Effect of Multileaf Collimator Leaf Position Error Determined by Picket Fence Test on Gamma Index Value in Patient-Specific Quality Assurance of Volumetric-Modulated Arc Therapy Plans

被引:2
|
作者
Ceylan, Cemile [1 ,2 ]
Inal, Serpil Yondem [3 ]
Senol, Elif [3 ]
Yilmaz, Berrin [1 ]
Sahin, Sevim [4 ]
机构
[1] Istanbul Oncol Hosp, Radiat Oncol Dept, Istanbul, Turkey
[2] Yeditepe Univ, Hlth Sci Inst, Istanbul, Turkey
[3] Mem Bahcelievler Hosp, Radiat Oncol Dept, Istanbul, Turkey
[4] Fenerbahce Univ, Med Imaging Dept, Istanbul, Turkey
关键词
mlc error; patient-specific quality assurance; gamma-index evaluation; DYNAMIC IMRT; SLIDING WINDOW; MLC; ACCURACY; SENSITIVITY; PERFORMANCE; DELIVERY; CONSEQUENCES;
D O I
10.7759/cureus.12684
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim The correlation between the MLC QA (IBA Dosimetry, Germany) results of the picket fence test created with intentional errors and the patient's quality assurance (QA) evaluation was investigated to assess the impact of multileaf collimator (MLC) positioning error on patient QA. Materials and methods The picket fence, including error-free and intentional MLC errors, defined in Bank In, Bank Out, and Bank Both were analyzed using MLC QA. The QA of 15 plans consisting of stereotactic radiosurgery (SRS), stereotactic body radiotherapy (SBRT), and conventionally fractionated volumetric-modulated arc therapy (VMAT) acquired with electronic portal imaging devices (EPID) was evaluated in the presence of error-free and MLC errors. The QA of plans were analyzed with 2%/2 mm and 3%/3 mm criteria. Results The passing rates of the picket fence test were 97%, 92%, 91%, and 87% for error-free and intentional errors. The criterion of 3%/3 mm wasn't able to detect an MLC error for either SRS/SBRT or conventionally fractionated VMAT. The criterion of 2%/2mm was more sensitive to detect MLC error for the conventionally fractionated VMAT than SRS/SBRT. While only two of SBRT plans had <90%, four of conventionally fractionated VMAT plans had a <90% passing rate. Conclusion We found that the systematic MLC positioning errors defined with picket fence have a smaller but measurable impact on SRS/SBRT than the VMAT plan for a conventionally fractionated and relatively complex plan such as head and neck and endometrium cases.
引用
下载
收藏
页数:13
相关论文
共 48 条
  • [21] A study on the correlation between plan complexity and gamma index analysis in patient specific quality assurance of volumetric modulated arc therapy
    Rajasekaran, Dhanabalan
    Jeevanandam, Prakash
    Sukumar, Prabakar
    Ranganathan, Arulpandiyan
    Johnjothi, Samdevakumar
    Nagarajan, Vivekanandan
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2015, 20 (01) : 57 - 65
  • [22] Error detection using a convolutional neural network with dose difference maps in patient-specific quality assurance for volumetric modulated arc therapy
    Kimura, Yuto
    Kadoya, Noriyuki
    Tomori, Seiji
    Oku, Yohei
    Jingu, Keiichi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 73 : 57 - 64
  • [23] Treatment plan complexity quantification for predicting gamma passing rates in patient-specific quality assurance for stereotactic volumetric modulated arc therapy
    Xue, Xudong
    Luan, Shunyao
    Ding, Yi
    Li, Xiangbin
    Li, Dan
    Wang, Jingya
    Ma, Chi
    Jiang, Man
    Wei, Wei
    Wang, Xiao
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024,
  • [24] Sensitivity of Three Patient-Specific Quality Assurance Systems to MLC Aperture Errors With Volumetric Modulated Arc Therapy
    Hu, Jinyou
    Gu, Shaoxian
    Wang, Ningyu
    Cui, Fengjie
    Zhang, Shengyuan
    Yin, Chuou
    Cai, Yunzhu
    Gou, Chengjun
    Zou, Lian
    Wu, Zhangwen
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2022, 21
  • [25] Cloud-Based Monte Carlo Patient-Specific Quality Assurance (QA) Method for Volumetric-Modulated Arc Therapy (VMAT): Clinical and Educational Impacts
    Chen, X.
    Xing, L.
    Mok, E.
    Luxton, G.
    Bush, K.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [26] Treatment Plan Complexity Quantification for Predicting Patient-Specific Quality Assurance of Stereotactic Volumetric Modulated Arc Therapy
    Xudong, X.
    Ding, Y.
    Wei, W.
    Ma, C.
    Wang, X.
    MEDICAL PHYSICS, 2022, 49 (06) : E286 - E287
  • [27] Systematic method for a deep learning-based prediction model for gamma evaluation in patient-specific quality assurance of volumetric modulated arc therapy
    Tomori, Seiji
    Kadoya, Noriyuki
    Kajikawa, Tomohiro
    Kimura, Yuto
    Narazaki, Kakutarou
    Ochi, Takahiro
    Jingu, Keiichi
    MEDICAL PHYSICS, 2021, 48 (03) : 1003 - 1018
  • [28] Development of a deep learning-based error detection system without error dose maps in the patient-specific quality assurance of volumetric modulated arc therapy
    Kimura, Yuto
    Kadoya, Noriyuki
    Oku, Yohei
    Jingu, Keiichi
    JOURNAL OF RADIATION RESEARCH, 2023, 64 (04) : 728 - 737
  • [29] Statistical analysis of correlation of gamma passing results for two quality assurance phantoms used for patient-specific quality assurance in volumetric modulated arc radiotherapy
    Kunii, Yuki
    Tanabe, Yoshinori
    Nakamoto, Akira
    Nishioka, Kunio
    MEDICAL DOSIMETRY, 2022, 47 (04) : 329 - 333
  • [30] Integrating plan complexity and dosiomics features with deep learning in patient-specific quality assurance for volumetric modulated arc therapy
    Han, Ce
    Zhang, Ji
    Yu, Bing
    Zheng, Haoze
    Wu, Yibo
    Lin, Zhixi
    Ning, Boda
    Yi, Jinling
    Xie, Congying
    Jin, Xiance
    RADIATION ONCOLOGY, 2023, 18 (01)