Multi-Material Decomposition using Low-Current X-Ray and a Photon-Counting CZT Detector

被引:0
|
作者
Kim, Sangtaek [1 ]
Hernandez, Andrew
Alhassen, Fares [1 ]
Pivovaroff, Michael [2 ]
Cho, Hyo-Min [3 ]
Gould, Robert G. [1 ]
Seo, Youngho [1 ]
机构
[1] Univ Calif San Francisco, Phys Res Lab, San Francisco, CA 94107 USA
[2] Lawrence Livermore Natl Lab, Livermore, CA 94550 USA
[3] Yonsei Univ, Wonju, South Korea
基金
美国国家卫生研究院;
关键词
CT; DENSITY; SYSTEM;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We developed and evaluated an x-ray photon-counting imaging system using an energy-resolving cadmium zinc telluride (CZT) detector coupled with application specific integrated circuit (ASIC) readouts. This x-ray imaging system can be used to identify different materials inside the object. The CZT detector has a large active area (5x5 array of 25 CZT modules, each with 16x16 pixels, cover a total area of 200 mm x 200 mm), high stopping efficiency for x-ray photons (similar to 100 % at 60 keV and 5 mm thickness). We explored the performance of this system by applying different energy windows around the absorption edges of target materials, silver and indium, in order to distinguish one material from another. The photon-counting CZT-based x-ray imaging system was able to distinguish between the materials, demonstrating its capability as a radiation-spectroscopic decomposition system.
引用
收藏
页码:4735 / 4738
页数:4
相关论文
共 50 条
  • [21] Performance of an improved readout IC for multi-energy photon-counting x-ray detector arrays
    Clajus, Martin
    Cajipe, Victoria B.
    Hayakawa, Satoshi
    Tuemer, Tuemay O.
    HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS X, 2008, 7079
  • [22] The effects of photon flux on energy spectra and imaging characteristics in a photon-counting x-ray detector
    Cho, H-M
    Kim, H-J
    Choi, Y-N
    Lee, S-W
    Ryu, H-J
    Lee, Y-J
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (14): : 4865 - 4879
  • [23] Microcomputed Tomography with A Second Generation Photon-counting X-ray Detector-Contrast Analysis and Material Separation
    Wang, X.
    Meier, D.
    Oya, P.
    Maehlum, G. E.
    Wagenaar, D. J.
    Tsui, B. M. W.
    Patt, B. E.
    Frey, E. C.
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [24] Material separation in x-ray CT with energy resolved photon-counting detectors
    Wang, Xiaolan
    Meier, Dirk
    Taguchi, Katsuyuki
    Wagenaar, Douglas J.
    Patt, Bradley E.
    Frey, Eric C.
    MEDICAL PHYSICS, 2011, 38 (03) : 1534 - 1546
  • [25] Multi-contrast CT imaging using a high energy resolution CdTe detector and a CZT photon-counting detector
    Clements, N.
    Richtsmeier, D.
    Hart, A.
    Bazalova-Carter, M.
    JOURNAL OF INSTRUMENTATION, 2022, 17 (01)
  • [26] Material Separation in X-ray CT with Energy Resolved Photon-Counting Detectors
    Wang, X.
    Meier, D.
    Taguchi, K.
    Wagenaar, D. J.
    Patt, B. E.
    Frey, E. C.
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [27] Evaluation of a photon counting Medipix3RX CZT spectral x-ray detector
    Jorgensen, Steven M.
    Vercnocke, Andrew J.
    Rundle, David S.
    Butler, Philip H.
    McCollough, Cynthia H.
    Ritman, Erik L.
    RADIATION DETECTORS: SYSTEMS AND APPLICATIONS XVII, 2016, 9969
  • [28] Tilted angle CZT detector for photon counting/energy weighting x-ray and CT imaging
    Shikhaliev, Polad M
    Physics in Medicine and Biology, 2006, 51 (17): : 4267 - 4287
  • [29] Development of a pixel readout ASIC for CZT detectors for spectral X-ray photon-counting imaging applications
    Wang, X.
    Deng, Z.
    Liu, W.
    Liu, Y.
    JOURNAL OF INSTRUMENTATION, 2020, 15 (01):
  • [30] Photon-counting X-ray computed tomography using a YAP(Ce)-PMT detector and beam hardening
    Watanabe, Manabu
    Sato, Eiichi
    Yoshida, Sohei
    Yoshioka, Kunihiro
    Oda, Yasuyuki
    Moriyama, Hodaka
    Hagiwara, Osahiko
    Enomoto, Toshiyuki
    HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS XXI, 2019, 11114