Dosimetric effects of intrafractional isocenter variation during deep inspiration breath-hold for breast cancer patients using surface-guided radiotherapy

被引:43
|
作者
Kugele, Malin [1 ,2 ]
Edvardsson, Anneli [2 ]
Berg, Lovisa [1 ]
Alkner, Sara [1 ]
Ljus, Carina Andersson [1 ]
Ceberg, Sofie [1 ,2 ]
机构
[1] Skane Univ Hosp, Dept Hematol Oncol & Radiat Phys, Lund, Sweden
[2] Lund Univ, Dept Clin Sci, Med Radiat Phys, Lund, Sweden
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2018年 / 19卷 / 01期
关键词
breath hold; intrafractional isocenter variation; optical surface scanning; treatment planning; RADIATION-THERAPY; HEART-DISEASE; RISK; WOMEN; REPRODUCIBILITY; MORTALITY; COHORT;
D O I
10.1002/acm2.12214
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to investigate potential dose reductions to the heart, left anterior descending coronary artery (LAD), and ipsilateral lung for left-sided breast cancer using visually guided deep inspiration breath-hold (DIBH) with the optical surface scanning system Catalyst (TM), and how these potential dosimetric benefits are affected by intrafractional motion in between breath holds. For both DIBH and free breathing (FB), treatment plans were created for 20 tangential and 20 locoregional left-sided breast cancer patients. During DIBH treatment, beam-on was triggered by a region of interest on the xiphoid process using a 3 mm gating window. Using a novel nonrigid algorithm, the Catalyst (TM) system allows for simultaneous real-time tracking of the isocenter position, which was used to calculate the intrafractional DIBH isocenter reproducibility. The 50% and 90% cumulative probabilities and maximum values of the intrafractional DIBH isocenter reproducibility were calculated and to obtain the dosimetric effect isocenter shifts corresponding to these values were performed in the treatment planning system. For both tangential and locoregional treatment, the dose to the heart, LAD and ipsilateral lung was significantly reduced for DIBH compared to FB. The intrafractional DIBH isocenter reproducibility was very good for the majority of the treatment sessions, with median values of approximately 1 mm in all three translational directions. However, for a few treatment sessions, intrafractional DIBH isocenter reproducibility of up to 5 mm was observed, which resulted in large dosimetric effects on the target volume and organs at risk. Hence, it is of importance to set tolerance levels on the intrafractional isocenter motion and not only perform DIBH based on the xiphoid process.
引用
收藏
页码:25 / 38
页数:14
相关论文
共 50 条
  • [1] Surface-guided radiotherapy for breast cancer in deep inspiration breath hold
    Beer, K. T.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S539 - S541
  • [2] Application and dosimetric comparison of surface-guided deep inspiration breath-hold for lung stereotactic body radiotherapy
    Wang, Jiaxin
    Dong, Tingting
    Meng, Xiangyin
    Li, Wenbo
    Li, Nan
    Wang, Yijun
    Yang, Bo
    Qiu, Jie
    MEDICAL DOSIMETRY, 2024, 49 (04) : 372 - 379
  • [3] Evaluation of image-guided and surface-guided radiotherapy for breast cancer patients treated in deep inspiration breath-hold: A single institution experience
    Penninkhof, Joan
    Fremeijer, Kimm
    Harten, Kirsten Offereins-van
    van Wanrooij, Cynthia
    Quint, Sandra
    Kunnen, Britt
    Hoffmans-Holtzer, Nienke
    Swaak, Annemarie
    Baaijens, Margreet
    Dirkx, Maarten
    TECHNICAL INNOVATIONS & PATIENT SUPPORT IN RADIATION ONCOLOGY, 2022, 21 : 51 - 57
  • [4] Dosimetric and clinical advantages of deep inspiration breath-hold (DIBH) during radiotherapy of breast cancer
    Vicente Bruzzaniti
    Armando Abate
    Paola Pinnarò
    Marco D’Andrea
    Erminia Infusino
    Valeria Landoni
    Antonella Soriani
    Carolina Giordano
    Anna Maria Ferraro
    Lidia Strigari
    Journal of Experimental & Clinical Cancer Research, 32
  • [5] Dosimetric and clinical advantages of deep inspiration breath-hold (DIBH) during radiotherapy of breast cancer
    Bruzzaniti, Vicente
    Abate, Armando
    Pinnaro, Paola
    D'Andrea, Marco
    Infusino, Erminia
    Landoni, Valeria
    Soriani, Antonella
    Giordano, Carolina
    Ferraro, Anna Maria
    Strigari, Lidia
    JOURNAL OF EXPERIMENTAL & CLINICAL CANCER RESEARCH, 2013, 32
  • [6] Accuracy and potential improvements of surface-guided breast cancer radiotherapy in deep inspiration breath-hold with daily image-guidance
    Nankali, Saber
    Hansen, Rune
    Worm, Esben
    Yates, Esben S.
    Thomsen, Mette Skovhus
    Offersen, Birgitte
    Poulsen, Per Rugaard
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (19):
  • [7] Dose coverage and breath-hold analysis of breast cancer patients treated with surface-guided radiotherapy
    Dekker, Janita
    Essers, Marion
    Verheij, Marcel
    Kusters, Martijn
    de Kruijf, Willy
    RADIATION ONCOLOGY, 2023, 18 (01)
  • [8] Dose coverage and breath-hold analysis of breast cancer patients treated with surface-guided radiotherapy
    Janita Dekker
    Marion Essers
    Marcel Verheij
    Martijn Kusters
    Willy de Kruijf
    Radiation Oncology, 18
  • [9] Impact of deep inspiration breath hold, surface-guided radiotherapy, and daily CBCT on the organs at risk in breast cancer radiotherapy
    Rudat, Volker
    Zhao, Ruping
    Wang, Bin
    Zhang, Lihua
    Shi, Yanyan
    SCIENTIFIC REPORTS, 2024, 14 (01):
  • [10] Surface-guided radiotherapy systems in locoregional deep inspiration breath hold radiotherapy for breast cancer - a multicenter study on the setup accuracy
    Laaksomaa, Marko
    Aula, Antti
    Sarudis, Sebastian
    Keyrilainen, Jani
    Ahlroth, Jenni
    Murtola, Anna
    Pynnonen, Kiira
    Lehtonen, Turkka
    Bjorkqvist, Mikko
    Jarvinen, Lauri
    Rossi, Maija
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2024, 29 (02) : 176 - 186