Pacemaker lead extraction and recapture of venous access: Technical problems arising from extensive venous obstruction

被引:11
|
作者
Kutarski, Andrzej [1 ]
Pietura, Radoslaw [2 ]
Mlynarczyk, Krzysztof [3 ]
Malecka, Barbara [4 ]
Glowniak, Andrzej [1 ]
机构
[1] Med Univ Lublin, Dept Cardiol, PL-20954 Lublin, Poland
[2] Med Univ Lublin, Dept Intervent Radiol & Neuroradiol, PL-20954 Lublin, Poland
[3] Specialist Hosp, Dept Cardiol, Tarnow, Poland
[4] John Paul 2 Hosp, Dept Electrocardiol, Krakow, Poland
关键词
lead-related venous occlusion; lead extraction; recapture of venous access; VEIN; OCCLUSION; IMPLANTATION; REPLACEMENT; VENOPLASTY;
D O I
10.5603/CJ.2012.0093
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We report the case of the extraction of 18 year-old leads in a patient with a DDD pacemaker, and chronic obstruction of the left subclavian and innominate, veins coexisting with extensive stenoses in the upper caved vein: After removal of pacing leads, angiographic guidewires were introduced via the Byrd dilatators and new pacing leads introduced with the use of long sheaths originally dedicated for transvenous left ventricular leads implantation. With this case, we discuss the problemarising during reimplantation of pacing leads in patients with chronic venous occlusion. (Cardiol J 2012; 19, 5: 513-517)
引用
收藏
页码:513 / 517
页数:5
相关论文
共 22 条
  • [21] Multicenter Experience with the Surfacer Inside-Out Access Catheter System in Patients with Thoracic Venous Obstruction: Results from the SAVE Registry
    Gallieni, Maurizio
    Matoussevitch, Vladimir
    Steinke, Tobias
    Ebner, Adrian
    Brunkwall, Silke
    Cariati, Maurizio
    Gallo, Santiago
    Reindl-Schwaighofer, Roman
    Sengoelge, Guerkan
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2020, 31 (10) : 1654 - +
  • [22] Modification of cochlear potentials produced by streptomycin poisoning and by extensive venous obstruction (Reprinted from Laryngoscope, vol 68, pg 596-627, 1958)
    Mattox, DE
    LARYNGOSCOPE, 1997, 107 (02): : 162 - 165