Performance evaluation of a CyberKnife® G4 image-guided robotic stereotactic radiosurgery system

被引:98
|
作者
Antypas, Christos [1 ,2 ]
Pantelis, Evaggelos [1 ]
机构
[1] Clin & Diagnost Ctr, Dept Med Phys, Athens 15231, Greece
[2] Univ Athens, Aretaieion Hosp, Dept Radiol, Athens 11528, Greece
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2008年 / 53卷 / 17期
关键词
D O I
10.1088/0031-9155/53/17/016
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of the current work was to present the performance evaluation procedures implemented at our department for the commissioning of a G4 CyberKnife system. This system consists of a robotic manipulator, a target-locating system and a lightweight 6-MV linac. Individual quality assurance procedures were performed for each of the CyberKnife subsystems. The system was checked for the mechanical accuracy of its robotic manipulator. The performance of the target-locating system was evaluated in terms of mechanical accuracy of both cameras' alignment and quality assurance tests of the x-ray generators and the flat-panel detectors. The traditional linac 6-MV beam characteristics and beam output parameters were also measured. Results revealed a manipulator mechanical mean accuracy of similar to 0.1 mm, with individual maximum position uncertainties less than 0.25 mm. The target-locating system mechanical accuracy was found within the acceptance limits. For the most clinically used parameters in the CyberKnife practice, e. g. 100-120 kV and 50-200 ms, kV and exposure time accuracy error were measured as less than 2%, while the precision error of the kV was determined as less than 1%. The acquired images of the ETR grid pattern revealed no geometrical distortion while the critical frequency f(50) values for cameras A and B were calculated as 1.5 lp mm(-1) and 1.4 lp mm(-1), respectively. Dose placement measurements were performed in a head and neck phantom. Results revealed sub-millimeter beam delivery precision whereas the total clinical accuracy of the system was measured equal to 0.44 +/- 0.12 mm, 0.29 +/- 0.10 mm and 0.53 +/- 0.16 mm for the skull, fiducial and Xsight spine tracking methods, respectively. The results of this work certify the G4 CyberKnife SRS system capable of delivering high dose distributions with sub-millimeter accuracy and precision to intracranial and extracranial lesions. Moreover, total clinical accuracy of the investigated G4 system was found to be improved for the skull and fiducial tracking methods
引用
收藏
页码:4697 / 4718
页数:22
相关论文
共 50 条
  • [1] Image guidance quality assurance of a G4 CyberKnife robotic stereotactic radiosurgery system
    Pantelis, E.
    Petrokokkinos, L.
    Antypas, C.
    JOURNAL OF INSTRUMENTATION, 2009, 4
  • [2] Image-guided stereotactic radiosurgery with the CyberKnife for pituitary adenomas
    Kajiwara, K
    Saito, K
    Yoshikawa, K
    Kato, S
    Akimura, T
    Nomura, S
    Ishihara, H
    Suzuki, M
    MINIMALLY INVASIVE NEUROSURGERY, 2005, 48 (02) : 91 - 96
  • [3] Lung Metastases treated by CyberKnife® image-guided robotic stereotactic radiosurgery at 4.1 months
    Brown, William T.
    Wu, Xiaodong
    Fowler, John F.
    Garcia, Silvio
    Fayad, Fahed
    Amendola, Beatriz E.
    de la Zerda, Alberto
    Schwade, James G.
    SOUTHERN MEDICAL JOURNAL, 2008, 101 (04) : 376 - 382
  • [4] Early results of CyberKnife image-guided robotic stereotactic radiosurgery for treatment of lung tumors
    Brown, William T.
    Wu, Xiaodong
    Wen, B-Chen
    Fowler, John F.
    Fayad, Fahed
    Amendola, Beatriz E.
    Garcia, Silvio
    De la Zerda, Alberto
    Huang, Zhicong
    Schwade, James G.
    COMPUTER AIDED SURGERY, 2007, 12 (05) : 253 - 261
  • [5] Dose linearity and monitor unit stability of a G4 type cyberknife robotic stereotactic radiosurgery system
    Sudahar, H.
    Kurup, P. G. G.
    Murali, V.
    Velmurugan, J.
    JOURNAL OF MEDICAL PHYSICS, 2012, 37 (01) : 4 - 7
  • [6] Preliminary clinical experience with the cyberknife: Image-guided stereotactic radiosurgery
    Adler, JR
    Cox, RS
    RADIOSURGERY 1995, 1996, 1 : 316 - 326
  • [7] Image-guided robotic radiosurgery: Clinical and radiographic results with the CyberKnife
    Chang, SD
    Murphy, MJ
    Martin, DP
    Hancock, SL
    Doty, JR
    Adler, JR
    RADIOSURGERY 1999, 2000, 3 : 23 - 33
  • [8] CyberKnife image-guided stereotactic radiosurgery in the treatment of benign spinal tumors
    Dodd, RL
    Gibbs, I
    Adler, J
    Chang, SD
    JOURNAL OF NEUROSURGERY, 2004, 100 (04) : 787 - 788
  • [9] CYBERKNIFE IMAGE-GUIDED RADIOSURGERY FOREWORD
    Adler, John R., Jr.
    Chang, Steven D.
    NEUROSURGERY, 2009, 64 (02) : A1 - A1
  • [10] Evaluation of CyberKnife frameless real-time image-guided stereotactic radiosurgery for spinal lesions
    Gerszten, PC
    Ozhasoglu, C
    Burton, SA
    Vogel, W
    Atkins, B
    Kalnicki, S
    Welch, WC
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2003, 81 (1-4) : 84 - 89