Simulated Design Strategies for SPECT Collimators to Reduce the Eddy Currents Induced by MRI Gradient Fields

被引:6
|
作者
Samoudi, Amine M. [1 ]
Van Audenhaege, Karen
Vermeeren, Gunter
Verhoyen, Gregory
Martens, Luc
Van Holen, Roel
Joseph, Wout
机构
[1] Ghent Univ iMinds, Dept Informat Technol INTEC, B-9000 Ghent, Belgium
关键词
Collimator; eddy currents; final design; MRI; SPECT; Tungsten; SYSTEM; COMPENSATION; PERFORMANCE; INSERT; COILS;
D O I
10.1109/TNS.2015.2476601
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Combining single photon emission computed tomography (SPECT) with magnetic resonance imaging (MRI) requires the insertion of highly conductive SPECT collimators inside the MRI scanner, resulting in an induced eddy current disturbing the combined system. We reduced the eddy currents due to the insert of a novel tungsten collimator inside transverse and longitudinal gradient coils. The collimator was produced with metal additive manufacturing, that is part of a microSPECT insert for a preclinical SPECT/MRI scanner. We characterized the induced magnetic field due to the gradient field and adapted the collimators to reduce the induced eddy currents. We modeled the x-, y-, and z-gradient coil and the different collimator designs and simulated them with FEKO, a three-dimensional method of moments / finite element methods (MoM/FEM) full-wave simulation tool. We used a time analysis approach to generate the pulsed magnetic field gradient. Simulation results show that the maximum induced field can be reduced by 50.82% in the final design bringing the maximum induced magnetic field to less than 2% of the applied gradient for all the gradient coils. The numerical model was validated with measurements and was proposed as a tool for studying the effect of a SPECT collimator within the MRI gradient coils.
引用
收藏
页码:2017 / 2022
页数:6
相关论文
共 50 条
  • [31] A combinatorial method based on simulated annealing to design a biplanar MRI gradient system
    Grau-Ruiz, D.
    Sanchez, H.
    Rigla, J. P.
    Diaz-Caballero, E.
    Gonzalez, J. M.
    Puchalt, G.
    Aguilar, A.
    Gonzalez-Montoro, A.
    Gonzalez, A. J.
    Sanchez, S.
    Benlloch, J. M.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [32] Multilayer Integral Method for Simulation of Eddy Currents in Thin Volumes of Arbitrary Geometry Produced by MRI Gradient Coils
    Lopez, Hector Sanchez
    Freschi, Fabio
    Trakic, Adnan
    Smith, Elliot
    Herbert, Jeremy
    Fuentes, Miguel
    Wilson, Stephen
    Liu, Limei
    Repetto, Maurizio
    Crozier, Stuart
    MAGNETIC RESONANCE IN MEDICINE, 2014, 71 (05) : 1912 - 1922
  • [33] Mitigation of Intra-coil Eddy Currents in Split Gradient Coils in a Hybrid MRI-LINAC System
    Tang, Fangfang
    Freschi, Fabio
    Repetto, Maurizio
    Li, Yu
    Liu, Feng
    Crozier, Stuart
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2017, 64 (03) : 725 - 732
  • [34] Induced electric fields and currents in the body by movements in a MRI facility: a numerical analysis
    Romeo, Stefania
    Sannino, Anna
    Zeni, Olga
    Scarfi, Maria Rosaria
    Massa, Rita
    Cerciello, Vincenzo
    d'Angelo, Raffaele
    2015 1st URSI Atlantic Radio Science Conference (URSI AT-RASC), 2015,
  • [35] Finite-difference time-domain-based studies of MRI pulsed field gradient-induced eddy currents inside the human body
    Liu, F
    Crozier, S
    Zhao, HW
    Lawrence, B
    CONCEPTS IN MAGNETIC RESONANCE, 2002, 15 (01): : 26 - 36
  • [36] Simulation study of the potential hazards of cardiac stimulation by induced eddy-currents-in modern MRI
    Xia, L
    Liu, F
    Zhao, H
    COMPUTERS IN CARDIOLOGY 2002, VOL 29, 2002, 29 : 725 - 728
  • [37] DYNAMICS OF DOMAIN-STRUCTURE OF FERROMAGNETS AND MICRO-EDDY CURRENTS INDUCED BY MICROWAVE FIELDS
    MINAKOV, VI
    FEDOSOV, VN
    IZVESTIYA VYSSHIKH UCHEBNYKH ZAVEDENII FIZIKA, 1977, (08): : 95 - 99
  • [38] Three-dimensional optimal shape design of magnetic pole in permanent magnet assembly for MRI taking account of Eddy currents due to gradient coil field
    Yao, YY
    Koh, CS
    Xie, DX
    IEEE TRANSACTIONS ON MAGNETICS, 2004, 40 (02) : 1164 - 1167
  • [39] Heating of metallic rods induced by time-varying gradient fields in MRI
    El Bannan, Khaled
    Handler, William
    Chronik, Blaine
    Salisbury, Shaun P.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2013, 38 (02) : 411 - 416
  • [40] Segmented RF shield design to minimize eddy currents for low-field Halbach MRI systems
    de Vos, Bart
    Remis, Rob
    Webb, Andrew
    JOURNAL OF MAGNETIC RESONANCE, 2024, 362