An evaluation of some pertinent parameters that influence the dosimetric performance of synthetic diamond detectors

被引:6
|
作者
Ade, N. [1 ,2 ]
Nam, T. L. [1 ,2 ]
Mhlanga, S. H. [3 ]
机构
[1] Univ Witwatersrand, Sch Phys, Radiat & Hlth Phys Unit, ZA-2050 Johannesburg, South Africa
[2] Univ Witwatersrand, DST NRF Ctr Excellence Strong Mat, ZA-2050 Johannesburg, South Africa
[3] Charlotte Maxeke Johannesburg Acad Hosp, Dept Radiat Sci, Div Med Phys, ZA-2000 Johannesburg, South Africa
基金
新加坡国家研究基金会;
关键词
Diamond; Pre-irradiation; Dose rate dependence; Relative dose; Angular dependence; HIGH-ENERGY ELECTRON; CHEMICAL-VAPOR-DEPOSITION; DOSE-RATE DEPENDENCE; CVD-DIAMOND; PHOTON BEAMS; RADIATION-THERAPY; RADIOTHERAPY; DEFECTS; PROBE; DISTRIBUTIONS;
D O I
10.1016/j.radphyschem.2013.01.038
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Although the near-tissue equivalence of diamond allows the direct measurement of dose for clinical applications without the need for energy-corrections, it is often cited that diamond detectors require pre-irradiation, a procedure necessary to stabilize the response or sensitivity of a diamond detector before dose measurements. In addition it has been pointed out that the relative dose measured with a diamond detector requires dose rate dependence correction and that the angular dependence of a detector could be due to its mechanical design or to the intrinsic angular sensitivity of the detection process. While the cause of instability of response has not been meticulously investigated, the issue of dose rate dependence correction is uncertain as some studies ignored it but reported good results. The aims of this study were therefore to investigate, in particular (1) the major cause of the unstable response of diamond detectors requiring pre-irradiation; (2) the influence of dose rate dependence correction in relative dose measurements; and (3) the angular dependence of the diamond detectors. The study was conducted with low-energy X-rays and electron therapy beams on HPHT and CVD synthesized diamonds. Ionization chambers were used for comparative measurements. Through systematic investigations, the major cause of the unstable response of diamond detectors requiring the recommended pre-irradiation step was isolated and attributed to the presence and effects of ambient light. The variation in detector's response between measurements in light and dark conditions could be as high as 63% for a CVD diamond. Dose rate dependence parameters (Delta values) of 0.950 and 1.035 were found for the HPHT and CVD diamond detectors, respectively. Without corrections based on dose rate dependence, the relative differences between depth-doses measured with the diamond detectors and a Markus chamber for exposures to 7 and 14 MeV electron beams were within 2.5%. A dose rate dependence correction using the Delta values obtained seemed to worsen the performance of the HPHT sample (up to about 3.3%) but it had a marginal effect on the performance of the CVD sample. In addition, the angular response of the CVD diamond detector was shown to be comparable with that of a cylindrical chamber. This study concludes that once the responses of the diamond detectors have been stabilised and they are properly shielded from ambient light, pre-irradiation prior to each measurement is not required. Also, the relative dose measured with the diamond detectors do not require dose rate dependence corrections as the required correction is only marginal and could have no dosimetric significance. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:42 / 51
页数:10
相关论文
共 50 条
  • [1] An evaluation of a novel synthetic diamond probe for dosimetric applications
    Ade, N.
    Nam, T. L.
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2015, 115 : 97 - 106
  • [2] Thermoluminescence of secondary peak on synthetic quartz: Influence of annealing time on kinetic parameters and some dosimetric features
    Dawam, R. R.
    Masok, F. B.
    Fierkwap, S. B.
    [J]. JOURNAL OF LUMINESCENCE, 2021, 233
  • [3] Evaluation of the Dosimetric Properties of a Synthetic Single Crystal Diamond Detector in Clinical Proton Beams
    Mandapaka, A.
    Verona-Rinati, G.
    Ghebremedhin, A.
    Patyal, B.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [4] Influence of some variable parameters on bremsstrahlung dosimetric characteristic for electron linac
    Baltateanu, N.
    Spanulescu, I.
    Pirnau, Mironela
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2008, 266 (10): : 2134 - 2137
  • [5] Evaluation of Some Essential Dosimetric Parameters for Beam Matching of Similar Accelerators
    Bhangle, Janhavi
    Narayanan, V. K. Sathiya
    Krishnakumar, Namitha
    Joseph, Jolly
    Nirhali, Amit
    Vaitheeswaran, R.
    [J]. MEDICAL PHYSICS, 2010, 37 (06) : 3256 - +
  • [6] The influence of uniformity of synthetic diamond grits on the drilling tool performance
    N. V. Novikov
    G. P. Bogatyreva
    G. D. Il’nitskaya
    G. A. Petasyuk
    A. M. Isonkin
    I. N. Zaitseva
    [J]. Journal of Superhard Materials, 2012, 34 : 141 - 147
  • [7] The influence of uniformity of synthetic diamond grits on the drilling tool performance
    Novikov, N. V.
    Bogatyreva, G. P.
    Il'nitskaya, G. D.
    Petasyuk, G. A.
    Isonkin, A. M.
    Zaitseva, I. N.
    [J]. JOURNAL OF SUPERHARD MATERIALS, 2012, 34 (02) : 141 - 147
  • [8] Improved performance in synthetic diamond neutron detectors: Application to boron neutron capture therapy
    Almaviva, S.
    Marinelli, Marco
    Milani, E.
    Prestopino, G.
    Tucciarone, A.
    Verona, C.
    Verona-Rinati, G.
    Angelone, M.
    Pillon, M.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 612 (03): : 580 - 582
  • [9] The effect of fast neutron irradiation on the performance of synthetic single crystal diamond particle detectors
    Lohstroh, A.
    Sellin, P. J.
    Gkoumas, S.
    Al-Barakaty, H.
    Veeramani, P.
    Ozsan, M. E.
    Prekas, G.
    Veale, M. C.
    Parkin, J. M.
    Davies, A. W.
    [J]. DIAMOND AND RELATED MATERIALS, 2010, 19 (7-9) : 841 - 845
  • [10] Evaluation of the dosimetric properties of a synthetic single crystal diamond detector in high energy clinical proton beams
    Mandapaka, A. K.
    Ghebremedhin, A.
    Patyal, B.
    Marinelli, Marco
    Prestopino, G.
    Verona, C.
    Verona-Rinati, G.
    [J]. MEDICAL PHYSICS, 2013, 40 (12)