Development of an automated sequential injection on-line solvent extraction-back extraction procedure as demonstrated for the determination of cadmium with detection by electrothermal atomic absorption spectrometry

被引:69
|
作者
Wang, JH [1 ]
Hansen, EH [1 ]
机构
[1] Tech Univ Denmark, Dept Chem, DK-2800 Lyngby, Denmark
关键词
extraction-back extraction; sequential injection; electrothermal atomic absorption spectrometry; assay of certified reference materials; cadmium;
D O I
10.1016/S0003-2670(02)00026-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automated sequential injection (SI) on-line solvent extraction-back extraction separation/preconcentration procedure is described. Demonstrated for the assay of cadmium by electrothermal atomic absorption spectrometry (ETAAS), the analyte is initially complexed with ammonium pyrrolidinedithiocarbamate (APDC) in citrate buffer and the chelate is extracted into isobutyl methyl ketone (IBMK), which is separated from the aqueous phase by means of a newly designed dual-conical gravitational phase separator. A metered amount of the organic eluate is aspirated and stored in the PTFE holding coil (HC) of the SI-system. Afterwards, it is dispensed and mixed with an aqueous back extractant of dilute nitric acid containing Hg(II) ions as stripping agent, thereby facilitating a rapid metal-exchange reaction with the APDC ligand and transfer of the Cd into the aqueous phase. The aqueous phase is separated in a second dual-conical gravitational phase separator, and 30 mul of it is entrapped and metered in a sample loop (SL) and subsequently introduced via air segmentation into the graphite tube for analyte quantification. The ETAAS determination is performed in parallel with the separation/preconcentration process of the ensuing sample. An enrichment factor of 21.4, a detection limit of 2.7 ng l(-1), along with a sampling frequency of 13 h(-1) were obtained at a sample flow rate of 6.0 ml min(-1). The precision (R.S.D.) at the 0.4 mug l(-1) level was 1.8% as compared to 3.2% when quantifying the organic extractant directly. The applicability of the procedure is demonstrated for the determination of trace levels of cadmium in three certified reference materials. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:283 / 292
页数:10
相关论文
共 50 条
  • [21] Determination of traces of iron in indium phosphide by electrothermal atomic absorption spectrometry combined with solvent extraction
    Taddia, M
    Bagnoli, G
    Cerroni, MG
    Fornari, R
    Righetti, G
    ANNALI DI CHIMICA, 2002, 92 (5-6) : 491 - 500
  • [22] Determination of cadmium in seawater using automated on-line preconcentration and direct injection graphite furnace atomic absorption spectrometry
    Colbert, D
    Johnson, KS
    Coale, KH
    ANALYTICA CHIMICA ACTA, 1998, 377 (2-3) : 255 - 262
  • [23] SOLVENT-EXTRACTION OF CADMIUM AS A PREVIOUS STEP FOR ITS DETERMINATION IN BIOLOGICAL SAMPLES BY ELECTROTHERMAL ATOMIZATION ATOMIC-ABSORPTION SPECTROMETRY
    ALMENDRO, JME
    OJEDA, CB
    DETORRES, AG
    PAVON, JMC
    TALANTA, 1993, 40 (11) : 1643 - 1648
  • [24] Automated determination of tin and nickel in brass by on-line anodic electrodissolution and electrothermal atomic absorption spectrometry
    da Silva, JBB
    Giacomelli, MBO
    de Souza, IG
    Curtius, AJ
    TALANTA, 1998, 47 (05) : 1191 - 1198
  • [25] Automated on-line separation-preconcentration system for platinum determination by electrothermal atomic absorption spectrometry
    Ojeda, CB
    Rojas, FS
    Pavón, JMC
    de Torres, AG
    ANALYTICA CHIMICA ACTA, 2003, 494 (1-2) : 97 - 103
  • [26] An automated flow injection system for metal determination by flame atomic absorption spectrometry involving on-line fabric disk sorptive extraction technique
    Anthemidis, A.
    Kazantzi, V.
    Samanidou, V.
    Kabir, A.
    Furton, K. G.
    TALANTA, 2016, 156 : 64 - 70
  • [27] Flow injection on-line hydrophobic sorbent extraction for flame atomic absorption spectrometric determination of cadmium in water samples
    Anthemidis, Aristidis N.
    Karapatouchas, Christos-Paraskevas P.
    MICROCHIMICA ACTA, 2008, 160 (04) : 455 - 460
  • [28] Flow injection on-line hydrophobic sorbent extraction for flame atomic absorption spectrometric determination of cadmium in water samples
    Aristidis N. Anthemidis
    Christos-Paraskevas P. Karapatouchas
    Microchimica Acta, 2008, 160 : 455 - 460
  • [29] Development of a slurry-extraction procedure for direct determination of cobalt by electrothermal atomic absorption spectrometry in complex environmental samples
    Felipe-Sotelo, M
    Carlosena, A
    Andrade, JM
    Fernández, E
    López-Mahía, P
    Muniategui, S
    Prada, D
    ANALYTICA CHIMICA ACTA, 2004, 522 (02) : 259 - 266
  • [30] Determination of Cadmium by Electrothermal Atomic Absorption Spectrometry Using Single Drop Micro Extraction in Real Samples
    Chamsaz, M.
    Arbab-Zavar, M. H.
    Riazi, M. M.
    Takjoo, R.
    ASIAN JOURNAL OF CHEMISTRY, 2011, 23 (02) : 547 - 550