Thermoelastic femoral stress imaging for experimental evaluation of hip prosthesis design

被引:9
|
作者
Hyodo, K
Inomoto, M
Ma, W
Miyakawa, S
Tateishi, T
机构
[1] Natl Inst Adv Ind Sci & Technol, AIST, Inst Human Sci & Biomed Engn, Biomed Sensing & Imaging Grp, Tsukuba, Ibaraki 3058564, Japan
[2] Univ Tsukuba, Inst Clin Med, Tsukuba, Ibaraki 3058577, Japan
[3] Univ Tokyo, Grad Sch Engn, Bunkyo Ku, Tokyo 1138654, Japan
关键词
biomechanics; muscle and skeleton; optical measurement; thermoelastic stress analysis; artificial hip joint; synthetic femur; stress distribution; mechanical biocompatibility;
D O I
10.1299/jsmec.44.1065
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
An experimental system using the thermoelastic stress analysis method and a synthetic femur was utilized to perform reliable and convenient mechanical biocompatibility evaluation of hip prosthesis design. Unlike the conventional technique, the unique advantage of the thermoelastic stress analysis method is its ability to image whole-surface stress (Delta(sigma(1) + sigma(2))) distribution in specimens. The mechanical properties of synthetic femurs agreed well with those of cadaveric femurs with little variability between specimens. We applied this experimental system for stress distribution visualization of the intact femur, and the femurs implanted with an artificial joint. The surface stress distribution of the femurs sensitively, reflected the prosthesis design and the contact condition between the stem and the bone. By analyzing the relationship between the stress distribution and the clinical results of the artificial joint, this technique can be used in mechanical biocompatibility evaluation and pre-clinical performance prediction of new artificial joint design.
引用
收藏
页码:1065 / 1071
页数:7
相关论文
共 50 条
  • [21] Application of thermoelastic stress analysis for the experimental evaluation of the effective stress intensity factor
    Diaz, Francisco A.
    Patterson, Eann A.
    Yates, John R.
    FRATTURA ED INTEGRITA STRUTTURALE, 2013, (25): : 109 - 116
  • [22] 3-DIMENSIONAL STRESS-ANALYSIS OF THE FEMORAL STEM OF A TOTAL HIP-PROSTHESIS
    HAMPTON, SJ
    ANDRIACCHI, TP
    GALANTE, JO
    JOURNAL OF BIOMECHANICS, 1980, 13 (05) : 443 - 448
  • [23] Reducing Contact Stress of the Surface by Modifying Different Hardness of Femoral Head and Cup in Hip Prosthesis
    Saputra, Eko
    Anwar, Iwan Budiwan
    Jamari, J.
    van der Heide, Emile
    FRONTIERS IN MECHANICAL ENGINEERING-SWITZERLAND, 2021, 7
  • [24] Femoral Stress Changes after Total Hip Arthroplasty with the Ribbed Prosthesis: A Finite Element Analysis
    Luo, Changqi
    Wu, Xiang-Dong
    Wan, Yifei
    Liao, Junyi
    Cheng, Qiang
    Tian, Mian
    Bai, Zhibiao
    Huang, Wei
    BIOMED RESEARCH INTERNATIONAL, 2020, 2020
  • [25] FEMORAL-SHAFT FRACTURE BELOW HIP PROSTHESIS
    AUFRANC, OE
    JONES, WN
    TURNER, RH
    HARRIS, WH
    WYMAN, ET
    JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1965, 194 (13): : 1378 - &
  • [26] Metallurgical aspects of failure in a broken femoral HIP prosthesis
    Khouzani, M. Kiani
    Bahrami, A.
    Eslami, A.
    ENGINEERING FAILURE ANALYSIS, 2018, 90 : 168 - 178
  • [27] A novel technique for the explantation of a cemented femoral hip prosthesis
    Hillier, D.
    Kazi, H. A.
    ANNALS OF THE ROYAL COLLEGE OF SURGEONS OF ENGLAND, 2017, 99 (02) : 176 - 177
  • [28] Intrapelvic dislocation of trial femoral prosthesis in hip arthroplasty
    Sivakumar, Brahman Shankar
    Suzuki, Arnold
    Lorentzos, Peter
    Suthersan, Mayuran
    Dave, Jaykar
    ANZ JOURNAL OF SURGERY, 2018, 88 (1-2) : E97 - E98
  • [29] ABOUT DAMAGE OF THE FEMORAL HEAD OF TOTAL HIP PROSTHESIS
    Capitanu, L.
    Badita, L. L.
    Bursuc, D. C.
    JOURNAL OF THE BALKAN TRIBOLOGICAL ASSOCIATION, 2012, 18 (04): : 642 - 653
  • [30] Failure of the uncoated titanium ProxiLock femoral hip prosthesis
    Stoller, Craig
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (455) : 283 - 283