Evaluation of the Method for Analyzing Chromium, Cobalt and Titanium Ion Levels in the Blood Following Hip Replacement with a Metal-on-Metal Prosthesis

被引:20
|
作者
Barry, Janie [1 ]
Lavigne, Martin [1 ]
Vendittoli, Pascal-Andre [1 ]
机构
[1] Univ Montreal, Dept Surg, Maisonneuve Rosemont Hosp, Montreal, PQ H3C 3J7, Canada
关键词
PLASMA-MASS SPECTROMETRY; DIOXIDE NANOPARTICLES; ARTHROPLASTY; RELEASE; WEAR; OSTEOARTHRITIS; MORTALITY; BEARINGS;
D O I
10.1093/jat/bks090
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
High-resolution inductively coupled plasma mass spectrometry (HR-ICP-MS) can be used to measure metal ion levels in whole blood, although the accuracy and repeatability of this method is not yet known in the clinical setting. In this study, chromium, cobalt and titanium ion levels were measured in three whole blood samples, each collected at the same moment from 101 patients undergoing total hip arthroplasty with a metal-on-metal (MoM) bearing. The first sample (purge sample) had direct contact with the metal needle used during insertion of the catheter, whereas the second sample (reference sample) and the third sample (reserve sample) did not. The absolute difference between reference samples and reserve samples was greater than the limit of quantification of the HR-ICP-MS device for all three ions (0.84 versus 0.35 g/L for chromium, 0.74 versus 0.07 for cobalt, and 0.88 versus 0.70 g/L for titanium), although the levels were very small in most cases, they exceeded the clinical significant threshold in 19 to 31 of the cases. No clinically significant difference was observed between reference samples and purge samples. Therefore, HR-ICP-MS is a clinically acceptable method to evaluate metal ion levels in blood following hip replacement with an MoM prosthesis, and it is not necessary to discard the purge sample to obtain repeatable results.
引用
收藏
页码:90 / 96
页数:7
相关论文
共 50 条
  • [21] Low Cobalt and Chromium Levels from a Single Metal-on-Metal Total Hip Arthroplasty Implant
    Chen, Antonia F.
    Pigott, Matthew T.
    McClain, Edward J.
    Klatt, Brian A.
    [J]. METAL-ON-METAL TOTAL HIP REPLACEMENT DEVICES, 2013, 1560 : 61 - 72
  • [22] Chromium speciation in the blood of metal-on-metal hip implant patients
    Finley, Brent
    Scott, Paul K.
    Glynn, Meghan E.
    Paustenbach, Dennis
    Donovan, Ellen
    Thuett, Kerry A.
    [J]. TOXICOLOGICAL AND ENVIRONMENTAL CHEMISTRY, 2017, 99 (01): : 48 - 64
  • [23] The clinical significance of metal ion release from cobalt-chromium metal-on-metal hip joint arthroplasty
    Cobb, AG
    Schmalzreid, TP
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) : 385 - 398
  • [24] Detection of cobalt in synovial fluid from metal-on-metal hip prosthesis: correlation with the ion haematic level
    Beraudi, Alina
    Catalani, Simona
    Montesi, Monica
    Stea, Susanna
    Sudanese, Alessandra
    Apostoli, Pietro
    Toni, Aldo
    [J]. BIOMARKERS, 2013, 18 (08) : 699 - 705
  • [25] The value of serial metal ion levels in following up patients with metal-on-metal hip arthroplasty
    McHugh, Gavin
    Merchant, Rajiv
    Kelly, Gabrielle E.
    Bergin, Karen M.
    McCoy, Gerald F.
    Wozniak, Andrew P.
    Quinlan, John F.
    [J]. HIP INTERNATIONAL, 2017, 27 (01) : 14 - 20
  • [26] Ion production and excretion in a patient with a metal-on-metal bearing hip prosthesis
    Bitsch, Rudi G.
    Zamorano, Michelle
    Loidolt, Travis
    Heisel, Christian
    Jacobs, Joshua J.
    Schmalzried, Thomas P.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (12): : 2758 - 2763
  • [27] Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement
    De Haan, R.
    Pattyn, C.
    Gill, H. S.
    Murray, D. W.
    Campbell, P. A.
    De Smet, K.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (10): : 1291 - 1297
  • [28] No association between blood count levels and whole-blood cobalt and chromium levels in 1,900 patients with metal-on-metal hip arthroplasty
    Honkasaari, Noora
    Lainiala, Olli
    Laine, Outi
    Reito, Aleksi
    Eskelinen, Antti
    [J]. ACTA ORTHOPAEDICA, 2020, 91 (06) : 711 - 716
  • [29] Circulating levels of cobalt and chromium from metal-on-metal hip replacement are associated with CD8+ T-cell lymphopenia
    Hart, A. J.
    Skinner, J. A.
    Winship, P.
    Faria, N.
    Kulinskaya, E.
    Webster, D.
    Muirhead-Allwood, S.
    Aldam, C. H.
    Anwar, H.
    Powell, J. J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2009, 91B (06): : 835 - 842
  • [30] How do serum cobalt and chromium levels change after metal-on-metal hip resurfacing? Reply
    Back, D. L.
    Young, D. A.
    Shimmin, A.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (451) : 293 - 293