Measurement of Tear Film Thickness Using Ultrahigh-Resolution Optical Coherence Tomography

被引:111
|
作者
Werkmeister, Rene M. [1 ]
Alex, Aneesh [1 ]
Kaya, Semira [2 ]
Unterhuber, Angelika [1 ]
Hofer, Bernd [1 ]
Riedl, Jasmin [1 ]
Bronhagl, Michael [1 ]
Vietauer, Martin [1 ]
Schmidl, Doreen [2 ]
Schmoll, Tilman [1 ]
Garhoefer, Gerhard [2 ]
Drexler, Wolfgang [1 ]
Leitgeb, Rainer A. [1 ]
Groeschl, Martin [3 ]
Schmetterer, Leopold [1 ,2 ]
机构
[1] Med Univ Vienna, Ctr Med Phys & Biomed Engn, Vienna, Austria
[2] Med Univ Vienna, Dept Clin Pharmacol, Vienna, Austria
[3] Vienna Univ Technol, Inst Appl Phys, A-1040 Vienna, Austria
关键词
optical coherence tomography; tear film thickness; dry eyes; EYE WORKSHOP 2007; INTERNATIONAL WORKSHOP; PRECORNEAL; SUBCOMMITTEE; DISEASE; PRELENS; INDEX;
D O I
10.1167/iovs.13-11920
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To visualize the precorneal tear film with ultrahigh-resolution spectral domain optical coherence tomography, enabling quantification of tear film thickness in healthy subjects. METHODS. A custom-built spectral domain optical coherence tomography system comprising a broadband titanium: sapphire laser operating at 800 nm and a high-speed charge coupled device (CCD) camera with a read-out rate of 47 kHz was used for measurement of precorneal tear film thickness. The system provides a theoretical axial resolution of 1.2 mu m in tissue. The signal-to-noise ratio close to the zero delay was measured with 94 dB. A total of 26 healthy subjects were included in this study. Measurement was started immediately after blinking and averaged over a period of 1 second. In a subset of eight healthy subjects, the reproducibility of the approach was studied by measuring the tear film thickness every 10 minutes over 1 hour. RESULTS. The average central tear film thickness of the measured population was 4.79 +/- 0.88 mu m. Reproducibility was very high, with an intraclass correlation coefficient of 0.97. A breakup of the tear film was observed in one subject after 14 seconds. CONCLUSIONS. Our data indicate that the human precorneal tear film can be measured with excellent reproducibility using ultrahigh-resolution optical coherence tomography. This technique may be a valuable tool in the management of dry eye syndrome. (ClinicalTrials.gov number, NCT01746602.)
引用
收藏
页码:5578 / 5583
页数:6
相关论文
共 50 条
  • [1] Ultrahigh-Resolution Measurement by Optical Coherence Tomography of Dynamic Tear Film Changes on Contact Lenses
    Chen, Qi
    Wang, Jianhua
    Tao, Aizhu
    Shen, Meixiao
    Jiao, Shuliang
    Lu, Fan
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (04) : 1988 - 1993
  • [2] Effect of different lubricant eye gels on tear film thickness as measured with ultrahigh-resolution optical coherence tomography
    Wozniak, Piotr A.
    Schmidl, Doreen
    Bata, Ahmed M.
    Fondi, Klemens
    Witkowska, Katarzyna J.
    dos Santos, Valentin Aranha
    Baar, Carina
    Room, Kim I.
    Nepp, Johannes
    Baumgartner, Isabella
    Popa-Cherecheanu, Alina
    Garhoefer, Gerhard
    Werkmeister, Rene M.
    Schmetterer, Leopold
    [J]. ACTA OPHTHALMOLOGICA, 2017, 95 (04) : E307 - E313
  • [3] Automatic assessment of tear film and tear meniscus parameters in healthy subjects using ultrahigh-resolution optical coherence tomography
    Stegmann, Hannes
    dos Santos, Valentin Aranha
    Messner, Alina
    Unterhuber, Angelika
    Schmidl, Doreen
    Garhoefer, Gerhard
    Schmetterer, Leopold
    Werkmeister, Rene Marcel
    [J]. BIOMEDICAL OPTICS EXPRESS, 2019, 10 (06) : 2744 - 2756
  • [4] Ultrahigh-resolution optical coherence tomography
    Drexler, W
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2004, 9 (01) : 47 - 74
  • [5] In vivo ultrahigh-resolution optical coherence tomography
    Drexler, W
    Morgner, U
    Kärtner, FX
    Pitris, C
    Boppart, SA
    Li, XD
    Ippen, EP
    Fujimoto, JG
    [J]. OPTICS LETTERS, 1999, 24 (17) : 1221 - 1223
  • [6] Ultrahigh-resolution optical coherence tomography in glaucoma
    Wollstein, G
    Paunescu, LA
    Ko, TH
    Fujimoto, JG
    Kowalevicz, A
    Hartl, I
    Beaton, S
    Ishikawa, H
    Mattox, C
    Singh, O
    Duker, J
    Drexler, W
    Schuman, JS
    [J]. OPHTHALMOLOGY, 2005, 112 (02) : 229 - 237
  • [7] Ultrahigh-resolution ophthalmic optical coherence tomography
    Drexler, W
    Morgner, U
    Ghanta, RK
    Kärtner, FX
    Schuman, JS
    Fujimoto, JG
    [J]. NATURE MEDICINE, 2001, 7 (04) : 502 - 507
  • [8] Ultrahigh-resolution ophthalmic optical coherence tomography
    Wolfgang Drexler
    Uwe Morgner
    Ravi K. Ghanta
    Franz X. Kärtner
    Joel S. Schuman
    James G. Fujimoto
    [J]. Nature Medicine, 2001, 7 : 502 - 507
  • [9] Retinal oximetry using ultrahigh-resolution optical coherence tomography
    Ye, Yufeng
    Jiang, Hong
    Shen, Meixiao
    Lam, Byron L.
    DeBuc, Delia Cabrera
    Ge, Lili
    Sehi, Mitra
    Wang, Jianhua
    [J]. CLINICAL OPHTHALMOLOGY, 2012, 6 : 2085 - 2092