Performance evaluation of a CyberKnife® G4 image-guided robotic stereotactic radiosurgery system

被引:98
|
作者
Antypas, Christos [1 ,2 ]
Pantelis, Evaggelos [1 ]
机构
[1] Clin & Diagnost Ctr, Dept Med Phys, Athens 15231, Greece
[2] Univ Athens, Aretaieion Hosp, Dept Radiol, Athens 11528, Greece
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2008年 / 53卷 / 17期
关键词
D O I
10.1088/0031-9155/53/17/016
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of the current work was to present the performance evaluation procedures implemented at our department for the commissioning of a G4 CyberKnife system. This system consists of a robotic manipulator, a target-locating system and a lightweight 6-MV linac. Individual quality assurance procedures were performed for each of the CyberKnife subsystems. The system was checked for the mechanical accuracy of its robotic manipulator. The performance of the target-locating system was evaluated in terms of mechanical accuracy of both cameras' alignment and quality assurance tests of the x-ray generators and the flat-panel detectors. The traditional linac 6-MV beam characteristics and beam output parameters were also measured. Results revealed a manipulator mechanical mean accuracy of similar to 0.1 mm, with individual maximum position uncertainties less than 0.25 mm. The target-locating system mechanical accuracy was found within the acceptance limits. For the most clinically used parameters in the CyberKnife practice, e. g. 100-120 kV and 50-200 ms, kV and exposure time accuracy error were measured as less than 2%, while the precision error of the kV was determined as less than 1%. The acquired images of the ETR grid pattern revealed no geometrical distortion while the critical frequency f(50) values for cameras A and B were calculated as 1.5 lp mm(-1) and 1.4 lp mm(-1), respectively. Dose placement measurements were performed in a head and neck phantom. Results revealed sub-millimeter beam delivery precision whereas the total clinical accuracy of the system was measured equal to 0.44 +/- 0.12 mm, 0.29 +/- 0.10 mm and 0.53 +/- 0.16 mm for the skull, fiducial and Xsight spine tracking methods, respectively. The results of this work certify the G4 CyberKnife SRS system capable of delivering high dose distributions with sub-millimeter accuracy and precision to intracranial and extracranial lesions. Moreover, total clinical accuracy of the investigated G4 system was found to be improved for the skull and fiducial tracking methods
引用
收藏
页码:4697 / 4718
页数:22
相关论文
共 50 条
  • [21] CLINICAL EFFECTIVENESS OF ROBOTIC IMAGE-GUIDED STEREOTACTIC RADIOSURGERY (CYBERKNIFE) IN SELECTED PRIMARY AND SECONDARY SOFT TISSUE NEOPLASMS: A SYSTEMATIC REVIEW
    Wlodarczyk, A.
    Ceglowska, U.
    Slomka, M.
    Luchowska, K.
    VALUE IN HEALTH, 2014, 17 (07) : A616 - A616
  • [22] A Second Course of Stereotactic Image-Guided Robotic Radiosurgery for Patients with Cerebral Metastasis
    Jiang, Xuechao
    Wang, Hui
    Song, Yongchun
    Wang, Xiaoguang
    Li, Fengtong
    Dong, Yang
    Wang, Jingsheng
    Chen, Huaming
    Yuan, Zhiyong
    WORLD NEUROSURGERY, 2019, 123 : E621 - E628
  • [23] Image-guided Robotic Stereotactic Radiosurgery for Unresectable Liver Metastases: Preliminary Results
    Ambrosino, Giovanni
    Polistina, Francesco
    Costantin, Giorgio
    Francescon, Paolo
    Guglielmi, Rosabianca
    Zanco, Pierluigi
    Casamassima, Franco
    Febbraro, Antonio
    Gerunda, Giorgio
    Lumachi, Franco
    ANTICANCER RESEARCH, 2009, 29 (08) : 3381 - 3384
  • [24] Fiducial migration following small peripheral lung tumor image-guided CyberKnife stereotactic radiosurgery
    Strulik, Konrad L.
    Cho, Min H.
    Collins, Brian T.
    Khan, Noureen
    Banovac, Filip
    Slack, Rebecca
    Cleary, Kevin
    MEDICAL IMAGING 2008: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND MODELING, PTS 1 AND 2, 2008, 6918
  • [25] CyberKnife robotic image-guided stereotactic radiotherapy for oligometastic cancer A prospective evaluation of 95 patients/118 lesions
    Jereczek-Fossa, B. A.
    Bossi-Zanetti, I.
    Mauro, R.
    Beltramo, G.
    Fariselli, L.
    Bianchi, L. C.
    Fodor, C.
    Fossati, P.
    Baroni, G.
    Orecchia, R.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2013, 189 (06) : 448 - 455
  • [26] Relativistic protons for image-guided stereotactic radiosurgery
    Durante, M.
    Stoecker, H.
    1ST NANO-IBCT CONFERENCE 2011 - RADIATION DAMAGE OF BIOMOLECULAR SYSTEMS: NANOSCALE INSIGHTS INTO ION BEAM CANCER THERAPY, 2012, 373
  • [27] Clinical experience with image-guided robotic radiosurgery (the Cyberknife) in the treatment of brain and spinal cord tumors
    Chang, SD
    Murphy, M
    Geis, P
    Martin, DP
    Hancock, SL
    Doty, JR
    Adler, JR
    NEUROLOGIA MEDICO-CHIRURGICA, 1998, 38 (11): : 780 - 783
  • [28] Image-Guided Robotic Radiosurgery for Trigeminal Neuralgia
    Romanelli, Pantaleo
    Conti, Alfredo
    Bianchi, Livia
    Bergantin, Achille
    Martinotti, Anna
    Beltramo, Giancarlo
    NEUROSURGERY, 2018, 83 (05) : 1023 - 1030
  • [29] Image-guided robotic radiosurgery for spinal metastases
    Gibbs, Iris C.
    Kamnerdsupaphon, Pimkhuan
    Ryu, Mi-Ryeong
    Dodd, Robert
    Kiernan, Michaela
    Change, Steven D.
    Adler, John R., Jr.
    RADIOTHERAPY AND ONCOLOGY, 2007, 82 (02) : 185 - 190
  • [30] Single fraction image-guided robotic stereotactic radiosurgery for primary and secondary lung cancer
    Senger, C.
    Pasemann, D.
    Kluge, A.
    Acker, G.
    Kalinauskaite, G.
    Kreimeier, A.
    Budach, V
    Stromberger, C.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2019, 195 : S140 - S140