共 31 条
Percutaneous Microwave Tumor Ablation Is Safe in Patients with Cardiovascular Implantable Electronic Devices: A Single-Institutional Retrospective Review
被引:3
|作者:
Swietlik, John F.
[1
]
Longo, Katherine C.
[1
]
Knott, Emily A.
[1
]
Hinshaw, J. Louis
[1
]
Brace, Christopher L.
[2
]
Alexander, Marci
[1
]
Lubner, Meghan G.
[1
]
Wells, Shane A.
[1
]
Lee, Fred T., Jr.
[1
]
Ziemlewicz, Timothy J.
[1
]
机构:
[1] Univ Wisconsin, Dept Radiol, 600 Highland Ave, Madison, WI 53792 USA
[2] Univ Wisconsin, Dept Med Phys, 600 Highland Ave, Madison, WI 53792 USA
关键词:
RADIOFREQUENCY ABLATION;
CARDIAC DEVICES;
INTERFERENCE;
EFFICACY;
D O I:
10.1016/j.jvir.2018.12.001
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
The risk of electromagnetic interference between microwave (MW) ablation and cardiac implantable electronic devices (CIEDs), ie, pacemakers and defibrillators, has not been fully evaluated. Fourteen MW ablations (kidney, n = 8; liver, n = 5; lung, n = 1) were performed in 13 patients with CIEDs in normal operating mode. Electrocardiography tracings, cardiovascular complications, tumor size, tumor-to-CIED distance, and tumor-to-device lead distance were recorded. Mean tumor size was 2.9 cm, mean tumor-to-CIED distance was 26.4 cm (range, 9-30 cm), and mean tumor-to-lead distance was 12.1 cm (range, 3.5-20 cm). No device-based cardiovascular complications or class C or higher complications per Society of Interventional Radiology criteria were identified. MW ablation appears to be safe in select patients with CIEDs.