Theoretical Framework for the Dual-Energy Cone-Beam CT Noise-Power Spectrum, NEQ, and Tasked-Based Detectability Index

被引:0
|
作者
Gang, G. J. [1 ,2 ]
Zbijewski, W. [2 ]
Stayman, J. W. [2 ]
Carrino, J. A. [3 ]
Siewerdsen, J. H. [1 ,3 ]
机构
[1] Univ Toronto, Inst Biomat & Biomed Engn, Toronto, ON M5G 2M9, Canada
[2] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21205 USA
[3] Johns Hopkins Univ, Russell H Morgan Dept Radiol, Baltimore, MD 21218 USA
关键词
dual-energy CT; cone-beam CT; cascaded systems analysis; noise-power spectrum; noise-equivalent quanta; detectability index; PERFORMANCE; OPTIMIZATION;
D O I
10.1117/12.911817
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The optimization of dual-energy computed tomography (DE-CT) is challenged by the lack of a theoretical foundation for image quality. This work reports a cascaded systems analysis model that was used to derive signal and noise propagation in DE-CBCT in prevalent Fourier metrics such as the noise-power spectrum (NPS) and noise-equivalent quanta (NEQ). The model was validated in comparison to measurements of the 3D NPS and NEQ in DE-CBCT images acquired using an experimental imaging bench. Task-based detectability index was derived using DE-NPS and NEQ as an objective function in optimizing DE imaging parameters such as the dose allocation factor (D-A) and kVp pair. The resulting dose allocation optimization is in agreement with the practice of assigning more dose to the high-energy image (D-A < 0.5), and the model provides a quantitative basis for examining the optimal dose allocation as a function of total dose, kVp pair, the presence of electronics noise, and the imaging task. An example optimization is shown for a breast tumor detection task. Using DE decomposition to cancel fibroglandular tissue (rendering a DE-CBCT image of breast tumor against an adipose tissue background) and assuming a total dose of 15mGy, the optimal kVp pair is identified at [45, 105] kVp with D-A=0.46. The model is sufficiently general for applications beyond this example, demonstrating utility in the optimization in a broad range of imaging parameters. The model provides a new, valuable framework for understanding the theoretical limits of DE-CBCT imaging performance and maximizing image quality while minimizing radiation dose.
引用
收藏
页数:8
相关论文
共 46 条
  • [21] Optimizing Proton Stopping Power Ratio Prediction with Dual-Energy Cone-Beam CT Using the Cramer-Rao Lower Bound
    Leibold, David
    Schaart, Dennis R.
    Goorden, Marlies C.
    [J]. MEDICAL IMAGING 2024: PHYSICS OF MEDICAL IMAGING, PT 1, 2024, 12925
  • [22] Low-dose dual-energy cone-beam CT using a total-variation minimization algorithm
    Min, Jonghwan
    Lee, Taewon
    Kim, Kyungwoo
    Cho, Gyuseong
    Cho, Seungryong
    [J]. MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [23] Evaluation of Imaging Dose for Modulation Layer Based Dual Energy Cone-Beam CT
    Ju, Eunbin
    Ahn, SoHyun
    Cho, Samju
    Keum, Ki Chang
    Lee, Rena
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3387 - 3387
  • [24] Feasibility study of three-material decomposition in dual-energy cone-beam CT imaging with deep learning
    Zhu, Jiongtao
    Su, Ting
    Zhang, Xin
    Yang, Jiecheng
    Mi, Donghua
    Zhang, Yunxin
    Gao, Xiang
    Zheng, Hairong
    Liang, Dong
    Ge, Yongshuai
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (14):
  • [25] Dual-energy cone-beam CT with a flat-panel detector: Effect of reconstruction algorithm on material classification
    Zbijewski, W.
    Gang, G. J.
    Xu, J.
    Wang, A. S.
    Stayman, J. W.
    Taguchi, K.
    Carrino, J. A.
    Siewerdsen, J. H.
    [J]. MEDICAL PHYSICS, 2014, 41 (02)
  • [26] Feasibility of Dual-Energy Cone-Beam CT of Bone Marrow Edema Using Dual-Layer Flat Panel Detectors
    Liu, Stephen Z.
    Zhao, Chumin
    Herbst, Magdalena
    Weber, Thomas
    Vogt, Sebastian
    Ritschl, Ludwig
    Kappler, Steffen
    Siewerdsen, Jeffrey H.
    Zbijewski, Wojciech
    [J]. MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING, 2022, 12031
  • [27] Local and global 3D noise power spectrum in cone-beam CT system with FDK reconstruction
    Baek, Jongduk
    Pelc, Norbert J.
    [J]. MEDICAL PHYSICS, 2011, 38 (04) : 2122 - 2131
  • [28] Are Spectral Modulator and Flying Focal Spot Beneficial for Fast-KV Switching Dual-Energy Cone-Beam CT ?
    Zhou, Hao
    Deng, Yifan
    Gao, Hewei
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [29] Technical note: TIGRE-DE for the creation of virtual monoenergetic images from dual-energy cone-beam CT
    Keeler, Andrew
    Lehmann, Mathias
    Luce, Jason
    Kaur, Mandeep
    Roeske, John
    Kang, Hyejoo
    [J]. MEDICAL PHYSICS, 2024, 51 (04) : 2975 - 2982
  • [30] Dual-Energy Imaging of Bone Marrow Edema on a Dedicated Multi-Source Cone-Beam CT System for the Extremities
    Zbijewski, W.
    Sisniega, A.
    Stayman, J. W.
    Thawait, G.
    Packard, N.
    Yorkston, J.
    Demehri, S.
    Fritz, J.
    Siewerdsen, J. H.
    [J]. MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412