Evaluation of the dose variation for prostate heavy charged particle therapy using four-dimensional computed tomography

被引:5
|
作者
Kumagai, Motoki [1 ]
Okada, Tohru [1 ]
Mori, Shinichiro [1 ]
Kandatsu, Susumu [1 ]
Tsuji, Hiroshi [1 ]
机构
[1] Natl Inst Radiol Sci, Res Ctr Charged Particle Therapy, Inage Ku, Chiba 2638555, Japan
关键词
charged particle therapy; 4DCT; intrafractional motion; prostate; REAL-TIME; HYPOFRACTIONATED RADIOTHERAPY; CONFORMAL RADIOTHERAPY; PERFORMANCE EVALUATION; ION RADIOTHERAPY; CT-SCANNER; MOTION; CANCER; IRRADIATION; TRACKING;
D O I
10.1093/jrr/rrs106
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
We quantified dose variation effects due to respiratory-induced intrafractional motion in conventional carbon-ion prostate treatment by using four-dimensional computed tomography (4DCT). 4DCT scans of 20 patients were acquired under free-breathing conditions using a 256 multi-slice CT scanner. The clinical target volume (CTV) was defined as the prostate and the seminal vesicle. Two types of planning target volumes (PTVs) were defined to minimize excessive dose to the rectum. The first PTV (= PTV1) was calculated by adding a 3D uniform margin to the CTV. The second PTV (= PTV2) was cut in a straight line from the top surface of the rectum from PTV1. Compensating boli were designed for the respective PTVs at the peak-exhalation phase, and carbon-ion dose distributions for a single respiratory cycle were calculated using these boli. Dose conformation to prostate, CTV, PTV1 and PTV2 were unchanged for all respiratory phases. The dose for > 95% volume irradiation (D95) was 97.7% for prostate, 92.5% for CTV, 74.1% for PTV1 and 96.1% for PTV2 averaged over all patients. The rectum volume at inhalation phase receiving < 50% of the prescribed dose was smaller than the planning dose due to the abdominal thickness variation. The target dose is not affected by intrafractional respiration in carbon-ion prostate treatment. Small dose variations, however, were observed due to respiratory-induced abdominal thickness variation; therefore the geometrical changes should be considered for prostate particle therapy.
引用
收藏
页码:357 / 366
页数:10
相关论文
共 50 条
  • [31] Modeling lung motion using consistent image registration in four-dimensional computed tomography for radiation therapy
    Lu, Wei
    Song, Joo Hyun
    Christensen, Gary E.
    Parikh, Parag J.
    Bradley, Jeffrey D.
    Low, Daniel A.
    MEDICAL IMAGING 2006: IMAGE PROCESSING, PTS 1-3, 2006, 6144
  • [32] The preoperative assessment of thoracic wall adhesions using four-dimensional computed tomography
    Jun Suzuki
    Satoshi Shiono
    Katsuyuki Suzuki
    Hikaru Watanabe
    Satoshi Takamori
    Takayuki Sasage
    Kaito Sato
    Yusuke Matsui
    Testsuro Uchida
    Fumika Watarai
    General Thoracic and Cardiovascular Surgery, 2023, 71 : 464 - 471
  • [33] Acquiring a four-dimensional computed tomography dataset using an external respiratory signal
    Vedam, SS
    Keall, PJ
    Kini, VR
    Mostafavi, H
    Shukla, HP
    Mohan, R
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (01): : 45 - 62
  • [34] Dynamic cine imaging of the Eustachian tube using four-dimensional computed tomography
    Jufas, N.
    Deveau, N.
    Bance, M.
    JOURNAL OF LARYNGOLOGY AND OTOLOGY, 2016, 130 (12): : 1162 - 1164
  • [35] Dynamic assessment of sternoclavicular joint instability using four-dimensional computed tomography
    Goh, Yin Peng
    Moaveni, Ash Kamali
    Hoy, Gregory
    Tate, Julie
    Rotstein, Andrew
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2019, 63 (02) : 216 - 221
  • [36] The preoperative assessment of thoracic wall adhesions using four-dimensional computed tomography
    Suzuki, Jun
    Shiono, Satoshi
    Suzuki, Katsuyuki
    Watanabe, Hikaru
    Takamori, Satoshi
    Sasage, Takayuki
    Sato, Kaito
    Matsui, Yusuke
    Uchida, Testsuro
    Watarai, Fumika
    GENERAL THORACIC AND CARDIOVASCULAR SURGERY, 2023, 71 (08) : 464 - 471
  • [37] Automated analysis of trapeziometacarpal joint kinematics using four-dimensional computed tomography
    Van Royen, Kjell
    Keelson, Benyameen
    Jager, Tjeerd
    Buls, Nico
    Scheerlinck, Thierry
    Cattrysse, Erik
    JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2024, 49 (09) : 1095 - 1103
  • [38] Both four-dimensional computed tomography and four-dimensional cone beam computed tomography under-predict lung target motion during radiotherapy
    Steiner, Elisabeth
    Shieh, Chun-Chien
    Caillet, Vincent
    Booth, Jeremy
    O'Brien, Ricky
    Briggs, Adam
    Hardcastle, Nicholas
    Jayamanne, Dasantha
    Szymura, Kathryn
    Eade, Thomas
    Keall, Paul
    RADIOTHERAPY AND ONCOLOGY, 2019, 135 : 65 - 73
  • [39] Four-dimensional computed tomography evaluation of shoulder joint motion in collegiate baseball pitchers
    Momma, Daisuke
    Orias, Alejandro A. Espinoza
    Irie, Tohru
    Irie, Tomoyo
    Kondo, Eiji
    Iwasaki, Norimasa
    Inoue, Nozomu
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [40] Target displacement evaluation for fluoroscopic and four-dimensional cone-beam computed tomography
    Iramina, H.
    Nakamura, M.
    Iizuka, Y.
    Matsuo, Y.
    Mizowaki, T.
    Hiraoka, M.
    Kanno, I.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S148 - S148