Lowering the limit of detection in high spatial resolution electron beam microanalysis with the microcalorimeter energy dispersive X-ray spectrometer

被引:19
|
作者
Newbury, D
Wollman, D
Irwin, K
Hilton, G
Martinis, J
机构
[1] Natl Inst Stand & Technol, Gaithersburg, MD 20899 USA
[2] Natl Inst Stand & Technol, Boulder, CO 80303 USA
关键词
electron probe X-ray microanalysis; energy dispersive X-ray spectrometry; microanalysis; microcalorimetry; scanning electron microscopy; X-ray microanalysis; X-ray spectrometry;
D O I
10.1016/S0304-3991(99)00028-5
中图分类号
TH742 [显微镜];
学科分类号
摘要
Low-beam-energy X-ray microanalysis with the field-emission-gun scanning electron microscope suffers limitations due to physical factors of X-ray generation. Instrumental limitations are imposed by the poor resolution of the conventional semiconductor energy dispersive X-ray spectrometry. Wavelength dispersive X-ray spectrometry provides sufficient resolution to solve spectroscopic problems, but the poor geometric efficiency and the single channel nature of spectrum measurement restrict its practical use for low-beam-energy microanalysis. The microcalorimeter energy dispersive X-ray spectrometer combines high resolution ( < 10 eV) with energy dispersive operation. The utility of this new spectrometer is examined in four categories: (1) qualitative analysis; (2) quantitative analysis; (3) chemical speciation studies, and (4) measurement of trace constituents. For the low-beam-energy regime, the microcalorimeter energy dispersive X-ray spectrometer provides important new performance capabilities for qualitative analysis and chemical speciation studies. However, there are limitations for quantitative studies imposed by detector geometrical characteristics. In addition, trace element detection is impractical below 0.001 mass fraction with low-beam energy ( < 5 keV) because of count rate limitations. Published by Elsevier Science B.V.
引用
收藏
页码:73 / 88
页数:16
相关论文
共 50 条
  • [31] Energy dispersive X-ray microanalysis for characterization of biomineralization processes
    Wiesmann, HP
    Kleinheinz, J
    Adelt, T
    Joos, U
    Plate, U
    Hohling, HJ
    Zierold, K
    [J]. EUROPEAN JOURNAL OF CELL BIOLOGY, 1997, 74 : 105 - 105
  • [32] What is the Best Beam Energy for X-Ray Microanalysis of Nanomaterials in Electron Microscopy?
    Gauvin, Raynald
    Michaud, Pierre
    Trudeau, Michel L.
    [J]. MICROSCOPY AND MICROANALYSIS, 2009, 15 : 460 - 461
  • [33] High spatial resolution X-ray beam position monitor
    Kocsis, M.
    Fajardo, P.
    [J]. 2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 1223 - 1225
  • [34] LIMIT OF DETECTABILITY IN X-RAY ELECTRON-PROBE MICROANALYSIS
    KOTRBA, Z
    [J]. MIKROCHIMICA ACTA, 1977, 2 (3-4) : 97 - 106
  • [35] Spatial resolution of electron probe X-ray microanalysis on sections of organic (biological) material
    Krefting, ER
    Felsmann, M
    Recker, A
    Feja, B
    Höhling, HJ
    Reichelt, R
    Reimer, L
    [J]. ULTRAMICROSCOPY, 1999, 77 (1-2) : 13 - 21
  • [36] Specific Features of Energy Dispersive X-Ray Electron Probe Microanalysis in the Low Vacuum Mode
    D. E. Pukhov
    S. V. Kurbatov
    [J]. Journal of Analytical Chemistry, 2018, 73 : 249 - 256
  • [37] THE APPLICATION OF SCANNING ELECTRON MICROSCOPY AND X-RAY ENERGY DISPERSIVE MICROANALYSIS TO THE STUDY OF THE HUMAN TESTIS
    Siew, Shirley
    [J]. ANDROLOGY, 2013, 1 : 76 - 76
  • [38] Specific Features of Energy Dispersive X-Ray Electron Probe Microanalysis in the Low Vacuum Mode
    Pukhov, D. E.
    Kurbatov, S. V.
    [J]. JOURNAL OF ANALYTICAL CHEMISTRY, 2018, 73 (03) : 249 - 256
  • [39] NON-DISPERSIVE X-RAY MICROANALYSIS WITH ENERGY SELECTION ON SCANNING ELECTRON-MICROSCOPE
    MENY, L
    CHAMPIGN.M
    [J]. JOURNAL DE MICROSCOPIE, 1972, 13 (03): : 287 - +
  • [40] Development of a high energy resolution X-ray microcalorimeter using Ti/Au TES
    Kudo, H
    Yokoyama, Y
    Shoji, S
    [J]. DESIGN, CHARACTERIZATION, AND PACKAGING FOR MEMS AND MICROELECTRONICS, 1999, 3893 : 241 - 249