Evaluation of the LABGEO PT10 Point-Of-Care Testing Device: Comparison of Analyte Measurements in Capillary Whole Blood and Lithium Heparin Whole Blood Samples With Those in Central Laboratory

被引:0
|
作者
Lim, Jinsook [1 ]
Kim, Hyunjin [1 ]
Koo, Sun Hoe [1 ]
Kwon, Gye Cheol [1 ]
机构
[1] Chungnam Natl Univ Hosp, Dept Lab Med, 33 Munhwa Ro, Daejeon 35015, South Korea
关键词
Analytic chemistry technique; capillary blood; POCT; point-of-care testing; PERFORMANCE; POCT;
D O I
10.1002/jcla.22050
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Point-of-care (POC) testing device has been widely used because of its rapid availability of results making diagnosis and management as early as possible. Capillary blood can reduce the difficulty of obtaining samples compared to venous blood and allows the prompt testing results. In this study, we evaluated the usefulness of capillary blood in Samsung LABGEO PT10. Methods: Fifty-one patients and 18 healthy adults aged between 20 and 65 were enrolled and their capillary and venous blood samples were collected. Venous blood samples were split into lithium heparin (LiHep) tube and serumseparating tube. Measurements using capillary blood and LiHep whole blood were performed in LABGEO PT10. Serum was used for measurement by Toshiba 2000FR NEO in central laboratory. Results: In comparison between measurements in LABGEO PT10 using capillary and LiHep whole blood, the slope ranged between 0.9289 and 1.0471, correlation coefficients (R-2) were over 0.95 except albumin, high-density lipoprotein, and total protein. Comparison of measurements in capillary and LiHep whole blood using LABGEO PT10 with those in the central laboratory revealed that the slope ranged between 0.6433 and 1.1364 for capillary whole blood and 0.6255 and 1.1602 for LiHep whole blood except alkaline phosphatase. For most of analytes, R-2 were over 0.95. Conclusion: Measurements in LABGEO PT10 using capillary blood was well correlated with those in LABGEO PT10 using LiHep whole blood and also with in the central laboratory. In conclusion, capillary blood provides reliable measurements and can be trustfully used in LABGEO PT10. (C) 2016 Wiley Periodicals, Inc.
引用
收藏
页数:11
相关论文
共 37 条
  • [21] European concerted action on anticoagulation - Comparison of fresh plasma and whole blood multicentre ISI calibrations of CoaguChek mini and TAS PT-NC whole blood prothrombin time point-of-care monitors
    Poller, L
    Keown, M
    Chauhan, N
    van den Besselaar, AMHP
    Tripodi, A
    Shiach, C
    Jespersen, J
    [J]. THROMBOSIS AND HAEMOSTASIS, 2002, 87 (05) : 859 - 866
  • [22] Comparability of point-of-care whole-blood electrolyte and substrate testing using a Stat Profile Critical Care Xpress analyzer and standard laboratory methods
    Flegar-Mestric, Zlata
    Perkov, Sonja
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2006, 44 (07) : 898 - 903
  • [23] Evaluation of Total Bilirubin Determination in Neonatal Whole-Blood Samples by Multiwavelength Photometry on the Roche OMNI S Point-of-Care Analyzer
    Rolinski, Boris
    Okorodudu, Anthony O.
    Kost, Gerald
    Roser, Markus
    Wu, Jiaxi
    Goerlach-Graw, Ada
    Kuester, Helmut
    [J]. POINT OF CARE, 2005, 4 (01): : 3 - 8
  • [24] Laboratory Evaluation of the Liat HIV Quant (IQuum) Whole-Blood and Plasma HIV-1 Viral Load Assays for Point-of-Care Testing in South Africa
    Scott, Lesley
    Gous, Natasha
    Carmona, Sergio
    Stevens, Wendy
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2015, 53 (05) : 1616 - 1621
  • [25] Evaluation of the Xpert HCV Viral Load point-of-care assay from venepuncture-collected and finger-stick capillary whole-blood samples: a cohort study
    Grebely, Jason
    Lamoury, Francois M. J.
    Hajarizadeh, Behzad
    Mowat, Yasmin
    Marshall, Alison D.
    Bajis, Sahar
    Marks, Philippa
    Amin, Janaki
    Smith, Julie
    Edwards, Michael
    Gorton, Carla
    Ezard, Nadine
    Persing, David
    Kleman, Marika
    Cunningham, Philip
    Catlett, Beth
    Dore, Gregory J.
    Applegate, Tanya L.
    [J]. LANCET GASTROENTEROLOGY & HEPATOLOGY, 2017, 2 (07): : 514 - 520
  • [26] Multicenter study of whole-blood creatinine, total carbon dioxide content, and chemistry profiling for laboratory and point-of-care testing in critical care in the United States
    Kost, GJ
    Vu, HT
    Inn, M
    DuPlantier, R
    Fleisher, M
    Kroll, MH
    Spinosa, JC
    [J]. CRITICAL CARE MEDICINE, 2000, 28 (07) : 2379 - 2389
  • [27] Development of a hybridization chain reaction-powered lab-on-fiber device for ultrafast point-of-care testing of circulating tuor DNA in whole blood
    Xu, Jiaxin
    Han, Xiangzhi
    Xu, Wenjuan
    Liu, Jiayuan
    Zhu, Anna
    Song, Dan
    Long, Feng
    [J]. TALANTA, 2023, 259
  • [28] Evaluation of the i-STAT point-of-care capillary whole blood hematocrit and hemoglobin: Comparison to the Siemens RAPIDLab 1200, Sysmex XE5000, and manual spun hematocrit
    Rudolf, Joseph
    Douglass, Joseph
    Baron, Jason
    Lewandrowski, Kent
    [J]. CLINICA CHIMICA ACTA, 2015, 446 : 37 - 42
  • [29] Bias and uncertainty of the International Normalized Ratio determined with a whole blood point-of-care prothrombin time test device by comparison to a new International Standard for thromboplastin
    van den Besselaar, Antonius M. H. P.
    van Rijn, Claudia J. J.
    Abdoel, Charmane F.
    Cobbaert, Christa M.
    [J]. THROMBOSIS RESEARCH, 2021, 202 : 1 - 7
  • [30] Field evaluation of Xpert® HCV Viral Load point-of-care test for detection of hepatitis C virus infection by venipuncture-collected and finger-stick capillary whole-blood samples
    Grebely, J.
    Lamoury, F.
    Hajarizadeh, B.
    Mowat, Y.
    Marshall, A.
    Bajis, S.
    Marks, P.
    Amin, J.
    Smith, J.
    Edwards, M.
    Gorton, C.
    Ezard, N.
    Persing, D.
    Kleman, M.
    Cunningham, P.
    Catlett, B.
    Dore, G.
    Applegate, T. L.
    [J]. JOURNAL OF HEPATOLOGY, 2017, 66 (01) : S709 - S709