Serial angiographic and intravascular ultrasound analysis of late stent strut fracture of sirolimus-eluting stents in native coronary arteries

被引:32
|
作者
Yamada, Ko P. [2 ]
Koizumi, Tomomi [1 ]
Yamaguchi, Hiroshi [3 ]
Kaneda, Hideaki [1 ]
Bonneau, Heidi N. [1 ]
Honda, Yasuhiro [1 ]
Fitzgerald, Peter J. [1 ]
机构
[1] Stanford Univ, Ctr Cardiovasc Technol, Med Ctr, Div Cardiovasc Med, Stanford, CA 94305 USA
[2] Stanford Univ, Med Ctr, Dept Med, Stanford, CA 94305 USA
[3] Tenyoukai Cent Hosp, Div Cardiol, Kagoshima, Japan
关键词
Stent fracture; SES; Sirolimus-eluting stent; Coronary angiography; IVUS; Intravascular ultrasound;
D O I
10.1016/j.ijcard.2007.08.082
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Stent fracture in sirolimus-eluting stents (SES) has been reported to be associated with late adverse events. However, a suitable method to diagnose stent fracture is not fully elucidated. Methods: One hundred and two consecutive SES implantations were performed in 83 lesions in 56 patients and underwent serial angiography with intravascular ultrasound (IVUS) at baseline and at 6 months follow-up. Angiographic stent strut fracture was defined as stent bending with separation of stent struts. Angiographic hinge movement was defined as stent shaft deviation without separation of stent struts. IVUS stent strut dissociation was defined as the disappearance of stent struts in more than one cross-sectional image which were previously visualized at baseline. Results: By angiography, no cases of stent fracture were detected at 6 months. One case of angiographic hinge movement was found at 12 months. However, three instances of stent fracturewere detected by IVUS at 6 months. One case of stent fracture showed a patent lumen area at 6 months but subsequently developed late stent restenosis at 12 months. The other two cases were associated with in-stent restenosis at 6 months. Conclusion: Compared to angiography, IVUS can more reliably detect stent fracture during follow-up evaluation. (C) 2007 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:255 / 259
页数:5
相关论文
共 50 条
  • [1] Prediction of late restenosis after sirolimus-eluting stent implantation using serial quantitative angiographic and intravascular ultrasound analysis
    Funada R.
    Oikawa Y.
    Yajima J.
    Matsuno S.
    Kano H.
    Kirigaya H.
    Nagashima K.
    Kurabayashi M.
    Aizawa T.
    Cardiovascular Intervention and Therapeutics, 2011, 26 (1) : 26 - 32
  • [2] Late incomplete stent apposition after sirolimus-eluting stent implantation - A serial intravascular ultrasound analysis
    Ako, J
    Morino, Y
    Honda, Y
    Hassan, A
    Sonoda, S
    Yock, PG
    Leon, MB
    Moses, JW
    Bonneau, HN
    Fitzgerald, PJ
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 46 (06) : 1002 - 1005
  • [3] Incidence of late angiographic stent thrombosis of sirolimus-eluting stents
    Kalaga, Ravishankar V.
    Mirza, Zulfiqar A.
    Shafeeq, Ahmed
    Kadakia, Hemal H.
    Ruffin, Richard T., Jr.
    Kokkirala, Aravind Rao
    Kiernan, Francis J.
    Lundbye, Justin B.
    Dacia, Marcin
    Malik, Amyn
    Murphy, Deborah
    Mather, Jeff
    McKay, Raymond G.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 49 (09) : 37B - 37B
  • [4] Intravascular ultrasound stent area of sirolimus-eluting stents and its impact on late outcome
    Cheneau, E
    Pichard, AD
    Satler, LF
    Suddath, WO
    Weissman, NJ
    Waksman, R
    AMERICAN JOURNAL OF CARDIOLOGY, 2005, 95 (10): : 1240 - 1242
  • [5] MECHANISM OF EDGE RESTENOSIS OF SIROLIMUS-ELUTING STENTS: SERIAL INTRAVASCULAR ULTRASOUND ANALYSIS
    Ichimoto, Eiji
    Kobayashi, Yoshio
    Kuroda, Nakabumi
    Komuro, Issei
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (10)
  • [6] Sirolimus-eluting stents abolish neointimal hyperplasia in patients with in-stent restenosis: Late angiographic and intravascular ultrasound results
    Serruys, PW
    Abizaid, A
    Foley, D
    Abizaid, A
    de Feyter, P
    Popma, J
    Degertakin, M
    Staico, R
    Pinto, L
    Wuelfert, E
    Sousa, A
    Sousa, JE
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (05) : 37A - 37A
  • [7] Serial intravascular ultrasound assessment of very late stent thrombosis after sirolimus-eluting stent placement
    Yamanaga, Kenshi
    Tsujita, Kenichi
    Shimomura, Hideki
    Nakamura, Yoshinori
    Ogura, Yuji
    Onoue, Yoshiro
    Chazono, Naomi
    Nagata, Takeshi
    Morisaki, Shogo
    Kudo, Takashi
    Yamada, Yoshihiro
    Komura, Naohiro
    Miyazaki, Takashi
    Akasaka, Tomonori
    Florio, Eiji
    Sato, Koji
    Arima, Yuichiro
    Kojima, Sunao
    Hokimoto, Seiji
    Ogawa, Hisao
    JOURNAL OF CARDIOLOGY, 2014, 64 (3-4) : 279 - 284
  • [8] Quantitative assessment of angiographic restenosis after sirolimus-eluting stent implantation in native coronary arteries
    Popma, JJ
    Leon, MB
    Moses, JW
    Holmes, DR
    Cox, N
    Fitzpatrick, M
    Douglas, J
    Lambert, C
    Mooney, M
    Yakubov, S
    Kuntz, RE
    CIRCULATION, 2004, 110 (25) : 3773 - 3780
  • [9] A Qualitative and Quantitative Angiographic Analysis of Stent Fracture Late Following Sirolimus-Eluting Stent Implantation
    Popma, Jeffrey J.
    Tiroch, Klaus
    Almonacid, Alexandra
    Cohen, Sidney
    Kandzari, David E.
    Leon, Martin B.
    AMERICAN JOURNAL OF CARDIOLOGY, 2009, 103 (07): : 923 - 929
  • [10] Serial angiographic findings of stent fracture lesion after sirolimus-eluting stent implantation
    Miyake, K.
    Kadota, K.
    Hosogi, H.
    Tada, T.
    Tanaka, T.
    Fuku, F.
    Katou, K.
    Yamamoto, Y.
    Gotou, G.
    Mitudou, M.
    EUROPEAN HEART JOURNAL, 2011, 32 : 956 - 956