In vivo proton range verification: a review

被引:408
|
作者
Knopf, Antje-Christin [1 ]
Lomax, Antony [1 ]
机构
[1] Paul Scherrer Inst, Ctr Proton Therapy, Villigen, Switzerland
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2013年 / 58卷 / 15期
关键词
POSITRON-EMISSION-TOMOGRAPHY; BEAM PET MEASUREMENTS; SCATTERING COMPTON CAMERA; MONTE-CARLO SIMULATIONS; PROMPT-GAMMA; RADIATION-THERAPY; BONE-MARROW; RADIOTHERAPY TREATMENT; IMPLANTABLE DOSIMETER; FILTERING APPROACH;
D O I
10.1088/0031-9155/58/15/R131
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Protons are an interesting modality for radiotherapy because of their well defined range and favourable depth dose characteristics. On the other hand, these same characteristics lead to added uncertainties in their delivery. This is particularly the case at the distal end of proton dose distributions, where the dose gradient can be extremely steep. In practice however, this gradient is rarely used to spare critical normal tissues due to such worries about its exact position in the patient. Reasons for this uncertainty are inaccuracies and non-uniqueness of the calibration from CT Hounsfield units to proton stopping powers, imaging artefacts (e. g. due to metal implants) and anatomical changes of the patient during treatment. In order to improve the precision of proton therapy therefore, it would be extremely desirable to verify proton range in vivo, either prior to, during, or after therapy. In this review, we describe and compare state-of-the art in vivo proton range verification methods currently being proposed, developed or clinically implemented.
引用
收藏
页码:131 / 160
页数:30
相关论文
共 50 条
  • [41] Time-resolved diode dosimetry for in vivo proton therapy range verification: calibration through numerical modeling
    Toltz, Allison
    Hoesl, Michaela
    Schuemann, Jan
    Seuntjens, Jan
    Lu, Hsiao-Ming
    Paganetti, Harald
    MEDICAL PHYSICS, 2016, 43 (08) : 4939 - 4939
  • [42] Feasibility study of in vivo MRI based dosimetric verification of proton end-of-range for liver cancer patients
    Yuan, Yading
    Andronesi, Ovidiu C.
    Bortfeld, Thomas R.
    Richter, Christian
    Wolf, Russell
    Guimaraes, Alexander R.
    Hong, Theodore S.
    Seco, Joao
    RADIOTHERAPY AND ONCOLOGY, 2013, 106 (03) : 378 - 382
  • [43] Validation of An In-Vivo Proton Range Verification Method for Reducing the Risk of Permanent Alopecia in the Treatment of Pediatric Medulloblastoma
    Lucconi, G.
    Bentefour, E.
    Deepak, S.
    Janssens, G.
    Weaver, K.
    Moteabbed, M.
    Lu, H-M
    MEDICAL PHYSICS, 2016, 43 (06) : 3512 - 3512
  • [44] First in vivo proof-of-concept of nanodroplet-mediated ultrasound-based proton range verification
    Carlier, B.
    Collado-Lara, G.
    Heymans, S.
    Toumia, Y.
    Musetta, L.
    Paradossi, G.
    Vos, H.
    Van Den Abeele, K.
    D'hooge, J.
    Himmelreich, U.
    Sterpin, E.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S562 - S563
  • [45] First Demonstration of a Prompt Gamma Slit-Camera for in Vivo Proton Range Verification Based On Semiconductor Detectors
    Wang, H.
    Lu, Y.
    Kuang, Y.
    MEDICAL PHYSICS, 2018, 45 (06) : E509 - E509
  • [46] Clinical commissioning of an in vivo range verification system for prostate cancer treatment with anterior and anterior oblique proton beams
    Hoesl, M.
    Deepak, S.
    Moteabbed, M.
    Jassens, G.
    Orban, J.
    Park, Y. K.
    Parodi, K.
    Bentefour, E. H.
    Lu, H. M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (08): : 3049 - 3062
  • [47] In-vivo range verification analysis with in-beam PET data for patients treated with proton therapy at CNAO
    Moglioni, Martina
    Kraan, Aafke Christine
    Baroni, Guido
    Battistoni, Giuseppe
    Belcari, Nicola
    Berti, Andrea
    Carra, Pietro
    Cerello, Piergiorgio
    Ciocca, Mario
    De Gregorio, Angelica
    De Simoni, Micol
    Del Sarto, Damiano
    Donetti, Marco
    Dong, Yunsheng
    Embriaco, Alessia
    Fantacci, Maria Evelina
    Ferrero, Veronica
    Fiorina, Elisa
    Fischetti, Marta
    Franciosini, Gaia
    Giraudo, Giuseppe
    Laruina, Francesco
    Maestri, Davide
    Magi, Marco
    Magro, Giuseppe
    Malekzadeh, Etesam
    Marafini, Michela
    Mattei, Ilaria
    Mazzoni, Enrico
    Mereu, Paolo
    Mirandola, Alfredo
    Morrocchi, Matteo
    Muraro, Silvia
    Orlandi, Ester
    Patera, Vincenzo
    Pennazio, Francesco
    Pullia, Marco
    Retico, Alessandra
    Rivetti, Angelo
    Rolo, Manuel Dionisio Da Rocha
    Rosso, Valeria
    Sarti, Alessio
    Schiavi, Angelo
    Sciubba, Adalberto
    Sportelli, Giancarlo
    Tampellini, Sara
    Toppi, Marco
    Traini, Giacomo
    Trigilio, Antonio
    Valle, Serena Marta
    FRONTIERS IN ONCOLOGY, 2022, 12
  • [48] Applications of Machine Learning to Improve the Clinical Viability of Compton Camera Based in vivo Range Verification in Proton Radiotherapy
    Polf, Jerimy C.
    Barajas, Carlos A.
    Peterson, Stephen W.
    Mackin, Dennis S.
    Beddar, Sam
    Ren, Lei
    Gobbert, Matthias K.
    FRONTIERS IN PHYSICS, 2022, 10
  • [49] Acoustic Range Verification of Proton Beams: Simulation of Heterogeneity and Clinical Proton Pulses
    Jones, K.
    Sehgal, C.
    Avery, S.
    MEDICAL PHYSICS, 2015, 42 (06) : 3582 - 3582
  • [50] Treatment Planning and Verification with Proton CT and Proton Radiography to Reduce Range Uncertainties in Proton Therapy
    Schulte, R.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S242 - S242