Serial angiographic and intravascular ultrasound analysis of late stent strut fracture of sirolimus-eluting stents in native coronary arteries

被引:32
|
作者
Yamada, Ko P. [2 ]
Koizumi, Tomomi [1 ]
Yamaguchi, Hiroshi [3 ]
Kaneda, Hideaki [1 ]
Bonneau, Heidi N. [1 ]
Honda, Yasuhiro [1 ]
Fitzgerald, Peter J. [1 ]
机构
[1] Stanford Univ, Ctr Cardiovasc Technol, Med Ctr, Div Cardiovasc Med, Stanford, CA 94305 USA
[2] Stanford Univ, Med Ctr, Dept Med, Stanford, CA 94305 USA
[3] Tenyoukai Cent Hosp, Div Cardiol, Kagoshima, Japan
关键词
Stent fracture; SES; Sirolimus-eluting stent; Coronary angiography; IVUS; Intravascular ultrasound;
D O I
10.1016/j.ijcard.2007.08.082
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Stent fracture in sirolimus-eluting stents (SES) has been reported to be associated with late adverse events. However, a suitable method to diagnose stent fracture is not fully elucidated. Methods: One hundred and two consecutive SES implantations were performed in 83 lesions in 56 patients and underwent serial angiography with intravascular ultrasound (IVUS) at baseline and at 6 months follow-up. Angiographic stent strut fracture was defined as stent bending with separation of stent struts. Angiographic hinge movement was defined as stent shaft deviation without separation of stent struts. IVUS stent strut dissociation was defined as the disappearance of stent struts in more than one cross-sectional image which were previously visualized at baseline. Results: By angiography, no cases of stent fracture were detected at 6 months. One case of angiographic hinge movement was found at 12 months. However, three instances of stent fracturewere detected by IVUS at 6 months. One case of stent fracture showed a patent lumen area at 6 months but subsequently developed late stent restenosis at 12 months. The other two cases were associated with in-stent restenosis at 6 months. Conclusion: Compared to angiography, IVUS can more reliably detect stent fracture during follow-up evaluation. (C) 2007 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:255 / 259
页数:5
相关论文
共 50 条
  • [31] Intravascular ultrasound study of effects of overlapping sirolimus-eluting stents
    Muñoz, JS
    Abizaid, A
    Mintz, GS
    Albertal, M
    Abizaid, AS
    Feres, F
    Centemero, M
    Staico, R
    Mattos, LA
    Pinto, L
    Sousa, A
    Sousa, JE
    AMERICAN JOURNAL OF CARDIOLOGY, 2004, 93 (04): : 470 - 473
  • [32] Late angiographic stent thrombosis after sirolimus-eluting stent implantation
    Takahashi, Saeko
    Kaneda, Hideaki
    Tanaka, Shinji
    Miyashita, Yusuke
    Shiono, Takaaki
    Taketani, Yoshio
    Domae, Hiroshi
    Matsumi, Junya
    Mizuno, Shingo
    Minami, Yoshiyasu
    Sugitatsu, Kazuya
    Saito, Shigern.
    CIRCULATION JOURNAL, 2007, 71 (02) : 226 - 228
  • [33] Comparison of Sirolimus-Eluting Stents With and Without Polymeric Coating: A Serial Angiography and Intravascular Ultrasound Study
    Chamie, Daniel
    Costa, J. Ribamar, Jr.
    Abizaid, Alexandre
    Almiro Silva, J. Fabio
    Feres, Fausto
    Abizaid, Andrea S.
    Tanajura, Luiz Fernando
    Mattos, Luiz Alberto P.
    Staico, Rodolfo
    Costa, Ricardo
    Sousa, Amanda G. M. R.
    Sousa, J. Eduardo
    AMERICAN JOURNAL OF CARDIOLOGY, 2008, 102 (8A): : 170I - 170I
  • [34] Serial intravascular ultrasound analysis of vascular remodeling at sites of incomplete stent apposition after sirolimus-eluting stent implantation
    Kimura, M
    Carlier, SG
    Mintz, GS
    Lui, J
    Costa, JD
    Sano, K
    Takebayashi, H
    Yasuda, T
    Fujii, K
    Mehran, R
    Dangas, G
    Stone, GW
    Moses, JW
    Leon, MB
    AMERICAN JOURNAL OF CARDIOLOGY, 2005, 96 (7A): : 171H - 172H
  • [35] Very late stent fracture associated with a sirolimus-eluting stent
    Leong, D. P.
    Dundon, B. K.
    Puri, R.
    Yeend, R. A. S.
    HEART LUNG AND CIRCULATION, 2008, 17 (05): : 426 - 428
  • [36] Overlapping sirolimus-eluting stents: Intravascular ultrasound analysis from the SIRIUS trial
    Morino, Y
    Ako, J
    Honda, Y
    Sonoda, S
    Terashima, M
    Hassan, A
    Holmes, DR
    Williams, DO
    Fischell, TA
    Keim, E
    Yock, PG
    Leon, MB
    Moses, JW
    Fitzgerald, PJ
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 41 (06) : 81A - 81A
  • [37] DIFFERENCES IN CORONARY INTRAVASCULAR ULTRASOUND FINDINGS IN EARLY, LATE, AND VERY LATE STENT THROMBOSIS AFTER SIROLIMUS-ELUTING STENT IMPLANTATION.
    Kimura, Masashi
    Ebisawa, Soichiro
    Tanaka, Nobuyoshi
    Nasu, Kenya
    Ehara, Mariko
    Ito, Tatsuya
    Kinoshita, Yoshihisa
    Tsuchikae, Etsuo
    Terashima, Mitsuyasu
    Asaskura, Yasushi
    Katoh, Osamu
    Suzuki, Takahiko
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (10)
  • [38] Intravascular Ultrasound Comparison of with Early, Late, and Very Late Stent Thrombosis After Sirolimus-Eluting Stent Implantation
    Kume, Teruyoshi
    Waseda, Katsuhisa
    Kozuma, Ken
    Aizawa, Tadanori
    Mitsudo, Kazuaki
    Miyazaki, Shinichi
    Yamaguchi, Tetsu
    Isshiki, Takaaki
    Kimura, Takeshi
    Fitzgerald, Peter J.
    Honda, Yasuhiro
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 56 (13) : B10 - B10
  • [39] Intravascular ultrasound assessment of lesions with sirolimus-eluting stent failure
    Takebayashi, H
    Kobayashi, Y
    Mintz, GS
    Fujii, K
    Yasuda, T
    Mehran, R
    Jacoboff, D
    Dangas, GD
    Lansky, AJ
    Moussa, I
    Collins, MB
    Kreps, E
    Stone, GW
    Leon, MB
    Moses, JW
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (05) : 65A - 66A
  • [40] Intravascular ultrasound analysis of sirolimus-eluting stents for the treatment of in-stent restenosis:: Results from the TROPICAL trial
    Sakurai, R
    Ako, J
    Kaneda, H
    Hassan, AHM
    Honda, Y
    Grube, E
    Mügge, A
    Aengevaeren, W
    Yock, PG
    Neumann, FJ
    Desmet, W
    Fitzgerald, PJ
    AMERICAN JOURNAL OF CARDIOLOGY, 2005, 96 (7A): : 95H - 95H