Ion mobility spectrometry-mass spectrometry studies of ion processes in air at atmospheric pressure and their application to thermal desorption of 2,4,6-trinitrotoluene

被引:10
|
作者
Sabo, Martin [1 ]
Malaskova, Michaela [1 ]
Matejcik, Stefan [1 ]
机构
[1] Comenius Univ, Fac Math Phys & Informat, Dept Expt Phys, Bratislava 84248, Slovakia
来源
PLASMA SOURCES SCIENCE & TECHNOLOGY | 2014年 / 23卷 / 01期
关键词
corona discharge; plasma ionization; negative ions; ion mobility spectrometry; mass spectrometry; surface analysis; TNT detection; CORONA DISCHARGE IONIZATION; VOLATILE ORGANIC-COMPOUNDS; REACTANT IONS; PLASMA; GAS; DESIGN;
D O I
10.1088/0963-0252/23/1/015025
中图分类号
O35 [流体力学]; O53 [等离子体物理学];
学科分类号
070204 ; 080103 ; 080704 ;
摘要
In this study we have investigated the negative reactant ion formation in a negative corona discharge (CD) using the corona discharge ion mobility spectrometry orthogonal acceleration time-of-flight (CD-IMS-oaTOF) technique. The reactant ions were formed in the CD operating in the reverse gas flow mode at an elevated temperature of 363.5K in synthetic and ambient air. Under these conditions mainly O-2(-) and their clusters were formed. We have also studied the influence of CCl4 admixture to air (dopant gas) on the composition of the reactant ions, which resulted in the formation of Cl- and its clusters with a reduced ion mobility of 3.05 cm(2) V-1 s(-1) as a major reactant ion peak. Additional IMS peaks with reduced ion mobilities of 2.49, 2.25 and 2.03 cm(2) V-1 s(-1) were detected, and Cl- center dot (NO2) and Cl- center dot (NO)(n) (n = 2, 3) anions were identified. The negative reactant ions were used to detect 2,4,6 trinitrotoluene (TNT) using the thermal desorption (TD) technique using a CD-IMS instrument. Using TD sampling and a negative CD ion source doped by CCl4 we have achieved a limit of detection of 350 pg for direct surface analysis of TNT.
引用
收藏
页数:7
相关论文
共 50 条
  • [11] The study of thermal decomposition of RDX by corona discharge-ion mobility spectrometry-mass spectrometry
    Lichvanova, Zuzana
    Sabo, Martin
    Matejcik, Stefan
    INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2015, 18 (1-2) : 59 - 66
  • [12] Atmospheric pressure ionization of chlorinated ethanes in ion mobility spectrometry and mass spectrometry
    Ewing, Robert G.
    Atkinson, David A.
    Benson, Michael T.
    INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2015, 18 (1-2) : 51 - 58
  • [13] Atmospheric pressure field desorption-trapped ion mobility-mass spectrometry coupling
    Gross, Juergen H.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2024, 416 (14) : 3313 - 3323
  • [14] Deciphering drift time measurements from travelling wave ion mobility spectrometry-mass spectrometry studies
    Smith, David P.
    Knapman, Tom W.
    Campuzano, Iain
    Malham, Richard W.
    Berryman, Joshua T.
    Radford, Sheena E.
    Ashcroft, Atison E.
    EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2009, 15 (02) : 113 - 130
  • [15] REACTIONS IN THE MASS-SPECTROMETRY OF A HYDRIDE MEISENHEIMER COMPLEX OF 2,4,6-TRINITROTOLUENE (TNT)
    YINON, J
    JOHNSON, JV
    BERNIER, UR
    YOST, RA
    MAYFIELD, HT
    MAHONE, WC
    VORBECK, C
    JOURNAL OF MASS SPECTROMETRY, 1995, 30 (05): : 715 - 722
  • [16] Atmospheric pressure chemical ionization of fluorinated phenols in atmospheric pressure chemical ionization mass spectrometry, tandem mass spectrometry, and ion mobility spectrometry
    Eiceman, GA
    Bergloff, JF
    Rodriguez, JE
    Munro, W
    Karpas, Z
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1999, 10 (11) : 1157 - 1165
  • [17] Ion mobility spectrometry-mass spectrometry (IMS-MS) for on- and offline analysis of atmospheric gas and aerosol species
    Krechmer, Jordan E.
    Groessl, Michael
    Zhang, Xuan
    Junninen, Heikki
    Massoli, Paola
    Lambe, Andrew T.
    Kimmel, Joel R.
    Cubison, Michael J.
    Graf, Stephan
    Lin, Ying-Hsuan
    Budisulistiorini, Sri H.
    Zhang, Haofei
    Surratt, Jason D.
    Knochenmuss, Richard
    Jayne, John T.
    Worsnop, Douglas R.
    Jimenez, Jose-Luis
    Canagaratna, Manjula R.
    ATMOSPHERIC MEASUREMENT TECHNIQUES, 2016, 9 (07) : 3245 - 3262
  • [18] METABOLIC STUDIES OF EXPLOSIVES .1. EI AND CI MASS-SPECTROMETRY OF METABOLITES OF 2,4,6-TRINITROTOLUENE
    YINON, J
    HWANG, DG
    BIOMEDICAL MASS SPECTROMETRY, 1984, 11 (11): : 594 - 600
  • [19] Atmospheric Solid Analysis Probe Coupled to Ion Mobility Spectrometry-Mass Spectrometry, a Fast and Simple Method for Polyalphaolefin Characterization
    Siqueira, Anna Luiza Mendes
    Beaumesnil, Mathieu
    Hubert-Roux, Marie
    Loutelier-Bourhis, Corinne
    Afonso, Carlos
    Bai, Yang
    Courtiade, Marion
    Racaud, Amandine
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2018, 29 (08) : 1678 - 1687
  • [20] Application of Ion Mobility Spectrometry-Mass Spectrometry for Compositional Characterization and Fingerprinting of a Library of Diverse Crude Oil Samples
    Cordova, Alexandra C.
    Dodds, James N.
    Tsai, Han-Hsuan D.
    Lloyd, Dillon T.
    Roman-Hubers, Alina T.
    Wright, Fred A.
    Chiu, Weihsueh A.
    McDonald, Thomas J.
    Zhu, Rui
    Newman, Galen
    Rusyn, Ivan
    ENVIRONMENTAL TOXICOLOGY AND CHEMISTRY, 2023, 42 (11) : 2336 - 2349