Clinical Results and Pump Analysis of the Gyro Pump for Long-Term Extracorporeal Life Support

被引:1
|
作者
Terasaki, Takamitsu [1 ]
Takano, Tamaki [1 ]
Michinaga, Yuuki [1 ]
Yokokawa, Michihiro [1 ]
Wada, Yuko [1 ]
Seto, Tatsuichirou [1 ]
Fukui, Daisuke [1 ]
Amano, Jun [1 ]
机构
[1] Shinshu Univ, Dept Cardiovasc Surg, Sch Med, Matsumoto, Nagano 3908621, Japan
关键词
Gyro pump; Extracorporeal life support; Long term; PERCUTANEOUS CARDIOPULMONARY SUPPORT; CENTRIFUGAL PUMP; VENTRICULAR SUPPORT; C1E3; BYPASS; OXYGENATOR; EXPERIENCE; MEMBRANE; IMPELLER; ASSIST;
D O I
10.1111/aor.12171
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Rescuing patients in severe cardiac failure with extracorporeal support remains challenging. The Gyro pump is a centrifugal blood pump and was now used for cardiopulmonary bypass, although it was originally developed for long-term cardiac assist. Little is known about clinical experiences using this pump. Here, we report on the clinical results of long-term extracorporeal life support for over 4 days using the Gyro pump with Excelung, a hollow fiber oxygenator coated with silicone and heparin. Seven patients underwent extracorporeal life support with 15 pump and oxygenator combinations. Gyro and Excelung were used for venoarterial extracorporeal support in six patients and for right ventricular support in one patient. Patient characteristics, pump driving conditions, and blood chemistry were obtained retrospectively. All pumps were subsequently disassembled and examined macroscopically, with 6 of 15 pumps also examined by scanning electron microscopy (SEM). The patient mortality rate was 57.1%. Mean duration of support was 10.5 +/- 7.2 days per pump and oxygenator combination. Lactate dehydrogenase and aspartate aminotransferase were generally maintained below 1000 and 100IU/L, respectively, after the first 4 days of pump driving. Thrombi were found in two pumps, one used without anticoagulation and the other driven at a very slow rotational speed. SEM revealed no wear in the male bearings and very low wear and deformation (0.02 +/- 0.03mm) in the female bearings. The combination of Gyro and Excelung may be applicable for long-term biventricular and right ventricular support, although proper anticoagulation should be administrated to avoid thrombus formation inside the pump.
引用
收藏
页码:810 / 815
页数:6
相关论文
共 50 条
  • [1] A SEAL-LESS CENTRIFUGAL PUMP (BAYLOR GYRO PUMP) FOR APPLICATION TO LONG-TERM CIRCULATORY SUPPORT
    MINATO, N
    SAKUMA, I
    SASAKI, T
    SHIONO, M
    OHARA, Y
    TAKATANI, S
    NOON, GP
    NOSE, Y
    ARTIFICIAL ORGANS, 1993, 17 (01) : 36 - 42
  • [2] BAYLOR GYRO PUMP - A COMPLETELY SEAL-LESS CENTRIFUGAL PUMP AIMING FOR LONG-TERM CIRCULATORY SUPPORT
    OHARA, Y
    SAKUMA, I
    MAKINOUCHI, K
    DAMM, G
    GLUECK, J
    MIZUGUCHI, K
    NAITO, K
    TASAI, K
    ORIME, Y
    TAKATANI, S
    NOON, GP
    NOSE, Y
    ARTIFICIAL ORGANS, 1993, 17 (07) : 599 - 604
  • [3] LONG-TERM CLINICAL USE OF EXTRACORPOREAL CARDIOPULMONARY ASSIST DEVICE WITH THE GYRO C1E3 CENTRIFUGAL BLOOD PUMP
    Tabayashi, N.
    Kameda, Y.
    Yoshikawa, Y.
    Abe, T.
    Naito, H.
    Hayata, Y.
    Hirose, T.
    Yokota, M.
    Iwashita, Y.
    Taniguchi, S.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2008, 31 (07): : 620 - 620
  • [4] A NEW HEART PUMP FOR LONG-TERM CIRCULATORY SUPPORT
    NUNN, DB
    BARILA, TG
    FORSEE, JH
    WOODWARD, KE
    MON, G
    STRAUB, H
    ARCHIVES OF SURGERY, 1963, 86 (05) : 797 - 803
  • [5] Preliminary In Vivo Testing of a Novel Pump for Short-Term Extracorporeal Life Support
    Spurlock, David J.
    Koch, Kelly
    Mazur, Daniel E.
    Fracz, Emilia M.
    Bartlett, Robert H.
    Haft, Jonathan W.
    ANNALS OF THORACIC SURGERY, 2012, 93 (01): : 141 - 147
  • [6] In Vivo Testing of a Novel Blood Pump for Short-Term Extracorporeal Life Support
    Teman, N. R.
    Demos, D. S.
    Bryner, B. S.
    Faliks, B.
    Fracz, E. M.
    Mazur, D. E.
    Rojas-Pena, A.
    Bartlett, R. H.
    Haft, J. W.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2013, 32 (04): : S148 - S149
  • [7] In Vivo Testing of a Novel Blood Pump for Short-Term Extracorporeal Life Support
    Teman, Nicholas R.
    Demos, David S.
    Bryner, Benjamin S.
    Faliks, Bradley
    Jahangir, Emilia M.
    Mazur, Daniel E.
    Rojas-Pena, Alvaro
    Bartlett, Robert H.
    Haft, Jonathan W.
    ANNALS OF THORACIC SURGERY, 2014, 98 (01): : 97 - 102
  • [8] LONG-TERM RESULTS WITH THE WHITE GLAUCOMA PUMP-SHUNT
    DAVIDOVSKI, F
    STEWART, RH
    KIMBROUGH, RL
    OPHTHALMIC SURGERY AND LASERS, 1990, 21 (04): : 288 - 293
  • [9] Long-Term Outcomes of Extracorporeal Life Support in Respiratory Failure
    Bursa, Filip
    Frelich, Michal
    Sklienka, Peter
    Jor, Ondrej
    Maca, Jan
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (16)
  • [10] Long-term biventricular support with the heartware implantable continuous flow pump
    Hetzer, Roland
    Krabatsch, Thomas
    Stepanenko, Alexander
    Hennig, Ewald
    Potapov, Evgenij V.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2010, 29 (07): : 822 - 824