24] Long-term survival after extracorporeal life support in children with neutropenic sepsis
    Sile Smith
    Warwick Butt
    Derek Best
    Graeme MacLaren
    Intensive Care Medicine, 2016, 42 : 942 - 943
  • [25] Long-term survival after extracorporeal life support in children with neutropenic sepsis
    Smith, Sile
    Butt, Warwick
    Best, Derek
    MacLaren, Graeme
    INTENSIVE CARE MEDICINE, 2016, 42 (05) : 942 - 943
  • [26] Long-Term Outcomes After Extracorporeal Life Support: What Happens to the Survivors?
    MacLaren, Graeme
    PEDIATRIC CRITICAL CARE MEDICINE, 2013, 14 (06) : 646 - 647
  • [27] Long-Term Outcomes After Extracorporeal Life Support for Acute Respiratory Failure
    McDonald, Michael D.
    Sandsmark, Danielle K.
    Palakshappa, Jessica A.
    Mikkelsen, Mark E.
    Anderson, Brian J.
    Gutsche, Jacob T.
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2019, 33 (01) : 72 - 79
  • [28] The Effect of Sedation on Long-Term Psychological Impairment After Extracorporeal Life Support
    McDonald, Michael D.
    Lane-Fall, Meghan
    Miano, Todd A.
    Henry, Madeline
    Gallagher, Colby
    Hadler, Rachel
    Laudanski, Krzysztof
    Mackay, Emily J.
    Usman, Asad A.
    Gutsche, Jacob
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2020, 34 (03) : 663 - 667
  • [29] LONG-TERM CLINICAL-EVALUATION OF A NEW MINIATURE INSULIN INFUSION PUMP
    BENDING, J
    PICKUP, J
    KEEN, H
    ROTHWELL, D
    SUTHERLAND, I
    DIABETES, 1983, 32 : A81 - A81
  • [30] Preclinical study of a novel hydrodynamically levitated centrifugal pump for long-term cardiopulmonary support
    Tsukiya, Tomonori
    Mizuno, Toshihide
    Takewa, Yoshiaki
    Tatsumi, Eisuke
    Taenaka, Yoshiyuki
    JOURNAL OF ARTIFICIAL ORGANS, 2015, 18 (04) : 300 - 306