Monte Carlo calculation of the efficiency response of a low-background well-type HPGe detector

被引:34
|
作者
Laborie, JM [1 ]
Le Petit, G [1 ]
Abt, D [1 ]
Girard, M [1 ]
机构
[1] CEA, DAM, Dept Analyse & Surveillance Environm, F-91680 Bruyeres Le Chatel, France
关键词
well-type HPGe detectors; gamma-ray spectrometry; efficiency calibration; Monte Carlo simulation; true coincidence-summing;
D O I
10.1016/S0168-9002(01)00942-1
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Well-type HPGe detectors are well suited for the analysis of small amounts of environmental samples, as they can combine both low background and high detection efficiency. A low-background well-type detector is installed in the Modane underground Laboratory. In the well geometry, coincidence-summing effects are high and make construction of the full energy peak efficiency curve difficult with the usual calibration standard, especially in the: high energy range. Using the GEANT code, taking into account a detailed description of the detector and the source, efficiency curves have been modelled for several filling heights of the vial. With a special routine, taking into account the decay schemes of the radionuclides corrections have been computed for coincidence-summing effects that occur when measuring samples containing U-238, Th-232 or Cs-134. The results are found to be in good agreement with experimental data. It is shown that triple coincidences effect on counting losses accounts for 7-15% of pair coincidences effect in the case of the 604 and 796 keV lines of Cs-134. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:618 / 630
页数:13
相关论文
共 50 条
  • [1] EFFICIENCY CALIBRATION OF A WELL-TYPE HPGe DETECTOR USING EXPERIMENTAL AND MONTE CARLO SIMULATION TECHNIQUES
    Dalaka, Ekaterini
    Kuburas, Georgios
    Eleftheriadis, Konstantinos
    Anagnostakis, Marios J.
    NUCLEAR TECHNOLOGY & RADIATION PROTECTION, 2020, 35 (02): : 121 - 129
  • [2] Derivation of an efficiency-calibration simulation for a well-type HPGe detector using the Monte Carlo approach and analytical techniques
    Belgin, E. Eren
    Aycik, G. A.
    RADIATION MEASUREMENTS, 2015, 73 : 36 - 45
  • [3] Simple Monte-Carlo method to calibrate well-type HPGe detectors
    Bochud, Francois
    Bailat, Claude J.
    Buchillier, Thierry
    Byrde, Francois
    Schmid, Ernst
    Laedermann, Jean-Pascal
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 569 (03): : 790 - 795
  • [4] Monte Carlo based geometrical model for efficiency calculation of an n-type HPGe detector
    Padilla Cabal, Fatima
    Lopez-Pino, Neivy
    Luis Bernal-Castillo, Jose
    Martinez-Palenzuela, Yisel
    Aguilar-Mena, Jimmy
    D'Alessandro, Katia
    Arbelo, Yuniesky
    Corrales, Yasser
    Diaz, Oscar
    APPLIED RADIATION AND ISOTOPES, 2010, 68 (12) : 2403 - 2408
  • [5] Development of a low-background HPGe detector at Kamioka Observatory
    Ichimura, K.
    Ikeda, H.
    Kishimoto, Y.
    Kurasawa, M.
    Suzuki, A. A.
    Gando, Y.
    Ikeda, M.
    Hosokawa, K.
    Sekiya, H.
    Ito, H.
    Minamino, A.
    Suzuki, S.
    PROGRESS OF THEORETICAL AND EXPERIMENTAL PHYSICS, 2023, 2023 (12):
  • [6] LARGE, LOW-BACKGROUND WELL-TYPE DETECTORS FOR MEASUREMENTS OF ENVIRONMENTAL-RADIOACTIVITY
    REYSS, JL
    SCHMIDT, S
    LEGELEUX, F
    BONTE, P
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 357 (2-3): : 391 - 397
  • [7] Monte Carlo calculation of the efficiency calibration curve and coincidence-summing corrections in low-level gamma-ray spectrometry using well-type HPGe detectors
    Laborie, JM
    Le Petit, G
    Abt, D
    Girard, M
    APPLIED RADIATION AND ISOTOPES, 2000, 53 (1-2) : 57 - 62
  • [8] Experimental and Monte Carlo determination of HPGe detector efficiency
    Jeskovsky, Miroslav
    Javornik, Andrej
    Breier, Robert
    Sluciak, Jarmila
    Povinec, Pavel P.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2019, 322 (03) : 1863 - 1869
  • [9] Experimental and Monte Carlo determination of HPGe detector efficiency
    Miroslav Ješkovský
    Andrej Javorník
    Róbert Breier
    Jarmila Slučiak
    Pavel P. Povinec
    Journal of Radioanalytical and Nuclear Chemistry, 2019, 322 : 1863 - 1869
  • [10] Experimental HPGe coaxial detector response and efficiency compared to Monte Carlo simulations
    Maidana, Nora L.
    Vanin, Vito R.
    Garcia-Alvarez, Juan A.
    Hermida-Lopez, Marcelino
    Brualla, Lorenzo
    APPLIED RADIATION AND ISOTOPES, 2016, 108 : 64 - 74