Development of low power high speed readout electronics for high resolution PET with LSO and APD arrays

被引:0
|
作者
Visser, G [1 ]
Cherry, S [1 ]
Clajus, M [1 ]
Shao, YP [1 ]
Tümer, TO [1 ]
机构
[1] NOVA R&D Inc, Riverside, CA 92507 USA
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
An ASIC-based readout electronics scheme is under development for high resolution, compact PET imagers based on independent readout of all channels of LSO scintillator and avalanche photodiode (APD) arrays. Depth of interaction is obtained by readout of both ends of the LSO crystals. A low power, highly integrated design is critical. We report here on a discrete electronics prototype, running at 22 mW per channel for the preamplifier and discriminator. The measured timing resolution is 3.6 ns FWHM, 9.2 ns full width at one tenth maximum, relative to an LSO/PMT detector, energy resolution is 13.3 % FWHM at 511 keV, and depth of interaction position resolution is 2.5 mm FWHM throughout the full length of the crystal. Work is in progress towards an ASIC and a full scale prototype detector module suitable for a high resolution compact PET instrument.
引用
收藏
页码:1988 / 1992
页数:5
相关论文
共 50 条
  • [1] Planar APD arrays for high resolution PET
    Shah, KS
    Farrell, R
    Cirignano, L
    Grazioso, R
    Bennett, P
    Cherry, SR
    Shao, YP
    MEDICAL APPLICATIONS OF PENETRATING RADIATION, 1999, 3770 : 104 - 111
  • [2] Studies with a prototype high resolution PET scanner based on LSO-APD modules
    Klinikum rechts der Isar der, Technischen Universitaet Muenchen, Muenchen, Germany
    IEEE Transactions on Nuclear Science, 1998, 45 (3 pt 2): : 1298 - 1302
  • [3] Studies with a prototype high resolution PET scanner based on LSO-APD modules
    Pichler, B
    Boning, G
    Lorenz, E
    Mirzoyan, R
    Pimpl, W
    Schwaiger, M
    Ziegler, SI
    1997 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1 & 2, 1998, : 1649 - 1653
  • [4] Studies with a prototype high resolution PET scanner based on LSO-APD modules
    Pichler, B
    Boning, G
    Lorenz, E
    Mirzoyan, R
    Pimpl, W
    Schwaiger, M
    Ziegler, SI
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) : 1298 - 1302
  • [6] Design studies of a high resolution PET detector using APD arrays
    Shao, Y
    Silverman, RW
    Farrell, R
    Cirignano, L
    Grazioso, R
    Shah, KS
    Visser, G
    Clajus, M
    Tümer, TO
    Cherry, SR
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (03) : 1051 - 1057
  • [7] Development of readout electronics for high resolution neutron scintillator detector
    Zhou S.
    Huang H.
    Wan Z.
    Chen S.
    Zhu Z.
    Tang B.
    Huang C.
    Wang X.
    Zeng L.
    Yu L.
    Yang H.
    Liu H.
    Yue X.
    Zhao Y.
    Sun Z.
    He Jishu/Nuclear Techniques, 2023, 46 (11):
  • [8] Investigation of low noise, low cost readout electronics for high sensitivity PET systems based on avalanche photodiode arrays
    Habte, F
    Levin, CS
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 661 - 665
  • [9] MadPET: High resolution animal pet with avalanche photodiode arrays and LSO crystals.
    Ziegler, SI
    Pichler, BJ
    Boening, G
    Rafecas, H
    Bieler, S
    Pimpl, W
    Lorenz, E
    Schwaiger, M
    JOURNAL OF NUCLEAR MEDICINE, 2000, 41 (05) : 20P - 20P
  • [10] Hamamatsu S8550 APD arrays for high-resolution scintillator matrices readout
    Kapusta, M
    Crespo, P
    Wolski, D
    Moszynski, M
    Enghardt, W
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 504 (1-3): : 139 - 142