Ex vivo investigations of laser auricular cartilage reshaping with carbon dioxide spray cooling in a rabbit model

被引:5
|
作者
Wu, Edward C. [1 ]
Sun, Victor [1 ]
Manuel, Cyrus T. [1 ]
Protsenko, Dmitriy E. [1 ]
Jia, Wangcun [1 ]
Nelson, J. Stuart [1 ]
Wong, Brian J. F. [1 ]
机构
[1] Univ Calif Irvine, Beckman Laser Inst & Med Clin, Irvine, CA 92612 USA
基金
美国国家卫生研究院;
关键词
Facial plastic surgery; Macrotia; Carbon dioxide spray; Laser cartilage reshaping; NASAL SEPTAL CARTILAGE; PORT-WINE STAIN; ER-YAG LASER; MECHANICAL-BEHAVIOR; STRESS-RELAXATION; ELASTIC-MODULUS; CRYOGEN; PORCINE; IRRADIATION; RESPONSES;
D O I
10.1007/s10103-012-1250-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Laser cartilage reshaping (LCR) with cryogen spray cooling is a promising modality for producing cartilage shape change while reducing cutaneous thermal injury. However, LCR in thicker tissues, such as auricular cartilage, requires higher laser power, thus increasing cooling requirements. To eliminate the risks of freeze injury characteristic of high cryogen spray pulse rates, a carbon dioxide (CO2) spray, which evaporates rapidly from the skin, has been proposed as the cooling medium. This study aims to identify parameter sets which produce clinically significant reshaping while producing minimal skin thermal injury in LCR with CO2 spray cooling in ex vivo rabbit auricular cartilage. Excised whole rabbit ears were mechanically deformed around a cylindrical jig and irradiated with a 1.45-mu m wavelength diode laser (fluence 12-14 J/cm(2) per pulse, four to six pulse cycles per irradiation site, five to six irradiation sites per row for four rows on each sample) with concomitant application of CO2 spray (pulse duration 33-85 ms) to the skin surface. Bend angle measurements were performed before and after irradiation, and the change quantified. Surface temperature distributions were measured during irradiation/cooling. Maximum skin surface temperature ranged between 49.0 to 97.6 A degrees C following four heating/cooling cycles. Significant reshaping was achieved with all laser dosimetry values with a 50-70 A degrees C difference noted between controls (no cooling) and irradiated ears. Increasing cooling pulse duration yielded progressively improved gross skin protection during irradiation. CO2 spray cooling may potentially serve as an alternative to traditional cryogen spray cooling in LCR and may be the preferred cooling medium for thicker tissues. Future studies evaluating preclinical efficacy in an in vivo rabbit model are in progress.
引用
收藏
页码:1475 / 1482
页数:8
相关论文
共 50 条
  • [1] Ex vivo investigations of laser auricular cartilage reshaping with carbon dioxide spray cooling in a rabbit model
    Edward C. Wu
    Victor Sun
    Cyrus T. Manuel
    Dmitriy E. Protsenko
    Wangcun Jia
    J. Stuart Nelson
    Brian J. F. Wong
    Lasers in Medical Science, 2013, 28 : 1475 - 1482
  • [2] SURFACE TEMPERATURE DISTRIBUTIONS IN RABBIT AURICULAR CARTILAGE FOLLOWING LASER CARTILAGE RESHAPING WITH CARBON DIOXIDE SPRAY COOLING
    Wu, Edward C.
    Sun, Victor
    Jia, Wangcun
    Protsenko, Dmitriy E.
    Manuel, Cyrus T.
    Wong, Brian J. F.
    Nelson, J. Stuart
    LASERS IN SURGERY AND MEDICINE, 2011, 43 : 962 - 962
  • [3] IN VIVO LASER CARTILAGE RESHAPING WITH CARBON DIOXIDE SPRAY COOLING IN A RABBIT EAR MODEL: A PILOT STUDY
    Wu, Edward
    Hamamoto, Ashley
    Sun, Victor
    Manuel, Cyrus
    Protsenko, Dmitriy
    Jia, Wangcun
    Wong, Brian
    Nelson, J. Stuart
    LASERS IN SURGERY AND MEDICINE, 2013, 45 : 68 - 68
  • [4] In Vivo Laser Cartilage Reshaping With Carbon Dioxide Spray Cooling in a Rabbit Ear Model: A Pilot Study
    Kuan, Edward C.
    Hamamoto, Ashley A.
    Sun, Victor
    Nguyen, Tony
    Manuel, Cyrus T.
    Protsenko, Dmitry E.
    Wong, Brian J. F.
    Nelson, J. Stuart
    Jia, Wangcun
    LASERS IN SURGERY AND MEDICINE, 2014, 46 (10) : 791 - 795
  • [5] Preliminary investigations of laser mediated cartilage reshaping in the in vivo rabbit model
    Chang, JC
    Diaz, S
    Wong, BJF
    LASERS IN SURGERY: ADVANCED CHARACTERIZATION, THERAPEUTICS, AND SYSTEMS XII, 2002, 4609 : 326 - 330
  • [6] Ex Vivo Electromechanical Reshaping of Costal Cartilage in the New Zealand White Rabbit Model
    Badran, Karam
    Manuel, Cyrus
    Waki, Curtis
    Protsenko, Dmitry
    Wong, Brian J. F.
    LARYNGOSCOPE, 2013, 123 (05): : 1143 - 1148
  • [7] Endoscopic laser reshaping of rabbit tracheal cartilage: Preliminary investigations
    Tsang, W
    Lam, A
    Protsenko, D
    Wong, B
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS, 2005, 5686 : 345 - 352
  • [8] Laser cartilage reshaping in an in vivo rabbit model using a 1.54 μm Er:Glass laser
    Mordon, S
    Wang, T
    Fleurisse, L
    Creusy, C
    LASERS IN SURGERY AND MEDICINE, 2004, 34 (04) : 315 - 322
  • [9] Preliminary investigations on therapy thresholds for laser dosimetry, cryogen spray cooling duration, and treatment cycles for laser cartilage reshaping in the New Zealand white rabbit auricle
    Chlebicki, Cara A.
    Protsenko, Dmitry E.
    Wong, Brian J.
    LASERS IN MEDICAL SCIENCE, 2014, 29 (03) : 1099 - 1109
  • [10] Preliminary investigations on therapy thresholds for laser dosimetry, cryogen spray cooling duration, and treatment cycles for laser cartilage reshaping in the New Zealand white rabbit auricle
    Cara A. Chlebicki
    Dmitry E. Protsenko
    Brian J. Wong
    Lasers in Medical Science, 2014, 29 : 1099 - 1109