Long-term wear of dual mobility total hip replacement cups: explant study

被引:29
|
作者
Boyer, Bertrand [1 ,2 ]
Neri, Thomas [2 ]
Geringer, Jean [1 ]
Di Iorio, Alexandre [2 ]
Philippot, Remi [2 ]
Farizon, Frederic [2 ]
机构
[1] Univ Lyon St Etienne, INSERM U1059, Unite SAINBIOSE, Mines St Etienne, Lyon, France
[2] Univ Jean Monnet, CHU St Etienne, Ctr Orthopedie Traumatol, Lab Physiol Exercice, F-42055 St Etienne 2, France
关键词
Dual mobility; Total hip arthroplasty; Retrieval; Wear; Long-term; FOLLOW-UP; ARTHROPLASTY; PROSTHESIS; DISLOCATION; SOCKET;
D O I
10.1007/s00264-017-3525-z
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose Dual mobility THA implantations are increasing. THA survival of more than 15 years is considered long-term survival based on the Charnley prosthesis, which has a median wear rate of 39 to 98 mm(3)/year. What is the median wear rate of dual mobility THA cups and its repartition on the liner volume? What are the factors influencing dual mobility wear? Methods Thirty-five explants were analysed. The liners were entirely mapped with a fully automatic stripe light scanner that uses green LED light. Results The liners had a median wear rate of 38 mm(3)/year. The distribution of the wear between the two bearings varied greatly. No factor was found influencing wear volumes. Conclusions The median annual wear of dual mobility liners is at least as low as the wear of cemented polyethylene liners and lower than equivalent cementless liners. The two articulations of the dual mobility THA do not cause more wear.
引用
下载
收藏
页码:41 / 47
页数:7
相关论文
共 50 条
  • [1] Long-term wear of dual mobility total hip replacement cups: explant study
    Bertrand Boyer
    Thomas Neri
    Jean Geringer
    Alexandre Di Iorio
    Remi Philippot
    Frederic Farizon
    International Orthopaedics, 2018, 42 : 41 - 47
  • [2] Understanding wear in dual mobility total hip replacement: first generation explant wear patterns
    Boyer, Bertrand
    Neri, Thomas
    Geringer, Jean
    Di Iorio, Alexandre
    Philippot, Remi
    Farizon, Frederic
    INTERNATIONAL ORTHOPAEDICS, 2017, 41 (03) : 529 - 533
  • [3] Understanding wear in dual mobility total hip replacement: first generation explant wear patterns
    Bertrand Boyer
    Thomas Neri
    Jean Geringer
    Alexandre Di Iorio
    Remi Philippot
    Frederic Farizon
    International Orthopaedics, 2017, 41 : 529 - 533
  • [4] Dual mobility cups in total hip arthroplasty
    De Martino, Ivan
    Triantafyllopoulos, Georgios Konstantinos
    Sculco, Peter Keyes
    Sculco, Thomas Peter
    WORLD JOURNAL OF ORTHOPEDICS, 2014, 5 (03): : 180 - 187
  • [5] The linear penetration rate is not relevant for evaluating wear of dual mobility cups: an explant study
    Bertrand Boyer
    Thomas Neri
    Alexandre Di Iorio
    Jean Geringer
    Remi Philippot
    Frederic Farizon
    International Orthopaedics, 2017, 41 : 599 - 603
  • [6] The linear penetration rate is not relevant for evaluating wear of dual mobility cups: an explant study
    Boyer, Bertrand
    Neri, Thomas
    Di Iorio, Alexandre
    Geringer, Jean
    Philippot, Remi
    Farizon, Frederic
    INTERNATIONAL ORTHOPAEDICS, 2017, 41 (03) : 599 - 603
  • [7] WEAR BEHAVIOR OF CAST COCRMO CUPS AND BALLS IN LONG-TERM IMPLANTED TOTAL HIP PROSTHESES
    SEMLITSCH, M
    STREICHER, RM
    WEBER, H
    ORTHOPADE, 1989, 18 (05): : 377 - 381
  • [8] Dual mobility cups in revision total hip arthroplasty
    Viste, Anthony
    Desmarchelier, Romain
    Fessy, Michel-Henri
    INTERNATIONAL ORTHOPAEDICS, 2017, 41 (03) : 535 - 542
  • [9] Dual mobility cups in revision total hip arthroplasty
    Anthony Viste
    Romain Desmarchelier
    Michel-Henri Fessy
    International Orthopaedics, 2017, 41 : 535 - 542
  • [10] Long-term study of total hip replacement in rheumatoid arthritis
    Wessinghage, D
    Kisslinger, E
    AKTUELLE RHEUMATOLOGIE, 1996, 21 (05) : 242 - 248