Numerical study of the influence of water evaporation on radiofrequency ablation

被引:27
|
作者
Zhu, Qing [1 ]
Shen, Yuanyuan [2 ]
Zhang, Aili [1 ]
Xu, Lisa X. [1 ]
机构
[1] Shanghai Jiao Tong Univ, MedX Res Inst, Sch Biomed Engn, State Key Lab Oncogenes & Related Genes, Shanghai 200030, Peoples R China
[2] Shenzhen Univ, Sch Med, Dept Biomed Engn, Natl Reg Key Technol Engn Lab Med Ultrasound, Shenzhen 518060, Peoples R China
基金
中国国家自然科学基金;
关键词
Radiofrequency ablation; Water evaporation; Water diffusion; Mathematical model; TEMPERATURE DISTRIBUTION; ELECTRICAL-CONDUCTIVITY; DIELECTRIC-PROPERTIES; THERMAL-CONDUCTIVITY; TISSUE TEMPERATURE; BIOLOGICAL TISSUES; HEATING PATTERNS; SELF-DIFFUSION; RF ABLATION; TUMOR;
D O I
10.1186/1475-925X-12-127
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Radiofrequency ablation is a promising minimal invasive treatment for tumor. However, water loss due to evaporation has been a major issue blocking further RF energy transmission and correspondently eliminating the therapeutic outcome of the treatment. Method: A 2D symmetric cylindrical mathematical model coupling the transport of the electrical current, heat, and the evaporation process in the tissue, has been developed to simulate the treatment process and investigate the influence of the excessive evaporation of the water on the treatment. Results: Our results show that the largest specific absorption rate (QSAR) occurs at the edge of the circular surface of the electrode. When excessive evaporation takes place, the water dehydration rate in this region is the highest, and after a certain time, the dehydrated tissue blocks the electrical energy transmission in the radial direction. It is found that there is an interval as long as 65 s between the beginning of the evaporation and the increase of the tissue impedance. The model is further used to investigate whether purposely terminating the treatment for a while allowing diffusion of the liquid water into the evaporated region would help. Results show it has no obvious improvement enlarging the treatment volume. Treatment with the cooled-tip electrode is also studied. It is found that the cooling conditions of the inside agent greatly affect the water loss pattern. When the convection coefficient of the cooling agent increases, excessive evaporation will start from near the central axis of the tissue cylinder instead of the edge of the electrode, and the coagulation volume obviously enlarges before a sudden increase of the impedance. It is also found that a higher convection coefficient will extend the treatment time. Though the sudden increase of the tissue impedance could be delayed by a larger convection coefficient; the rate of the impedance increase is also more dramatic compared to the case with smaller convection coefficient. Conclusion: The mathematical model simulates the water evaporation and diffusion during radiofrequency ablation and may be used for better clinical design of radiofrequency equipment and treatment protocol planning.
引用
收藏
页数:16
相关论文
共 50 条
  • [11] Pathobiological Influence of a Radiofrequency Ablation System
    Inoue, Keiji
    Karashima, Takashi
    Iiyama, Tatsuo
    Ashida, Shingo
    Kamada, Masayuki
    Shuin, Taro
    Rao, Parijatha
    Kurabayashi, Atsushi
    Furihata, Mutsuo
    Hotta, Akimitsu
    PATHOBIOLOGY, 2010, 77 (02) : 64 - 77
  • [12] Numerical study of heat transfer in bone tumor during radiofrequency ablation therapy
    Nie, Xiaohui
    Nan, Qun
    Zhai, Fei
    Guo, Xuemei
    INFORMATION TECHNOLOGY, 2015, : 269 - 271
  • [13] Laser and Radiofrequency Ablation Study (LARA study): A Randomised Study Comparing Radiofrequency Ablation and Endovenous Laser Ablation (810 nm)
    Goode, S. D.
    Chowdhury, A.
    Crockett, M.
    Beech, A.
    Simpson, R.
    Richards, T.
    Braithwaite, B. D.
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2010, 40 (02) : 246 - 253
  • [14] Experimental Study of the Influence of Water Temperature on Pan Evaporation
    Gao H.
    Chen Z.
    Shi Z.
    Yan C.
    Wang B.
    Zou Z.
    Qiu G.
    Beijing Daxue Xuebao (Ziran Kexue Ban)/Acta Scientiarum Naturalium Universitatis Pekinensis, 2022, 58 (01): : 147 - 156
  • [15] NUMERICAL STUDY ON EVOLUTION OF SESSILE WATER DROPLETS DURING EVAPORATION
    Hu, Dinghua
    Wu, Huiying
    Wu, Xinyu
    PROCEEDINGS OF THE ASME INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION - 2012, VOL 7, PTS A-D, 2013, : 2743 - 2749
  • [16] Simulation study on the influence of an electric field on water evaporation
    Okuno, Yoshishige
    Minagawa, Mie
    Matsumoto, Hidetoshi
    Tanioka, Akihiko
    JOURNAL OF MOLECULAR STRUCTURE-THEOCHEM, 2009, 904 (1-3): : 83 - 90
  • [17] Numerical and Experimental Characterization of Radiofrequency Ablation in Perfused Kidneys
    Frank, Kristian
    Lindenborn, Herbert
    Dahlhaus, Dirk
    2012 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2012, : 5707 - 5711
  • [18] Influence of radiofrequency catheter ablation on platelet aggregation
    Kozlovaite, Vilma
    Grybauskas, Pranas
    Cimbolaityte, Jurate
    Mongirdiene, Ausra
    Puodziukynas, Aras
    Kazakevicius, Tomas
    Ptasekas, Julius
    CLINICAL AND APPLIED THROMBOSIS-HEMOSTASIS, 2007, 13 (04) : 416 - 421
  • [19] RADIOFREQUENCY ABLATION - INFLUENCE OF MYOCARDIAL TISSUE CHARACTERISTICS
    SAXON, LA
    STEVENSON, WG
    MIDDLEKAUFF, HR
    NATTERSON, PD
    FONAROW, GC
    CHILD, JS
    CLINICAL RESEARCH, 1993, 41 (01): : A105 - A105
  • [20] Influence of radiofrequency catheter ablation on myocardial metabolism
    Lin, JM
    Li, YH
    Lin, JL
    Tseng, YZ
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 1996, 56 (01) : 29 - 34