Real-time, three-dimensional echocardiography: Feasibility of dynamic right ventricular volume measurement with saline contrast

被引:50
|
作者
Ota, T [1 ]
Fleishman, CE [1 ]
Strub, M [1 ]
Stetten, G [1 ]
Ohazama, CJ [1 ]
von Ramm, OT [1 ]
Kisslo, J [1 ]
机构
[1] Duke Univ, Med Ctr, Ctr Emerging Cardiovasc Technol, Div Med & Biomed Engn, Durham, NC 27710 USA
关键词
D O I
10.1016/S0002-8703(99)70422-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background:The asymmetry and complex shape of the right ventricle have made:it difficult to determine right ventricular (RV) volume with 2-dimensional echocardiography. Three-dimensional cardiac imaging improves visualization of cardiac anatomy but is also complex and time consuming. A newly developed volumetric scanning system holds promise of obviating past limitations. Methods Real-time, transthoracic 3-dimensional echocardiographic images of the right ventricle were obtained with a high-speed volumetric ultrasound system that uses a 16:1 parallel processing schema From a 2.5 MHz matrix phased-array scanner to interrogate an entire pyramidal volume in real time. The instrumentation was used to measure RV volume in 8 excised canine hearts; dynamic real-time 3-dimensional images were also obtained from 14 normal subjects. Results Three-dimensional images were obtained in vitro and in vivo during intravenous hand-agitated saline injection to determine RV volumes. The RV volumes by real-time 3-dimentional echocardiography are well correlated with those of drained in vitro (y = 1.26x -9.92, r = 0.97; P < .0001, standard error of the estimate = .3.26 mL). For human subjects, the end-diastolic and end-systolic RV volumes were calculated by tracing serial cross-sectional, inclined C scans; functional data were validated by comparing the scans with conventional 2-dimensional echocardiographic indexes of left ventricular stroke volume. Conclusions These data indicate that RV volume measurements of excised heart by real-time 3-dimensional echocardiography are accurate and that beat-to-beat RV quantitative measurement applying this imaging method is possible. The new application of real-time 3-dimensional echocardiography presents the opportunity to develop new descriptors of cardiac performance.
引用
收藏
页码:958 / 966
页数:9
相关论文
共 50 条
  • [1] In vitro validation of right ventricular volume and mass measurement by real-time three-dimensional echocardiography
    Chen, GZ
    Sun, K
    Huang, GY
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2006, 23 (05): : 395 - 399
  • [2] Measurement of left ventricular volume by real-time, three-dimensional echocardiography in dogs
    Ota, T
    Fleishman, CE
    Ohazama, CJ
    Stetten, G
    Lewis, CW
    Glower, DD
    Li, J
    Ryan, T
    Kisslo, J
    vonRamm, OT
    CIRCULATION, 1996, 94 (08) : 2211 - 2211
  • [3] Right ventricular visualization and quantification using contrast in real-time three-dimensional echocardiography
    Van Der Zwaan, H. B.
    Gommans, F.
    Geleijnse, M. L.
    Mcghie, J. S.
    Meijboom, F. J.
    Van Dijk, A. P.
    Roos-Hesselink, J. W.
    EUROPEAN HEART JOURNAL, 2011, 32 : 1020 - 1021
  • [4] Evaluation of Right Ventricular Volume and Systolic Function by Real-time Three-dimensional Echocardiography
    王静
    王新房
    谢明星
    杨娅
    吕清
    杨颖
    王良玉
    华中科技大学学报(医学英德文版), 2005, (01) : 94 - 96
  • [5] Left ventricular volume estimation for real-time three-dimensional echocardiography
    Corsi, C
    Saracino, G
    Sarti, A
    Lamberti, C
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (09) : 1202 - 1208
  • [6] Real-time three-dimensional echocardiography for determining right ventricular stroke volume in an animal model of chronic right ventricular volume overload
    Shiota, T
    Jones, M
    Chikada, M
    Fleishman, CE
    Castellucci, JB
    Cotter, B
    DeMaria, AN
    von Ramm, OT
    Kisslo, J
    Ryan, T
    Sahn, DJ
    CIRCULATION, 1998, 97 (19) : 1897 - 1900
  • [7] Real-time three-dimensional echocardiography for measurement of left ventricular volumes
    Schmidt, MA
    Ohazama, CJ
    Agyeman, KO
    Freidlin, RZ
    Jones, M
    Laurienzo, JM
    Brenneman, CL
    Arai, AE
    von Ramm, OT
    Panza, JA
    AMERICAN JOURNAL OF CARDIOLOGY, 1999, 84 (12): : 1434 - 1439
  • [9] Right ventricular function using real-time three-dimensional echocardiography according to the acute volume reduction
    Park, Chan Seok
    Gwon, A. M.
    Kim, G. H.
    Ahn, H. S.
    Cho, J. S.
    Youn, H. J.
    EUROPEAN JOURNAL OF HEART FAILURE, 2019, 21 : 52 - 53
  • [10] Accuracy of real-time three-dimensional echocardiography for assessment of right ventricular stroke volume in children and young adults
    Hui, L
    Li, XK
    Jerosch-Herold, M