The self expandable Nitinol stents or stentgrafts are typically used for miniinvasive treatment of stenosis and aneurysms in the cardiovascular system. The minimal traumatisation of the patient, shorter time of hospitalization are typical advantages of these methods. More than ten years of experience has yielded also important information about the performance of stents in interaction with biological system and the possible problems related with it. The leakage or the shift of stent are some typical disadvantages, that can be related among other in the construction of the stent. The problem is that the mechanical properties, dimensions and the dynamical properties of the stent do not exactly correspond to the properties of the vessel or generally of tissue where this stent is introduced. The measurement, the description and the comparison of the relations between the mechanical properties of stents and tissues can be one of the possible ways to minimize these disadvantages. The developed original computer controlled measuring system allows the measurement of mechanical properties of stents, the measurement of strain-stress curves or simulation of interaction of the stent and vessel for exactly defined hemodynamic conditions. We measured and compared the mechanical parameters of different selfexpandable Nitinol stents, which differed in geometry ( radius and length), in the type of construction ( number of branches and rising of winding) and in the diameter of used wire. The results of measurements confirmed the theoretical assumptions that just the diameter of the Nitinol wire significantly influences the rigidity and the level of compressibility of the stent as well. Acompromise must be found between the required rigidity of the stent and the minimal size of the delivery system. The exact description of the relation between the mechanical properties and geometry and construction of the stents enables to design the stent to fit the patient and it is expected that this access improves the efficiency of treatment. The results of measurement are also necessary for the design and identification of the parameters of the models of the stents.