Effect of Dosing Cup Size on the Aerosol Performance of High-Dose Carrier-Based Formulations in a Novel Dry Powder Inhaler

被引:3
|
作者
Yeung, Stewart [1 ,2 ]
Traini, Daniela [1 ,2 ]
Tweedie, Alan [3 ]
Lewis, David [3 ]
Church, Tanya [3 ]
Young, Paul M. [1 ,2 ]
机构
[1] Univ Sydney, Fac Med & Hlth, Resp Technol, Woolcock Inst Med Res, Sydney, NSW 2006, Australia
[2] Univ Sydney, Fac Med & Hlth, Discipline Pharmacol, Sydney, NSW 2006, Australia
[3] Chiesi Ltd, Chippenham, Wilts, England
基金
澳大利亚研究理事会;
关键词
dosing cup size; reservoir-based dry powder inhaler; carrier formulation; impaction forces; aerosol performance; COMPUTATIONAL FLUID-DYNAMICS; PART; DRUG; INHALATION; DESIGN; MIXTURES; DETACHMENT; ADHESION; DELIVERY; CAPSULE;
D O I
10.1016/j.xphs.2018.09.033
中图分类号
R914 [药物化学];
学科分类号
100701 ;
摘要
This study investigated how varying the dosing cup size of a novel reservoir dry powder inhaler (DPI) affects the detachment of a micronized active pharmaceutical ingredient from larger carrier particles, and the aerosol performance of a DPI carrier formulation. Three different-sized dosing cups were designed: 3D printed with cup volumes of 16.26 mm(3), 55.99 mm(3), and 133.04 mm(3), and tested with five different carrier type formulations with beclomethasone dipropionate (BDP) concentrations between 1% and 30% (w/w). The morphology of the BDP attached to the carrier was investigated using scanning electron microscopy and the aerosol performance using the Next Generation Impactor. Increasing the volume of the dosing cup led to a reduction of BDP deposition in the Next Generation Impactor preseparator, and an increase in BDP detachment from the carrier was observed, leading to increased aerosol performance. The decreased amount of BDP attached to carrier after aerosolization was attributed to the increased dosing cup void volume. This may enable greater particle-particle and particle-wall collisions, with greater BDP detachment from the carrier and deagglomeration of smaller agglomerates. The dosing cup volume was observed to have significant influence on particle dispersion and the overall aerosol performance of a DPI. (c) 2019 American Pharmacists Association (R). Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:949 / 959
页数:11
相关论文
共 50 条
  • [21] COMPARING THE AEROSOL IN VITRO PERFORMANCE AND SURFACE ENERGETICS OF DRY POWDER INHALER FORMULATIONS
    Jamal, A. J.
    Assi, K. H.
    Vangala, V. G.
    Alayoub, Y.
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2021, 34 (03) : A16 - A16
  • [22] DEM analysis of powder deaggregation and discharge from the capsule of a carrier-based Dry Powder Inhaler
    Alfano, Francesca Orsola
    Sommerfeld, Martin
    Di Maio, Francesco Paolo
    Di Renzo, Alberto
    ADVANCED POWDER TECHNOLOGY, 2022, 33 (12)
  • [23] Carrier-free high-dose dry powder inhaler formulation of ibuprofen: Physicochemical characterization and in vitro aerodynamic performance
    Yazdi, Ashkan K.
    Smyth, Hugh D. C.
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2016, 511 (01) : 403 - 414
  • [24] Modeling the performance of carrier-based dry powder inhalation formulations: Where are we, and how to get there?
    Elsayed, Mustafa M. A.
    Shalash, Ahmed O.
    JOURNAL OF CONTROLLED RELEASE, 2018, 279 : 251 - 261
  • [25] Influence of humidity on the electrostatic charge and aerosol performance of dry powder inhaler carrier based systems
    Young, Paul M.
    Sung, Adrian
    Traini, Daniela
    Kwok, Philip
    Chiou, Herbert
    Chan, Hak-Kim
    PHARMACEUTICAL RESEARCH, 2007, 24 (05) : 963 - 970
  • [26] Influence of Humidity on the Electrostatic Charge and Aerosol Performance of Dry Powder Inhaler Carrier based Systems
    Paul M Young
    Adrian Sung
    Daniela Traini
    Philip Kwok
    Herbert Chiou
    Hak-Kim Chan
    Pharmaceutical Research, 2007, 24 : 963 - 970
  • [27] Effect of Device Design on the Aerosolization of a Carrier-Based Dry Powder Inhaler—a Case Study on Aerolizer® Foradile®
    Qi Tony Zhou
    Zhenbo Tong
    Patricia Tang
    Mauro Citterio
    Runyu Yang
    Hak-Kim Chan
    The AAPS Journal, 2013, 15 : 511 - 522
  • [28] Effect of Sampling Inhalation Volume on Evaluation of Dispersion and Deposition Performance of a Carrier-based Dry Powder Inhaler by Three Different Impactors
    Lu, Huanhuan
    Wu, Fei
    Chen, Lan
    Chen, Donghao
    Chinese Journal of Pharmaceuticals, 2023, 54 (01) : 101 - 110
  • [29] Effect of carrier particle shape on dry powder inhaler performance
    Kaialy, Waseem
    Alhalaweh, Amjad
    Velaga, Sitaram P.
    Nokhodchi, Ali
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2011, 421 (01) : 12 - 23
  • [30] Recent developments in lactose blend formulations for carrier-based dry powder inhalation
    Hebbink, Gerald A.
    Jaspers, Maarten
    Peters, Harry J. W.
    Dickhoff, Bastiaan H. J.
    ADVANCED DRUG DELIVERY REVIEWS, 2022, 189