Technical Note: Algebraic iterative image reconstruction using a cylindrical image grid for tetrahedron beam computed tomography

被引:1
|
作者
Kim, Joshua [1 ,2 ]
Ionascu, Dan [3 ]
Zhang, Tiezhi [2 ,3 ]
机构
[1] Oakland Univ, Dept Phys, Rochester, MI 48309 USA
[2] TetraImaging Inc, Troy, MI 48098 USA
[3] William Beaumont Hosp, Dept Radiat Oncol, Royal Oak, MI 48073 USA
关键词
CBCT; TBCT; image reconstruction; SART; TECHNIQUE SART; SYSTEM; ART;
D O I
10.1118/1.4812886
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To accelerate iterative algebraic reconstruction algorithms using a cylindrical image grid. Methods: Tetrahedron beam computed tomography (TBCT) is designed to overcome the scatter and detector problems of cone beam computed tomography (CBCT). Iterative algebraic reconstruction algorithms have been shown to mitigate approximate reconstruction artifacts that appear at large cone angles, but clinical implementation is limited by their high computational cost. In this study, a cylindrical voxelization method on a cylindrical grid is developed in order to take advantage of the symmetries of the cylindrical geometry. The cylindrical geometry is a natural fit for the circular scanning trajectory employed in volumetric CT methods such as CBCT and TBCT. This method was implemented in combination with the simultaneous algebraic reconstruction technique (SART). Both two- and three-dimensional numerical phantoms as well as a patient CT image were utilized to generate the projection sets used for reconstruction. The reconstructed images were compared to the original phantoms using a set of three figures of merit (FOM). Results: The cylindrical voxelization on a cylindrical reconstruction grid was successfully implemented in combination with the SART reconstruction algorithm. The FOM results showed that the cylindrical reconstructions were able to maintain the accuracy of the Cartesian reconstructions. In three dimensions, the cylindrical method provided better accuracy than the Cartesian methods. At the same time, the cylindrical method was able to provide a speedup factor of approximately 40 while also reducing the system matrix storage size by 2 orders of magnitude. Conclusions: TBCT image reconstruction using a cylindrical image grid was able to provide a significant improvement in the reconstruction time and a more compact system matrix for storage on the hard drive and in memory while maintaining the image quality provided by the Cartesian voxelization on a Cartesian grid. (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Evaluation of Algebraic Iterative Image Reconstruction Methods for Tetrahedron Beam Computed Tomography Systems
    Kim, Joshua
    Guan, Huaiqun
    Gersten, David
    Zhang, Tiezhi
    INTERNATIONAL JOURNAL OF BIOMEDICAL IMAGING, 2013, 2013 (2013)
  • [2] Image reconstruction using an adaptive simultaneous algebraic reconstruction technique in computed tomography
    Qiao Q.
    Huang L.
    He Z.
    Huang, Liyu (huangly@mail.xidian.end.cn), 2016, Science Press (43): : 67 - 72
  • [3] Efficient algebraic image reconstruction technique for computed tomography
    Hanna, R.
    Sutcliffe, M.
    Charlton, P.
    Mosey, S.
    INSIGHT, 2022, 64 (06) : 326 - 333
  • [4] Computed Tomography Image Reconstruction with a Cylindrical Voxelization Method
    Kim, J.
    Ionascu, D.
    Zhang, T.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [5] Volume integral model for algebraic image reconstruction and computed tomography
    Hanna, R.
    Sutcliffe, M.
    Carswell, D.
    Charlton, P.
    Mosey, S.
    INSIGHT, 2023, 65 (09) : 484 - 491
  • [6] Iterative Image Reconstruction Techniques Cardiothoracic Computed Tomography Applications
    Cho, Young Jun
    Schoepf, U. Joseph
    Silverman, Justin R.
    Krazinski, Aleksander W.
    Canstein, Christian
    Deak, Zsuzsanna
    Grimm, Jochen
    Geyer, Lucas L.
    JOURNAL OF THORACIC IMAGING, 2014, 29 (04) : 198 - 208
  • [7] Bifurcation analysis of iterative image reconstruction method for computed tomography
    Yoshinaga, Tetsuya
    Imakura, Yoshihiro
    Fujimoto, Ken'ichi
    Ueta, Tetsushi
    INTERNATIONAL JOURNAL OF BIFURCATION AND CHAOS, 2008, 18 (04): : 1219 - 1225
  • [8] Iterative Image Reconstruction and Its Role in Cardiothoracic Computed Tomography
    Singh, Sarabjeet
    Khawaja, Ranish Deedar Ali
    Pourjabbar, Sarvenaz
    Padole, Atul
    Lira, Diego
    Kalra, Mannudeep K.
    JOURNAL OF THORACIC IMAGING, 2013, 28 (06) : 355 - 367
  • [9] Proton Computed Tomography: iterative image reconstruction and dose evaluation
    Civinini, C.
    Bonanno, D.
    Brianzi, M.
    Carpinelli, M.
    Cirrone, G. A. P.
    Cuttone, G.
    Lo Presti, D.
    Maccioni, G.
    Pallotta, S.
    Randazzo, N.
    Scaringella, M.
    Romano, F.
    Sipala, V.
    Talamonti, C.
    Vanzi, E.
    Bruzzi, M.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [10] Computed Tomography Image Reconstruction
    Seeram, Euclid
    RADIOLOGIC TECHNOLOGY, 2020, 92 (02) : 155CT - 171CT